Design and development strategy for the rotary blood pump

被引:54
|
作者
Nose, Y [1 ]
机构
[1] Baylor Coll Med, Dept Surg, Houston, TX 77030 USA
关键词
development strategy of blood pump; rotary blood pump; axial flow pump; centrifugal blood pump; 2 day pump; 2 week pump; permanent pump;
D O I
10.1046/j.1525-1594.1998.06098.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Development of an antitraumatic antithrombogenic and durable blood pump is a very difficult task. Based upon this author's experience of over 35 years in the development of various types of cardiac prostheses, development strategies for a rotary blood pump are described. A step-by-step development strategy is thus proposed. Initially, the development of a 2 day antitraumatic pump (Phase 1) would be made. Then, conversion of this pump to a 2 week antithrombogenic pump (Phase 2) should be attempted. After the successful development of the Phase 2 pump, the conversion of this device to a durable, implantable, and long-term blood pump (Phase 3) should be established. Based upon this development strategy, 2 rotary blood pumps, namely, the axial Row blood pump and the centrifugal blood pump, have been developed in less than 4 pears with modest development costs.
引用
收藏
页码:438 / 446
页数:9
相关论文
共 50 条
  • [11] The Progress in the Novel Pediatric Rotary Blood Pump Sputnik Development
    Telyshev, Dmitry
    Denisov, Maxim
    Pugovkin, Alexander
    Selishchev, Sergey
    Nesterenko, Igor
    [J]. ARTIFICIAL ORGANS, 2018, 42 (04) : 432 - 443
  • [12] Development of the MEDOS/HIA DeltaStream extracorporeal rotary blood pump
    Göbel, C
    Arvand, A
    Eilers, R
    Marseille, O
    Bals, C
    Meyns, B
    Flameng, W
    Rau, G
    Reul, H
    [J]. ARTIFICIAL ORGANS, 2001, 25 (05) : 358 - 365
  • [13] Development of a non-pulsatile permanent rotary blood pump
    Nose, Y
    Kawahito, K
    [J]. EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 1997, 11 : S32 - S38
  • [14] Development of rotary blood pump technology:: Past, present, and future
    Nosé, Y
    Yoshikawa, M
    Murabayashi, S
    Takano, T
    [J]. ARTIFICIAL ORGANS, 2000, 24 (06) : 412 - 420
  • [15] Blood transfusion with rotary pump
    Darmady, EM
    [J]. BRITISH MEDICAL JOURNAL, 1939, 1939 : 1304 - 1305
  • [16] Blood transfusion with the rotary pump
    Riddell, V
    [J]. BRITISH MEDICAL JOURNAL, 1939, 1939 : 1125 - 1129
  • [17] Modeling of a Rotary Blood Pump
    Nestler, Frank
    Bradley, Andrew P.
    Wilson, Stephen J.
    Timms, Daniel L.
    [J]. ARTIFICIAL ORGANS, 2014, 38 (03) : 182 - 190
  • [18] DESIGN OF TOTALLY HYDRODYNAMICALLY SUSPENDED IMPELLERS FOR MINIATURIZED ROTARY BLOOD PUMP
    Amaral, F. K.
    Timms, D.
    Bohning, F.
    Graefe, R.
    Schmitz-Rode, T.
    Steinseifer, U.
    [J]. INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2010, 33 (07): : 425 - 425
  • [19] OPTIMIZATION OF THE DESIGN OF A ROTARY PUMP
    Alibekov, S. Ya.
    Krasheninnikov, M. V.
    [J]. CHEMICAL AND PETROLEUM ENGINEERING, 2013, 49 (7-8) : 539 - 541
  • [20] Analysis of a new PM motor design for a rotary dynamic blood pump
    Xu, LY
    Wang, FX
    Fu, MH
    Medvedev, A
    Smith, WA
    Golding, LAR
    [J]. ASAIO JOURNAL, 1997, 43 (05) : M559 - M564