The Progress in the Novel Pediatric Rotary Blood Pump Sputnik Development

被引:26
|
作者
Telyshev, Dmitry [1 ]
Denisov, Maxim [1 ]
Pugovkin, Alexander [1 ]
Selishchev, Sergey [1 ]
Nesterenko, Igor [1 ]
机构
[1] Natl Res Univ Elect Technol, Biomed Syst Dept, Moscow 124635, Russia
关键词
Congestive heart failure; Mechanical circulatory support; Mock circulatory loop; Pediatric rotary blood pump; VENTRICULAR ASSIST DEVICE; COMPUTATIONAL FLUID-DYNAMICS; HEARTMATE II; FLOW; DESIGN;
D O I
10.1111/aor.13109
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In this work, the study results of an implantable pediatric rotary blood pump (PRBP) are presented. They show the results of the numerical simulation of fluid flow rates in the pump. The determination method of the backflows and stagnation regions is represented. The operating points corresponding to fluid flow rates of 1, 3, and 5 L/min for 75-80 mm Hg pressure head are investigated. The study results have shown that use of the pump in the 1 L/min operating point can potentially lead to the appearance of backflows and stagnation regions. In the case of using pumps in fluid flow rates ranging from 3 to 5 L/min, the number of stagnation regions decreases and the fluid flow rate changes marginally. Using the pump in this flow rate range is considered judicious. The study shows an increase in shear stress with an increase in fluid flow rates, while there is no increase in shear stress above the critical condition of 150 Pa (which does not allow us to reliably speak about the increased risk of blood cell damage). The aim of this work was to design, prototype, and study interaction of the Sputnik PRBP with the cardiovascular system. A three-dimensional model of Sputnik PRBP was designed with the following geometrical specifications: flow unit length of 51.5 mm, flow unit diameter of 10 mm, and spacing between the rotor and housing of 0.1 mm. Computational fluid dynamics studies were used to calculate head pressure-flow rate (H-Q) curves at rotor speeds ranging from 10 000 to 14 000 rpm (R-2 = 0.866 between numerical simulation and experiment) and comparing flow patterns at various points of the flow rate operating range (1, 3, and 5 L/min) for operating pressures ranging from 75 to 80 mm Hg. It is noted that when fluid flow rate changes from 1 L/min to 3 L/min, significant changes are observed in the distribution of zero flow zones. At the inlet and outlet of the pump, when going to the operating point of 3 L/min, zones of stagnation become minuscule. The shear stress distribution was calculated along the pump volume. The volume in which shear stress exceed 150 Pa is less than 0.38% of the total pump volume at flow rates of 1, 3, and 5 L/min. In this study, a mock circulatory system (MCS) allowing simulation of physiological cardiovascular characteristics was used to investigate the interaction of the Sputnik PRBP with the cardiovascular system. MCS allows reproducing the Frank-Starling autoregulation mechanism of the heart. PRBP behavior was tested in the speed range of 6 000 to 15 000 rpm. Decreased contractility can be expressed in a stroke volume decrease approximately from 18 to 4 mL and ventricle systolic pressure decrease approximately from 92 to 20 mm Hg. The left ventricle becomes fully supported at a pump speed of 10 000 rpm. At a pump speed of 14 000 rpm, the left ventricle goes into a suction state in which fluid almost does not accumulate in the ventricle and only passes through it to the pump. The proposed PRBP showed potential for improved clinical outcomes in pediatric patients with a body surface area greater than 0.6 m(2) and weight greater than 12 kg.
引用
收藏
页码:432 / 443
页数:12
相关论文
共 50 条
  • [1] Computational Fluid Dynamics Simulation of the Sputnik Pediatric Rotary Blood Pump
    Telyshev, Dmitry
    Denisov, Maxim
    Satyukova, Anna
    Le, Tatyana
    [J]. PROCEEDINGS OF 2018 IEEE EAST-WEST DESIGN & TEST SYMPOSIUM (EWDTS 2018), 2018,
  • [2] Development of Medical and Technical Requirements and Simulation of the Flow−Pressure Characteristics of the Sputnik Pediatric Rotary Blood Pump
    Denisov M.V.
    Selishchev S.V.
    Telyshev D.V.
    Frolova E.A.
    [J]. Biomedical Engineering, 2017, 50 (5) : 296 - 299
  • [3] Experimental Determination of the Normalized Index of Hemolysis for the Sputnik Implantable Pediatric Rotary Blood Pump
    Bockeria L.A.
    Bockeria O.L.
    Selishchev S.V.
    Telyshev D.V.
    Le T.G.
    Satyukova A.S.
    Shvartz V.A.
    Glushko L.A.
    [J]. Biomedical Engineering, 2017, 50 (6) : 416 - 419
  • [4] The Effect of Rotor Geometry on the H−Q Curves of the Sputnik Implantable Pediatric Rotary Blood Pump
    Telyshev D.V.
    Denisov M.V.
    Selishchev S.V.
    [J]. Biomedical Engineering, 2017, 50 (06) : 420 - 424
  • [5] Analysis of the preload and afterload sensitivity of the sputnik rotary blood pump
    Petukhov D.S.
    Telyshev D.V.
    [J]. Biomedical Engineering, 2016, 49 (6) : 362 - 365
  • [6] Characterization of a pediatric rotary blood pump
    de Melo, Thamiles Rodrigues
    Vasconcelos, Felipe José de Sousa
    Ribeiro, Luiz Henrique Ramalho Diniz
    Bacht, Simão
    Cestari, Idágene Aparecida
    Neto, José Sérgio da Rocha
    Lima, Antonio Marcus Nogueira
    [J]. Research on Biomedical Engineering, 2018, 34 (04) : 299 - 309
  • [7] A novel rotary blood pump for HFpEF
    Gude, E.
    Fiane, D. R.
    [J]. EUROPEAN JOURNAL OF HEART FAILURE, 2019, 21 : 348 - 348
  • [8] A Novel Rotary Blood Pump for HFpEF
    Gude, E.
    Fiane, A.
    [J]. JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2019, 38 (04): : S342 - S342
  • [9] Development of a novel shrouded impeller pediatric blood pump
    Khan, Talha Irfan
    Zad, Haris Sheh
    Lazoglu, Ismail
    Yalcin, Ozlem
    [J]. JOURNAL OF ARTIFICIAL ORGANS, 2018, 21 (02) : 142 - 149
  • [10] Development of a novel centrifugal pump: Magnetic rotary pump
    Naganuma, S
    Yambe, T
    Sonobe, T
    Kobayashi, S
    Nitta, S
    [J]. ARTIFICIAL ORGANS, 1997, 21 (07) : 746 - 750