Evaluation of a physiologic pulsatile pump system for neonate-infant cardiopulmonary bypass support

被引:0
|
作者
Ündar, A
Masai, T
Inman, R
Beyer, EA
Mueller, MA
McGarry, MC
Frazier, OH
Fraser, CD
机构
[1] Texas Childrens Hosp, Baylor Coll Med, Congenital Heart Surg Serv, Houston, TX 77030 USA
[2] Baylor Coll Med, Texas Heart Inst, Cullen Cardiovasc Surg Res Labs, Houston, TX 77030 USA
[3] Baylor Coll Med, Dept Surg, Sect Congenital Heart Surg, Houston, TX 77030 USA
关键词
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
An alternate physiologic pulsatile pump (PPP) system was designed and evaluated to produce sufficient pulsatility during neonate-infant open heart surgery. This hydraulically driven pump system has a unique "dual" pumping chamber mechanism. The first chamber is placed between the venous reservoir and oxygenator and the second chamber between the oxygenator and patient. Each chamber has two unidirectional tricuspid valves. Stroke volume (0.2-10 ml), upstroke rise time (10-350 msec), and pump rate (2-250 beats per minute [bpm]) can be adjusted independently to produce adequate pulsatility. This system has been tested in 3-kg piglets (n = 6), with a pump flow of 150 ml/kg/min, a pump rate of 150 bpm, and a pump ejection time of 110 msec. After initiation of cardiopulmonary bypass (CPB), all animals were subjected to 25 minutes of hypothermia to reduce the rectal temperatures to 18 degrees C, 60 minutes of deep hypothermic circulatory arrest (DHCA), then 10 minutes of cold perfusion with a full pump flow, and 40 minutes of rewarming. During CPB, mean arterial pressures were kept at less than 50 mm Hg. Mean extracorporeal circuit pressure (ECCP), the pressure drop of a 10 French aortic cannula, and the pulse pressure were 67 +/- 9, 21 +/- 6, and 16 +/- 2 mm Hg, respectively. All values are represented as mean +/- SD. No regurgitation or abnormal hemolysis has been detected during these experiments. The oxygenator had no damping effect on the quality of the pulsatility because of the dual chamber pumping mechanism. The ECCP was also significantly lower than any other known pulsatile system. We conclude that this system, with a 10 French aortic cannula and arterial filter, produces adequate pulsatility in 3 kg piglets.
引用
收藏
页码:53 / 58
页数:6
相关论文
共 50 条
  • [41] Preclinical evaluation of the Kyocera Gyro centrifugal blood pump for cardiopulmonary bypass
    Takami, Yoshiyuki
    Ohara, Yasuhisa
    Otsuka, Goro
    Nakazawa, Tadashi
    Nose, Yukihiko
    PERFUSION-UK, 1997, 12 (05): : 335 - 341
  • [42] DEVELOPMENT AND EVALUATION OF A NEWLY DESIGNED PULSATILE FLOW DEVICE FOR CARDIOPULMONARY BYPASS AND ASSISTED CIRCULATION
    TAKIDO, N
    FUJIMASA, I
    IMACHI, K
    SATO, N
    MIYAKE, H
    IWAI, N
    KOUNO, A
    ATSUMI, K
    PHYSICS IN MEDICINE AND BIOLOGY, 1980, 25 (05): : 973 - 973
  • [43] Evaluation of the results of patients who underwent coronary bypass grafting with or without cardiopulmonary bypass pump
    P Kwinecki
    T Stankowski
    A Kędziora
    A Pawęzowska
    SS Aboul-Hassan
    R Cichoń
    Journal of Cardiothoracic Surgery, 8 (Suppl 1)
  • [44] Clinical evaluation of the Gyro Pump C1e3 as a cardiopulmonary bypass pump
    Orime, Y
    Shiono, M
    Yagi, S
    Yamamoto, T
    Okumura, H
    Kimura, S
    Hata, M
    Sezai, A
    Kashiwazaki, S
    Choh, S
    Negishi, N
    Sezai, Y
    Matsui, T
    Suzuki, M
    ASAIO JOURNAL, 2000, 46 (01) : 128 - 133
  • [45] Novel Pulsatile Diagonal Pump for Pediatric Extracorporeal Life Support System
    Wang, Shigang
    Kunselman, Allen R.
    Uendar, Akif
    ARTIFICIAL ORGANS, 2013, 37 (01) : 37 - 47
  • [46] Microvascular fluid exchange during pulsatile cardiopulmonary bypass perfusion with the combined use of a nonpulsatile pump and intra-aortic balloon pump
    Lundemoen, Steinar
    Kvalheim, Venny Lise
    Mongstad, Arve
    Andersen, Knut Sverre
    Grong, Ketil
    Husby, Paul
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2013, 146 (05): : 1275 - 1282
  • [47] Off-pump coronary artery bypass surgery versus standard linear or pulsatile cardiopulmonary bypass: endothelial activation and inflammatory response
    Onorati, Francesco
    Rubino, Antonino S.
    Nucera, Sergio
    Foti, Daniela
    Sica, Vincenzo
    Santini, Francesco
    Gulletta, Elio
    Renzulli, Attilio
    EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2010, 37 (04) : 897 - 904
  • [48] BODY PERFUSION DURING ADULT CARDIOPULMONARY BYPASS IS IMPROVED BY PULSATILE FLOW WITH INTRA-AORTIC BALLOON PUMP
    Onorati, F.
    Santarpino, G.
    Caroleo, S.
    Dardano, A.
    Scalas, C.
    Gulletta, E.
    Santangelo, E.
    Renzutli, A.
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2008, 31 (07): : 619 - 619
  • [49] Body perfusion during adult cardiopulmonary bypass is improved by pulsatile flow with intra-aortic balloon pump
    Onorati, F.
    Santarpino, G.
    Rubino, A. S.
    Caroleo, S.
    Dardano, A.
    Scalas, C.
    Gulletta, E.
    Santangelo, E.
    Renzulli, A.
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2009, 32 (01): : 50 - 61
  • [50] Augmentation of abdominal organ perfusion during cardiopulmonary bypass with a novel intra-aortic pulsatile catheter pump
    Gu, YJ
    De Kroon, T
    Elstrodt, JM
    van Oeveren, W
    Boonstra, PW
    Rakhorst, G
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2005, 28 (01): : 35 - 43