Permeability of gloves used in nuclear medicine departments to [99mTc]-pertechnetate and [18F]-fluorodeoxyglucose: Radiation protection considerations

被引:1
|
作者
Ridone, S. [1 ]
Matheoud, R. [1 ]
Valzano, S. [1 ]
Di Martino, R. [2 ]
Vigna, L. [1 ]
Brambilla, M. [1 ]
机构
[1] Univ Hosp Maggiore Della Carita Novara, Dept Med Phys, I-28100 Novara, Italy
[2] Univ Oriental Piedmont, Sch Med, I-28100 Novara, Italy
来源
关键词
Gloves; Permeability [(99)mTc]-pertechnetate; F-18]-FDG; Radiation protection; SKIN CONTAMINATION; RADIONUCLIDE;
D O I
10.1016/j.ejmp.2013.01.005
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In order to evaluate the safety of the individual protection devices, the permeability of four different types of disposable gloves, commonly used in hospitals, was tested in relation to [Tc-99m]-pertechnetate and to [F-18]-fluorodeoxyglucose ([F-18]-FDG). From these radiopharmaceutical solutions, a drop was deposited on the external surface of the glove which was opened and stretched with the external surface placed upward. The smear test technique permitted to evaluate the activity onto the inner surface of the glove at different times. The smear tests were measured in a well sodium iodide detector calibrated in efficiency for Tc-99m and F-18. The permeability was tested on ten samples of each type of gloves and was expressed as the ratio of the activity onto the inner surface at each time interval to the activity deposited on the external surface of the glove. For each type of gloves and for each sampling time, mean value, standard deviation and percentage coefficient of variation of permeability were evaluated. One type of gloves showed a low resistance to permeation of both radiopharmaceuticals, while another one only to pertechnetate. The other gloves were good performers. The results of this study suggest to test permeability for gloves used for handling radiopharmaceuticals, before their adoption in the clinical routine. This practice will provide a more careful service of radiation protection for nuclear medicine department staff. (C) 2013 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:545 / 548
页数:4
相关论文
共 50 条
  • [1] Permeability of gloves used in Nuclear Medicine Departments to pertechentate and [18F]-fluorodeoxyglucose for staff radiation protection
    Matheoud, R.
    Ridone, S.
    Valzano, S.
    Vigna, L.
    Brambilla, M.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2012, 39 : S392 - S392
  • [2] Assessment of Radiation Exposure Dose for Nuclear Medicine Workers from 18F-FDG, 99mTc MDP, and 99mTc
    Al-esaei, Abdulrahman M.
    Khalil, Magdy M.
    El Shazly, Raed M.
    Kany, Amr M. I.
    Saleh, Emran Eisa
    Elmaghraby, Sharif
    CURRENT RADIOPHARMACEUTICALS, 2022, 15 (04) : 320 - 326
  • [3] Radiation dose to the hands and to the eyes of the technologists during hot lab work with 99mTc and 18F in a nuclear medicine practice
    Glomset, O
    Balazs, A
    Johansen, T
    Forfang, R
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2004, 31 : S456 - S457
  • [4] [18F]Fluorocholine PET/CT significantly reduces the amount of negative scan results compared to [99mTc]sestamibi and [99mTc]-pertechnetate scintigraphy SPECT/CT
    Rode, M.
    Moeller, H. Reilev
    Oestergaard, L. Lange
    Holdgaard, P.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2024, 51 : S154 - S154
  • [5] Implementing [18F]fluorocholine PET/CT leads to a reduced effective patient dose compared to standard [99mTc]-sestamibi and [99mTc]-pertechnetate scintigraphy SPECT/CT
    Rode, M.
    Moeller, H. Reilev
    Oestergaard, L. Lange
    Holdgaard, P.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2024, 51 : S311 - S311
  • [6] External radiation exposure of personnel in nuclear medicine from 18F, 99mTc and 131I with special reference to fingers, eyes and thyroid
    Leide-Svegborn, S.
    RADIATION PROTECTION DOSIMETRY, 2012, 149 (02) : 196 - 206
  • [7] Functional evaluation of myocardial viability by 99mTc tetrofosmin gated SPECT -: A quantitative comparison with 18F fluorodeoxyglucose positron emission CT (18F FDG PET)
    Kuwabara, Y
    Watanabe, S
    Nakaya, J
    Fujiwara, M
    Hasegawa, R
    Matsuno, K
    Kuroda, T
    Mikami, Y
    Fujii, K
    Himi, T
    Masuda, Y
    ANNALS OF NUCLEAR MEDICINE, 1999, 13 (03) : 135 - 140
  • [8] Dose Estimation for Extravasation of 177Lu, 99mTc, and 18F
    Tsorxe, Innocent Y.
    Hayes, Robert B.
    HEALTH PHYSICS, 2023, 124 (03): : 217 - 220
  • [9] In vivo simultaneous imaging with 99mTc and 18F using a Compton camera
    Sakai, Makato
    Yamaguchi, Mitsutaka
    Nagao, Yuto
    Kawachi, Naoki
    Kilcuchi, Mikiko
    Torikai, Kota
    Kamiya, Tomihiro
    Takeda, Shinichiro
    Watanabe, Shin
    Takahashi, Tadayuki
    Arakawa, Kazuo
    Nakano, Takashi
    PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (20):
  • [10] Synthesis and evaluation of 99mTc(CO)3(FEDA): a new dual-purpose 99mTc/18F renal imaging agent
    Klenc, Jeffrey
    Taylor, Andrew
    Lipowska, Malgorzata
    JOURNAL OF NUCLEAR MEDICINE, 2015, 56 (03)