Evaluation of neutron dose equivalent from the Mevion S250 proton accelerator: measurements and calculations

被引:17
|
作者
Chen, Kuan Ling [1 ]
Bloch, Charles D. [2 ]
Hill, Patrick M. [3 ]
Klein, Eric E. [1 ]
机构
[1] Washington Univ, Dept Radiat Oncol, St Louis, MO 63130 USA
[2] Seattle Canc Care Alliance Proton Therapy, Seattle, WA USA
[3] Univ Iowa Hosp & Clin, Dept Radiat Oncol, Iowa City, IA 52242 USA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2013年 / 58卷 / 24期
关键词
MONTE-CARLO; SECONDARY NEUTRON; DOUBLE SCATTERING; THERAPY; RADIATION; EXPOSURES; WENDI;
D O I
10.1088/0031-9155/58/24/8709
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Neutron production is of concern for proton therapy, especially for passive scattering proton beam delivery methods. The levels of neutron dose equivalent vary significantly with system design and treatment parameters. The purpose of this study was to examine neutron dose equivalent per therapeutic dose (H/D) around the Mevion S250 proton therapy system, a novel design of proton therapy systems. The benchmark comparisons between measurement and simulation were found to be within a factor of 2 for most cases. The H/D values were evaluated as a function of various parameters. The results showed that, at a standard reference condition (10 x 10 cm(2) field size, distance 1 m detector-to-isocenter lateral to the primary proton beam direction), the H/D values range from 0.72 to 3.37 mSv Gy(-1) for all configurations studied. The H/D values generally (1) decreased as the neutron detectors moved away from the isocenter, (2) decreased with increasing aperture field sizes, (3) increased with increasing angle from the initial beam axis and (4) were independent of treatment nozzle position. The H/D trends were consistent with other existing passive scattering proton accelerators reported in the literature.
引用
收藏
页码:8709 / 8723
页数:15
相关论文
共 50 条
  • [1] Shielding verification and neutron dose evaluation of the Mevion S250 proton therapy unit
    Prusator, Michael T.
    Ahmad, Salahuddin
    Chen, Yong
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2018, 19 (02): : 305 - 310
  • [2] Clinical Commissioning of the Pinnacle Proton Dose Engine for the Mevion S250 Delivery System
    Willoughby, T.
    Zeidan, O.
    [J]. MEDICAL PHYSICS, 2017, 44 (06)
  • [3] Systematic Analysis of the Neutron Equivalent Dose in a Mevion S250 Proton Therapy Machine Using Geant4 Monte Carlo Simulation
    Tatari, N.
    Baradaran-Ghahfarokhi, M.
    Roeder, B.
    Zhao, T.
    Khan, R.
    Sobotka, L.
    Darafsheh, A.
    [J]. MEDICAL PHYSICS, 2021, 48 (06)
  • [4] Routine Quality Assurance for the MEVION S250 Proton System
    Klein, E.
    Zhao, T.
    Grantham, K.
    Goddu, S.
    [J]. MEDICAL PHYSICS, 2013, 40 (06)
  • [5] TOPAS Simulation of the Mevion S250 compact proton therapy unit
    Prusator, Michael
    Ahmad, Salahuddin
    Chen, Yong
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2017, 18 (03): : 88 - 95
  • [6] Evaluation of Surface Dose for Various Treatment Sites Using Mevion S250 Passively Scattered Proton Therapy System
    Jin, H.
    Ahmad, S.
    [J]. MEDICAL PHYSICS, 2017, 44 (06)
  • [7] A Virtual Machine Source Model for Monte Carlo Simulations of the Mevion S250 Proton System
    Yuan, J.
    Jesseph, Federick
    Cheng, Chee-Wai
    Ellis, Rodney
    Machtay, Mitchell
    [J]. MEDICAL PHYSICS, 2018, 45 (06) : E432 - E432
  • [8] Experimental Assessment of Proton Dose Calculation Accuracy in Small-Field Delivery Using a Mevion S250 Proton Therapy System
    DePew, Kyle D.
    Ahmad, Salahuddin
    Jin, Hosang
    [J]. JOURNAL OF MEDICAL PHYSICS, 2018, 43 (04) : 221 - 229
  • [9] Measurements of fetal dose with Mevion S250i proton therapy system with HYPERSCAN
    Hopfensperger, Karolyn M.
    Li, Xing
    Paxton, Adam
    Sarkar, Vikren
    Su, Fan-Chi Frances
    Price, Ryan G.
    Salter, Bill
    St. James, Sara
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2023, 24 (05):
  • [10] Comparing Monte Carlo Simulations with Commissioning Beam Data for Mevion S250 Proton Therapy System
    Prusator, M.
    Jin, H.
    Ahmad, S.
    Chen, Y.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3495 - 3495