Measurements of fetal dose with Mevion S250i proton therapy system with HYPERSCAN

被引:3
|
作者
Hopfensperger, Karolyn M. [1 ,3 ]
Li, Xing [2 ]
Paxton, Adam [2 ]
Sarkar, Vikren [2 ]
Su, Fan-Chi Frances [2 ]
Price, Ryan G. [2 ]
Salter, Bill [2 ]
St. James, Sara [2 ]
机构
[1] Univ Kansas Med Ctr, Dept Radiat Oncol, Kansas City, KS USA
[2] Univ Utah, Huntsman Canc Inst, Dept Radiat Oncol, Salt Lake City, UT USA
[3] Univ KansasMed Ctr, Dept Radiat Oncol, 4001 Rainbow Blvd, RmRO1126, Kansas City, KS 66160 USA
来源
关键词
fetal dose; scanning beam proton therapy; NEUTRON; RADIOTHERAPY; PATIENT;
D O I
10.1002/acm2.13957
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To characterize potential dose to the fetus for all modes of delivery (dynamic adaptive aperture, static adaptive aperture, and no adaptive aperture) for the Mevion S250i Proton Therapy System with HYPERSCAN and compare the findings with those of other available proton systems. Materials and Methods: Fetal dose measurements were performed for all three modes of dose delivery on the Mevion S250i Proton therapy system with HYPERSCAN (static aperture, dynamic aperture and uncollimated). Standard treatment plans were created in RayStation for a left-sided brain lesion treated with a vertex field, a left lateral field, and a posterior field. Measurements were performed using WENDI and the RANDO with the detector placed at representative locations to mimic the growth and movement of the fetus at different gestational stages. Results: The fetal dose measurements varied with fetus position and the largest measured dose was 64.7 mu Sv per 2 Gy (RBE) fraction using the dynamic aperture. The smallest estimated fetal dose was 45.0 mu Sv per 2 Gy (RBE) at the base of the RANDO abdomen (47 cm from isocenter to the outer width of WENDI and 58.5 cm from the center of the WENDI detector) for the static aperture delivery. The vertex fields at all depths had larger contributions to the total dose than the other two and the dynamic aperture plans resulted in the highest dose measured for all depths. Conclusion: The reported doses are lower than reported doses using a double-scattering system. This work suggests that avoiding vertex fields and using the static aperture will help minimize dose to the fetus.
引用
收藏
页数:7
相关论文
共 50 条
  • [1] Estimates of Fetal Dose with the Mevion S250i Proton Therapy System with HYPERSCAN
    Hopfensperger, K.
    Paxton, A.
    Sarkar, V.
    Su, F.
    Price, R.
    Salter, B.
    St James, S.
    [J]. MEDICAL PHYSICS, 2022, 49 (06) : E493 - E493
  • [2] Commissioning of a Mevion S250i with Hyperscan Pencil-Beam Scanning Proton Therapy System
    Jin, H.
    Chen, Y.
    Kendall, E.
    Ahmad, S.
    [J]. MEDICAL PHYSICS, 2019, 46 (06) : E325 - E325
  • [3] Quantitative assessment of the production of radioactive materials by the Mevion S250i Hyperscan proton therapy system: a year-long survey
    Chiang, Bing-Hao
    Chen, Yong
    MacDurmon, George
    Ahmad, Salahuddin
    [J]. JOURNAL OF RADIOTHERAPY IN PRACTICE, 2021, 20 (03) : 361 - 364
  • [4] Developing a Monte Carlo model for MEVION S250i with HYPERSCAN and Adaptive Aperture™ pencil beam scanning proton therapy system
    Chiang, Bing-Hao
    Bunker, Austin
    Jin, Hosang
    Ahmad, Salahuddin
    Chen, Yong
    [J]. JOURNAL OF RADIOTHERAPY IN PRACTICE, 2021, 20 (03) : 279 - 286
  • [5] Comparison of Static and Dynamic Collimation of the Mevion S250i Proton Therapy System
    Jin, H.
    Grewal, H.
    Ahmad, S.
    [J]. MEDICAL PHYSICS, 2020, 47 (06) : E453 - E453
  • [6] A Systematic Log File Analysis Tool for the Mevion S250i Hyperscan System: First Year Experience
    Newpower, M.
    Jin, H.
    Bunker, A.
    Ahmad, S.
    Chen, Y.
    [J]. MEDICAL PHYSICS, 2020, 47 (06) : E691 - E691
  • [7] Determination of Acceptable Air Gap Differences From the Treatment Plan for the Mevion S250i Hyperscan System
    Newpower, M.
    Richards, Z.
    Ahmad, S.
    Jin, H.
    [J]. MEDICAL PHYSICS, 2020, 47 (06) : E739 - E739
  • [8] Patient Specific Quality Assurance Using Machine Log Files for the Mevion S250i Proton Therapy System
    Bunker, A.
    Ahmad, S.
    Chen, Y.
    [J]. MEDICAL PHYSICS, 2019, 46 (06) : E325 - E325
  • [9] Joint EURADOS WG9-WG11 rem-counter intercomparison in a Mevion S250i proton therapy facility with Hyperscan pulsed synchrocyclotron
    Zorloni, Gabriele
    Bosmans, Geert
    Brall, Thomas
    Caresana, Marco
    De Saint-Hubert, Marijke
    Domingo, Carles
    Ferrante, Christian
    Ferrulli, Francesca
    Kopec, Renata
    Leidner, Johannes
    Mares, Vladimir
    Nabha, Racell
    Olko, Pawel
    Caballero-Pacheco, Miguel Angel
    Ruehm, Werner
    Silari, Marco
    Stolarczyk, Liliana
    Vilches-Freixas, Gloria
    Swakon, Jan
    Tisi, Marco
    Trinkl, Sebastian
    Van Hoey, Olivier
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2022, 67 (07):
  • [10] Determination of Virtual Source Points for Mevion S250i Pencil Beam Scanning Proton System
    Chen, Y.
    Frame, C.
    Ma, T.
    Yuan, Y.
    Pang, D.
    [J]. MEDICAL PHYSICS, 2021, 48 (06)