Patient dosimetry and a novel approach to establishing Diagnostic Reference Levels in dental radiology

被引:10
|
作者
Walker, Christopher [1 ]
van der Putten, Wit [1 ,2 ]
机构
[1] St Lukes Hosp, Dept Phys, Dublin 6, Ireland
[2] Natl Univ Ireland, Sch Phys, Galway, Ireland
来源
关键词
Diagnostic Reference Levels; Dental X-ray units; Quality Assurance; Monte-Carlo; Intra-Oral; Pan-oral tomograph;
D O I
10.1016/j.ejmp.2010.12.003
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Diagnostic Reference Levels provide a method of ensuring that patient doses in medical procedures are kept at acceptable levels. Their application in dentistry can provide an indication of current dose levels and can assist in potentially significant dose reduction in Ireland given the number of patients screened annually. This study involved retrospective analyses of entrance surface dose and dose-width-product measurements obtained in Irish Dental Practices for both Intra-Oral and Panoramic units respectively, followed by comparisons with Monte-Carlo generated computer models of these procedures. Analysis was performed on data from 33 Intra-Oral units for an Adult Mandibular Molar entrance surface dose, 198 readings for a proposed mGy/mAs reference level and 50 Panoramic machines for a dose-width product investigation. The third quartile value of the entrance surface dose for a standard Adult Mandibular Molar Intra-Oral radiograph is (2.40 +/- 0.92)mGy, compared to a computer-modelled value of 2.60 mGy. The third quartile mGy/mAs value for Intra-Oral procedures is (1.03 +/- 0.38)mGy/mAs, compared to a computer-modelled value of 0.75 mGy/mAs. The third quartile dose width product for an Adult Panoramic radiograph is (59.89 +/- 20.97)mGymm, compared to a computer-modeled value of 62.40 mGymm. It is proposed to introduce Diagnostic Reference Levels of 2.4 mGy for an Adult Mandibular Molar Intra-Oral radiograph and 60 mGymm for an Adult Panoramic radiograph. The use of a new reference quantity in Intra-Oral radiology is also suggested. This has a value of 1 mGy/mAs and may be introduced alongside established procedures. These levels can be taken as guides to acceptable doses, but it should be noted that further reductions are practical under ALARA principles. (C) 2010 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:7 / 12
页数:6
相关论文
共 50 条
  • [21] Provision of Italian diagnostic reference levels for diagnostic and interventional radiology
    Compagnone, Gaetano
    Padovani, Renato
    D'Ercole, Loredana
    Orlacchio, Antonio
    Bernardi, Guglielmo
    D'Avanzo, Maria Antonietta
    Grande, Sveva
    Palma, Alessandra
    Campanella, Francesco
    Rosi, Antonella
    RADIOLOGIA MEDICA, 2021, 126 (01): : 99 - 105
  • [22] AUTOMATED QUALITY ASSURANCE AND PATIENT DOSIMETRY IN DIAGNOSTIC-RADIOLOGY
    FAULKNER, K
    CHAPPLE, CL
    HEDLEY, P
    KOTRE, CJ
    HARRISON, RM
    JOURNAL OF BIOMEDICAL ENGINEERING, 1990, 12 (03): : 228 - 232
  • [23] ESTABLISHING LOCAL AND NATIONAL DIAGNOSTIC AND INTERVENTIONAL CARDIOLOGY AND RADIOLOGY REFERENCE LEVELS IN A SMALL EUROPEAN STATE: THE CASE OF MALTA
    Pace, Eric
    Cortis, Kelvin
    Debono, Joseph
    Grech, Marvin
    Caruana, Carmel J.
    RADIATION PROTECTION DOSIMETRY, 2020, 191 (03) : 261 - 271
  • [24] Diagnostic reference levels and complexity indices in interventional radiology
    Vossou, Marialena
    Spiliopoulos, Stavros
    Palialexis, Kostas
    Antonakos, John
    Efstathopoulos, Efstathios P.
    RADIATION PROTECTION DOSIMETRY, 2023, 199 (03) : 254 - 261
  • [25] The implementation of diagnostic reference levels to Australian radiology practice
    Wallace, Anthony B.
    JOURNAL OF MEDICAL IMAGING AND RADIATION ONCOLOGY, 2010, 54 (05) : 465 - 471
  • [26] Evolution of diagnostic reference levels in Spanish intraoral radiology
    Alcaraz, M.
    Velasco, F.
    Martinez-Beneyto, Y.
    Alcaraz-Saura, M.
    Velasco, E.
    Achel, G. D.
    Canteras, M.
    RADIATION PROTECTION DOSIMETRY, 2012, 151 (01) : 166 - 171
  • [27] Establishing the European diagnostic reference levels for interventional cardiology
    Siiskonen, T.
    Ciraj-Bjelac, O.
    Dabin, J.
    Diklic, A.
    Domienik-Andrzejewska, J.
    Farah, J.
    Fernandez, J. M.
    Gallagher, A.
    Hourdakis, C. J.
    Jurkovic, S.
    Jarvinen, H.
    Jarvinen, J.
    Knezevic, Z.
    Koukorava, C.
    Maccia, C.
    Majer, M.
    Malchair, F.
    Riccardi, L.
    Rizk, C.
    Sanchez, R.
    Sandborg, M.
    Merce, M. Sans
    Segota, D.
    Sierpowska, J.
    Simantirakis, G.
    Sukupova, L.
    Thrapsanioti, Z.
    Vano, E.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2018, 54 : 42 - 48
  • [28] Establishing diagnostic reference levels for interventional procedures in Kenya
    Korir, Geoffrey K.
    Wambani, Jeska S.
    Yuko-Jowi, Christine A.
    Korir, Ian K.
    Tries, Mark
    Kidali, Mike M.
    RADIOGRAPHY, 2014, 20 (02) : 148 - 152
  • [29] Clinical Diagnostic Reference Levels in Diagnostic Radiology: Preliminary Results in Qatar
    Al Naemi, H.
    Kharita, M.
    Omar, A.
    Al Obadli, A.
    Al Kuwari, M.
    Mohamed, A.
    Alkhazzam, S.
    Aly, A.
    Tsapaki, V.
    MEDICAL PHYSICS, 2020, 47 (06) : E519 - E519
  • [30] Dose area product reference levels in dental panoramic radiology
    Tierris, CE
    Yakoumakis, EN
    Bramis, GN
    Georgiou, E
    RADIATION PROTECTION DOSIMETRY, 2004, 111 (03) : 283 - 287