Monte Carlo dose verification for intensity-modulated arc therapy

被引:31
|
作者
Li, XA [1 ]
Ma, LJ [1 ]
Naqvi, S [1 ]
Shih, RP [1 ]
Yu, C [1 ]
机构
[1] Univ Maryland, Dept Radiat Oncol, Baltimore, MD 21201 USA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2001年 / 46卷 / 09期
关键词
D O I
10.1088/0031-9155/46/9/301
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Intensity-modulated are therapy (IMAT), a technique which combines beam rotation and dynamic multileaf collimation, has been implemented in our clinic. Dosimetric errors can be created by the inability of the planning system to accurately account for the effects of tissue inhomogeneities and physical characteristics of the multileaf collimator (MLC). The objective of this study is to explore the use of Monte Carlo (MC) simulation for IMAT dose verification. The BEAM/DOSXYZ Monte Carlo system was implemented to perform dose verification for the IMAT treatment. The implementation includes the simulation of the linac head/MLC (Elekta SL20), the conversion of patient CT images and beam arrangement for 3D dose calculation, the calculation of gantry rotation and leaf motion by a series of static beams and the development of software to automate the entire MC process. The MC calculations were verified by measurements for conventional beam settings. The agreement was within 2%. The IMAT dose distributions generated by a commercial forward planning system (RenderPlan, Elekta) were compared with those calculated by the MC package. For the cases studied, discrepancies of over 10% were found between the MC and the RenderPlan dose calculations. These discrepancies were due in part to the inaccurate dose calculation of the RenderPlan system. The computation time for the IMAT MC calculation was in the range of 20-80 min on 15 Pentium-III computers. The MC method was also useful in verifying the beam apertures used in the IMAT treatments.
引用
下载
收藏
页码:2269 / 2282
页数:14
相关论文
共 50 条
  • [1] The impact of Monte Carlo dose calculations on intensity-modulated radiation therapy
    Siebers, JV
    Keall, PJ
    Mohan, R
    ADVANCED MONTE CARLO FOR RADIATION PHYSICS, PARTICLE TRANSPORT SIMULATION AND APPLICATIONS, 2001, : 205 - 210
  • [2] Monte Carlo Dose Verification for Volumetric Modulated Arc Therapy in Multiple Brain Metastases
    Tominaga, H.
    Araki, F.
    Kanetake, N.
    Sakata, J.
    MEDICAL PHYSICS, 2016, 43 (06) : 3546 - 3547
  • [3] Validation of a 2D dose verification systemfor intensity-modulated arc therapy
    Zhang, Ruohui
    Bai, Wenwen
    Gao, Yulan
    Wang, Shiguang
    Zhang, Weihui
    Dong, Zhiwei
    RADIATION PHYSICS AND CHEMISTRY, 2023, 213
  • [4] Dose rate optimization for intensity-modulated arc therapy
    Goldman, P.
    Chen, J.
    Wong, E.
    Battista, J.
    MEDICAL PHYSICS, 2006, 33 (06) : 2089 - 2089
  • [5] Fluoroscopic Verification of Intensity-Modulated Rotational Arc Therapy Delivery
    Yi, B.
    Han-Oh, Y.
    Lerma, F.
    Berman, B.
    Yu, C.
    MEDICAL PHYSICS, 2009, 36 (06)
  • [6] Comparison of Intensity-Modulated Radiation Therapy, Intensity-Modulated Arc Therapy and Arc-Modulated Radiation Therapy
    Tang, G.
    Earl, M.
    Luan, S.
    Wang, C.
    Naqvi, S.
    Yu, C.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [7] Commissioning a Fast and Accurate Monte Carlo Dose Engine (MCsquare) in Intensity-Modulated Proton Therapy
    Deng, W.
    Huang, S.
    Souris, K.
    Ding, X.
    Cohilis, M.
    Younkin, J.
    Bues, M.
    Lin, L.
    Shen, J.
    Liu, W.
    MEDICAL PHYSICS, 2019, 46 (06) : E168 - E168
  • [8] Relative profile and dose verification of intensity-modulated radiation therapy
    Chang, JW
    Mageras, GS
    Chui, CS
    Ling, CC
    Lutz, W
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2000, 47 (01): : 231 - 240
  • [9] Evaluation of uncertainty-based stopping criteria for Monte Carlo calculations of intensity-modulated radiotherapy and arc therapy patient dose distributions
    Vanderstraeten, Barbara
    Olteanu, Ana Maria Luiza
    Reynaert, Nick
    Leal, Antonio
    De Neve, Wilfried
    Thierens, Hubert
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 69 (02): : 628 - 637
  • [10] Development and commissioning of a multileaf collimator model in Monte Carlo dose calculations for intensity-modulated radiation therapy
    Jang, SY
    Vassiliev, ON
    Liu, HH
    Mohan, R
    Siebers, JV
    MEDICAL PHYSICS, 2006, 33 (03) : 770 - 781