Defects in Nickel-Titanium Instruments after Clinical Use. Part 2: Fractographic Analysis of Fractured Surface in a Cohort Study

被引:53
|
作者
Shen, Ya [1 ]
Cheung, Gary Shun-pan [2 ]
Peng, Bin [3 ]
Haapasalo, Markus [1 ]
机构
[1] Univ British Columbia, Fac Dent, Div Endodont, Dept Oral Biol & Med Sci, Vancouver, BC V6T 1Z3, Canada
[2] Univ Hong Kong, Fac Dent, Area Endodont, Hong Kong, Hong Kong, Peoples R China
[3] Wuhan Univ, Sch & Hosp Stomatol, Minist Educ, Key Lab Oral Biomed Engn, Wuhan 430072, Peoples R China
关键词
Clinical; failure; fatigue; nickel-titanium; rotary instrument; shear; NITI ROTARY INSTRUMENTS; ROOT-CANAL PREPARATION; LOW-CYCLE FATIGUE; PROTAPER-S1; INSTRUMENTS; ENDODONTIC INSTRUMENTS; BENDING PROPERTIES; PROFILE; FAILURE; FILES; STRESSES;
D O I
10.1016/j.joen.2008.10.013
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
The purpose of this study was to investigate the mode of failure of three brands of nickel-titanium (NiTi) instruments (ProTaper [Dentsply Maillefer, Ballaigues, Switzerland], ProTaper for Hand Use [Dentsply Maillefer], and K3 [SybronEndo, Orange, CA]) that separated during clinical use. A total of 79 fractured instruments were collected from three endodontic clinics over 16 months. The fracture surface of each fragment was examined by scanning electron microscope. Most of the rotary instruments (78% of K3 and 66% of ProTaper) failed because of fatigue fracture, whereas 91% of NiTi hand instruments failed as a result of shear (p < 0.01). The fracture mode of shaping files in rotary ProTaper was different between two different clinics (p < 0.05). All surfaces with fatigue fracture (n = 47) revealed the presence of either one or two crack origins. The vast majority (86%) of K3 fatigue failure had two crack origins that could be found not only at the cutting edge but also at various places along the flute compared with only 28% of ProTaper showing multiple crack origins (p < 0.01). The latter showed one crack origin in 81% of the fatigued shaping files but only 37% for finishing files (p < 0.01). The results of this study indicated that the failure mode of NiTi instruments is related to preparation technique. (J Endod 2009;35:133-136)
引用
收藏
页码:133 / 136
页数:4
相关论文
共 50 条
  • [21] Fractographic analysis of K3 nickel-titanium rotary instruments submitted to different modes of mechanical loading
    Galvao Barbosa, Fabiola Ormiga
    Ponciano Gomes, Jose Antonio da Cunha
    de Araujo, Marcos Cesar Pimenta
    JOURNAL OF ENDODONTICS, 2008, 34 (08) : 994 - 998
  • [22] Why do nickel-titanium archwires fracture intraorally? Fractographic analysis and failure mechanism of in-vivo fractured wires
    Zinelis, Spiros
    Eliades, Theodore
    Pandis, Nikolaos
    Eliades, George
    Bourauel, Christoph
    AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 2007, 132 (01) : 84 - 89
  • [23] Defects in ProTaper S1 instruments after clinical use: fractographic examination
    Cheung, GSP
    Peng, B
    Bian, Z
    Shen, Y
    Darvell, BW
    INTERNATIONAL ENDODONTIC JOURNAL, 2005, 38 (11) : 802 - 809
  • [24] Effect of clinical use on the cyclic fatigue resistance of ProTaper nickel-titanium rotary instruments
    Ounsi, Hani F.
    Salameh, Ziad
    Al-Shalan, Thakib
    Ferrari, Marco
    Grandini, Simone
    Pashley, David H.
    Tay, Franklin R.
    JOURNAL OF ENDODONTICS, 2007, 33 (06) : 737 - 741
  • [25] Influence of multiple clinical use on fatigue resistance of ProTaper rotary nickel-titanium instruments
    Vieira, E. P.
    Franca, E. C.
    Martins, R. C.
    Buono, V. T. L.
    Bahia, M. G. A.
    INTERNATIONAL ENDODONTIC JOURNAL, 2008, 41 (02) : 163 - 172
  • [26] Influence of simulated clinical use on the torsional behavior of nickel-titanium rotary endodontic instruments
    Bahia, MGA
    Melo, MCC
    Buono, VTL
    ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTOLOGY, 2006, 101 (05): : 675 - 680
  • [27] Evaluation of deformation and fracture rates for nickel-titanium rotary instruments according to the frequency of clinical use
    Yilmaz, Ayca
    Gokyay, Sitki Selcuk
    Daglaroglu, Rustu
    Kucukay, Isil Karagoz
    EUROPEAN ORAL RESEARCH, 2018, 52 (02) : 89 - 93
  • [28] Evaluation of an ultrasonic technique to remove fractured rotary nickel-titanium endodontic instruments from root canals: Clinical cases
    Ward, JR
    Parashos, P
    Messer, HH
    JOURNAL OF ENDODONTICS, 2003, 29 (11) : 764 - 767
  • [29] Mechanical behaviour of nickel-titanium rotary endodontic instruments in simulated clinical conditions: a computational study
    Necchi, S.
    Taschieri, S.
    Petrini, L.
    Migliavacca, F.
    INTERNATIONAL ENDODONTIC JOURNAL, 2008, 41 (11) : 939 - 949
  • [30] Cyclic fatigue of nickel-titanium rotary instruments after clinical use with low- and high-torque endodontic motors
    Gambarini, G