Defects in Nickel-Titanium Instruments after Clinical Use. Part 2: Fractographic Analysis of Fractured Surface in a Cohort Study

被引:53
|
作者
Shen, Ya [1 ]
Cheung, Gary Shun-pan [2 ]
Peng, Bin [3 ]
Haapasalo, Markus [1 ]
机构
[1] Univ British Columbia, Fac Dent, Div Endodont, Dept Oral Biol & Med Sci, Vancouver, BC V6T 1Z3, Canada
[2] Univ Hong Kong, Fac Dent, Area Endodont, Hong Kong, Hong Kong, Peoples R China
[3] Wuhan Univ, Sch & Hosp Stomatol, Minist Educ, Key Lab Oral Biomed Engn, Wuhan 430072, Peoples R China
关键词
Clinical; failure; fatigue; nickel-titanium; rotary instrument; shear; NITI ROTARY INSTRUMENTS; ROOT-CANAL PREPARATION; LOW-CYCLE FATIGUE; PROTAPER-S1; INSTRUMENTS; ENDODONTIC INSTRUMENTS; BENDING PROPERTIES; PROFILE; FAILURE; FILES; STRESSES;
D O I
10.1016/j.joen.2008.10.013
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
The purpose of this study was to investigate the mode of failure of three brands of nickel-titanium (NiTi) instruments (ProTaper [Dentsply Maillefer, Ballaigues, Switzerland], ProTaper for Hand Use [Dentsply Maillefer], and K3 [SybronEndo, Orange, CA]) that separated during clinical use. A total of 79 fractured instruments were collected from three endodontic clinics over 16 months. The fracture surface of each fragment was examined by scanning electron microscope. Most of the rotary instruments (78% of K3 and 66% of ProTaper) failed because of fatigue fracture, whereas 91% of NiTi hand instruments failed as a result of shear (p < 0.01). The fracture mode of shaping files in rotary ProTaper was different between two different clinics (p < 0.05). All surfaces with fatigue fracture (n = 47) revealed the presence of either one or two crack origins. The vast majority (86%) of K3 fatigue failure had two crack origins that could be found not only at the cutting edge but also at various places along the flute compared with only 28% of ProTaper showing multiple crack origins (p < 0.01). The latter showed one crack origin in 81% of the fatigued shaping files but only 37% for finishing files (p < 0.01). The results of this study indicated that the failure mode of NiTi instruments is related to preparation technique. (J Endod 2009;35:133-136)
引用
收藏
页码:133 / 136
页数:4
相关论文
共 50 条
  • [1] Defects in Nickel-Titanium Instruments after Clinical Use. Part 4: An Electropolished Instrument
    Shen, Ya
    Winestock, Earl
    Cheung, Gary Shun-Pan
    Haapasalo, Markus
    JOURNAL OF ENDODONTICS, 2009, 35 (02) : 197 - 201
  • [2] Defects in Nickel-Titanium Instruments after Clinical Use. Part 1: Relationship between Observed Imperfections and Factors Leading to Such Defects in a Cohort Study
    Shen, Ya
    Haapasalo, Markus
    Cheung, Gary Shun-pan
    Peng, Bin
    JOURNAL OF ENDODONTICS, 2009, 35 (01) : 129 - 132
  • [3] Defects in Nickel-Titanium Instruments after Clinical Use. Part 5: Single Use From Endodontic Specialty Practices
    Shen, Ya
    Coil, Jeffrey M.
    Mclean, Alex G. R.
    Hemerling, David L.
    Haapasalo, Markus
    JOURNAL OF ENDODONTICS, 2009, 35 (10) : 1363 - 1367
  • [4] Defects in Nickel-Titanium Instruments after Clinical Use. Part 3: A 4-Year Retrospective Study from an Undergraduate Clinic
    Shen, Ya
    Coil, Jeffrey M.
    Haapasalo, Markus
    JOURNAL OF ENDODONTICS, 2009, 35 (02) : 193 - 196
  • [5] Factors influencina defects of rotary nickel-titanium endodontic instruments after clinical use
    Parashos, P
    Gordon, I
    Messer, HH
    JOURNAL OF ENDODONTICS, 2004, 30 (10) : 722 - 725
  • [6] Defects in rotary nickel-titanium files after clinical use
    Sattapan, B
    Nervo, GJ
    Palamara, JEA
    Messer, HH
    JOURNAL OF ENDODONTICS, 2000, 26 (03) : 161 - 165
  • [7] SEM observations of nickel-titanium rotary endodontic instruments that fractured during clinical use
    Alapati, SB
    Brantley, WA
    Svec, TA
    Powers, JM
    Nusstein, JM
    Daehn, GS
    JOURNAL OF ENDODONTICS, 2005, 31 (01) : 40 - 43
  • [8] Ultrastructural Analysis of the Surface Changes on the Nickel-Titanium Endodontic Instruments after Multiple Use
    Stosic, Nenad
    Popovic, Jelena
    Apostolovic, Marija Andelkovic
    Mitic, Aleksandar
    Nikolic, Marija
    Barac, Radomir
    ACTA FACULTATIS MEDICAE NAISSENSIS, 2021, 38 (03) : 291 - 298
  • [9] Defects in ProFile rotary nickel-titanium files after clinical use
    Ferreira, Manuel
    Ferreira, Hugo
    Oliveiros, Barbara
    Carrilho, Eunice
    ENDO-ENDODONTIC PRACTICE TODAY, 2012, 6 (02): : 113 - 117
  • [10] Modes of failure of ProTaper nickel-titanium rotary instruments after clinical use
    Wei, Xi
    Ling, Junqi
    Jiang, Jin
    Huang, Xiangya
    Liu, Lu
    JOURNAL OF ENDODONTICS, 2007, 33 (03) : 276 - 279