Interrupting gel layer between Double cuffs prevents fluid leakage past tracheal tube cuffs

被引:9
|
作者
Hwang, J. Y. [1 ]
Han, S. H. [2 ]
Park, S. H. [2 ]
Park, S. J. [2 ]
Park, S. [2 ]
Oh, S. H. [3 ]
Kim, J. H. [2 ]
机构
[1] SMG SNU Boramae Med Ctr, Dept Anesthesiol & Pain Med, Seoul, South Korea
[2] Seoul Natl Univ, Bundang Hosp, Dept Anesthesiol & Pain Med, Gyeonggido, South Korea
[3] SMG SNU Boramae Med Ctr, Dept Biostat, Seoul, South Korea
关键词
cuffs tracheal; complications; equipment; respiratory aspiration; VENTILATOR-ASSOCIATED PNEUMONIA; LOW-PRESSURE CUFF; ENDOTRACHEAL-TUBES; HIGH-VOLUME; ASPIRATION; LUBRICATION; SHAPE;
D O I
10.1093/bja/aet152
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Current tracheal tubes (TTs) cannot guarantee a perfect seal against pulmonary aspiration of upper airway secretions. The purpose of this study was to investigate whether a gel layer between the tracheal tube with double cuffs (Double cuffs) prevents fluid leakage past TT cuffs. In vitro fluid leakage tests were performed using the Double cuff with or without a gel layer between the two cuffs and four commercially available TTs (Euromedical Standard TT, Mallinckrodt Hi-Lo, Microcuff, and Mallinckrodt TaperGuard) when placed in artificial tracheas with three-different internal diameters (ID; 16, 20, and 22 mm). Blue-dyed water (5 ml) was placed above the cuff, and the extent of fluid leakage was observed for 48 h. Each test was repeated five times with new tubes at six different intracuff pressures (15, 20, 25, 30, 40, and 50 cm H2O). In all of the conventional TTs and the Double cuff without a gel layer, fluid leakage was observed even at clinically acceptable intracuff pressures (2530 cm H2O). However, in the Double cuff with a gel layer, no fluid leakage was observed for 48 h at all intracuff pressures in three-different sized artificial tracheas. At an intracuff pressure of 20 cm H2O in a 20 mm ID trachea, the average volume of injected gel was 2.0 ml. After removal of the TT, the mean volume of the remaining gel in the trachea was 0.10 ml. A prototype TT with gel-layered Double cuffs completely blocked fluid leakage past the cuffs for 48 h in a bench-top model. Clinical studies are required to determine whether this TT reduces the risk of ventilator-associated pneumonia.
引用
收藏
页码:496 / 504
页数:9
相关论文
共 23 条
  • [21] FLUID LEAKAGE PAST TRACHEAL TUBE CUFF: EFFECT OF SUCTIONING MANOEUVRE. A CLINICAL STUDY
    Beuret, P.
    Philippon, B.
    Minacori, F.
    Fabre, X.
    Kaaki, M.
    Carton, M-J
    INTENSIVE CARE MEDICINE, 2011, 37 : S79 - S79
  • [22] A double-layer tracheal tube cuff designed to prevent leakage: a bench-top study
    Alberto Zanella
    Massimo Cressoni
    Myra Epp
    Mario Stylianou
    Theodor Kolobow
    Intensive Care Medicine, 2008, 34 : 1145 - 1149
  • [23] A double-layer tracheal tube cuff designed to prevent leakage: a bench-top study
    Zanella, Alberto
    Cressoni, Massimo
    Epp, Myra
    Stylianou, Mario
    Kolobow, Theodor
    INTENSIVE CARE MEDICINE, 2008, 34 (06) : 1145 - 1149