Investigation of polycapillary half lenses for quantitative confocal micro-X-ray fluorescence analysis

被引:6
|
作者
Iro, Michael [1 ]
Ingerle, Dieter [1 ]
Radtke, Martin [2 ]
Buzanich, Ana Guilherme [2 ]
Kregsamer, Peter [1 ]
Streli, Christina [1 ]
机构
[1] TU Wien, Atominst, Stadionallee 2, A-1020 Vienna, Austria
[2] Bundesanstalt Mat Forsch & Prufung BAM, Richard Willstatter Str 11, D-12489 Berlin, Germany
关键词
polycapillary optics; confocal micro-XRF; X-ray fluorescence; 3D elemental analysis; X-ray optics;
D O I
10.1107/S1600577522009699
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The use of polycapillary optics in confocal micro-X-ray fluorescence analysis (CMXRF) enables the destruction-free 3D investigation of the elemental composition of samples. The energy-dependent transmission properties, concerning intensity and spatial beam propagation of three polycapillary half lenses, which are vital for the quantitative interpretation of such CMXRF measurements, are investigated in a monochromatic confocal laboratory setup at the Atominstitut of TU Wien, and a synchrotron setup on the BAMline beamline at the BESSY II Synchrotron, Helmholtz-Zentrum-Berlin. The empirically established results, concerning the intensity of the transmitted beam, are compared with theoretical values calculated with the polycap software package and a newly presented analytical model for the transmission function. The resulting form of the newly modelled energy-dependent transmission function is shown to be in good agreement with Monte Carlo simulated results for the complete energy regime, as well as the empirically established results for the energy regime between 6 keV and 20 keV. An analysis of possible fabrication errors was conducted via pinhole scans showing only minor fabrication errors in two of the investigated polycapillary optics. The energy-dependent focal spot size of the primary polycapillary was investigated in the laboratory via the channel-wise evaluation of knife-edge scans. Experimental results are compared with data given by the manufacturer as well as geometric estimations for the minimal focal spot size. Again, the resulting measurement points show a trend in agreement with geometrically estimated results and manufacturer data.
引用
收藏
页码:1376 / 1384
页数:9
相关论文
共 50 条
  • [41] Characterization of a Fluidized Catalytic Cracking Catalyst on Ensemble and Individual Particle Level by X-ray Micro-and Nanotomography, Micro-X-ray Fluorescence, and Micro-X-ray Diffraction
    Bare, Simon R.
    Charochak, Meghan E.
    Kelly, Shelly D.
    Lai, Barry
    Wang, Jun
    Chen-Wiegart, Yu-chen Karen
    CHEMCATCHEM, 2014, 6 (05) : 1427 - 1437
  • [42] A beam path-based method for attenuation correction of confocal micro-X-ray fluorescence imaging data
    Liu, Peng
    Ptacek, Carol J.
    Blowes, David W.
    Finfrock, Y. Zou
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2017, 32 (08) : 1582 - 1589
  • [43] Study on Zn relative concentration and chemical state in broilers duodenum by micro-X-ray fluorescence and micro-X-ray absorption fine structure
    Liu, Hua-Wei
    Liu, Da-Sen
    Zheng, Li-Xin
    LIVESTOCK SCIENCE, 2014, 161 : 101 - 108
  • [44] Visualization of Urea Treatments Using Micro-X-Ray Fluorescence Spectrometry
    Brehm, Lora
    Keene, Ellen C.
    Ng, Sze-Sze
    PESTICIDE FORMULATION AND DELIVERY SYSTEMS: 37TH VOLUME, FORMULATIONS WITH INGREDIENTS ON THE EPA'S LIST OF MINIMAL CONCERN, 2018, 1602 : 120 - 135
  • [45] Micro-X-ray Fluorescence Analysis of Sr in Fish Otolith of Ayu Plecoglossus altivelis
    Nakazawa, Takashi
    Nakano, Kazuhiko
    Onchi, Hiromitsu
    Namita, Mayu
    Yamchi, Susumu
    Tsuji, Kouichi
    BUNSEKI KAGAKU, 2012, 61 (07) : 637 - 642
  • [46] Confocal three-dimensional micro-X-ray fluorescence spectroscopy used to retrieve the information covered in printed paper
    Sun, Xuepeng
    Zhang, Xiaoyun
    Wang, Xinyun
    Wang, Yabing
    Li, Yufei
    Peng, Shiqi
    Shao, Shangkun
    Liu, Zhiguo
    Shang, Hongzhong
    Sun, Tianxi
    X-RAY SPECTROMETRY, 2020, 49 (02) : 267 - 273
  • [47] Quantification of large scale micro-X-ray fluorescence elemental images
    Worley, CG
    Havrilla, GJ
    Dunn, PS
    APPLIED SPECTROSCOPY, 2001, 55 (11) : 1448 - 1454
  • [48] A micro X-ray fluorescence analysis method using polycapillary X-ray optics and grazing exit geometry
    Yang, Jun
    Tsuji, Kouichi
    Lin, Xiaoyan
    Han, Dongyan
    Ding, Xunliang
    THIN SOLID FILMS, 2009, 517 (11) : 3357 - 3361
  • [49] Limitation of micro-X-ray fluorescence for chemical quantification in nanocrystalline films
    Phan, Trang Thuy Thi
    Tran, Vinh Cao
    Pham, Anh Tuan Thanh
    SURFACES AND INTERFACES, 2024, 48
  • [50] A Method for Redox Mapping by Confocal Micro-X-ray Fluorescence Imaging: Using Chromium Species in a Biochar Particle as an Example
    Liu, Peng
    Ptacek, Carol J.
    Blowes, David W.
    Finfrock, Y. Zou
    Steinepreis, Mark
    Budimir, Filip
    ANALYTICAL CHEMISTRY, 2019, 91 (08) : 5142 - 5149