Attenuation-based Automatic Tube Voltage Selection and Tube Current Modulation for Dose Reduction at Contrast-enhanced Liver CT

被引:78
|
作者
Lee, Kyung Hee [1 ]
Lee, Jeong Min [1 ]
Moon, Sung Kyoung [1 ]
Baek, Jee Hyun [1 ]
Park, Ji Hoon [1 ]
Flohr, Thomas G. [2 ]
Kim, Kyung Won [1 ]
Kim, Soo Jin [1 ]
Han, Joon Koo [1 ]
Choi, Byung Ihn [1 ]
机构
[1] Seoul Natl Univ, Coll Med, Inst Radiat Med, Dept Radiol,Div Abdominal Imaging, Seoul 110744, South Korea
[2] Siemens Med Solut, Forchheim, Germany
关键词
STATISTICAL ITERATIVE RECONSTRUCTION; INITIAL CLINICAL-EXPERIENCE; IMPROVED IMAGE QUALITY; MULTIDETECTOR ROW CT; ABDOMINAL CT; COMPUTED-TOMOGRAPHY; RADIATION-EXPOSURE; PATIENT SIZE; OPTIMIZATION; FEASIBILITY;
D O I
10.1148/radiol.12112434
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To retrospectively determine whether the combined use of automatic tube voltage selection (ATVS) and automatic tube current modulation (ATCM) can effectively reduce radiation dose at contrast material-enhanced liver computed tomography (CT) while maintaining acceptable image quality compared with the use of ATCM alone. Materials and Methods: This study was approved by an institutional review board, and informed consent was waived. Three hundred fourteen consecutive patients suspected of having liver disease were divided into three groups. In two groups, both ATVS and ATCM were used (group A1, n = 97; group A2, n = 101) but with different contrast gain settings; in one group, only ATCM with a fixed tube potential of 120 kV (group B, n = 116) was used. Weighted volume CT dose index and dose-length product, contrast-to-noise ratios (CNRs), and mean image noise were assessed. Qualitative analysis was performed by two board-certified radiologists and one radiology resident. Statistical analysis was performed by using the one-way analysis of variance test, two-tailed paired t test, Kruskal-Wallis test, and noninferiority test. Results: In groups A1 and A2, a significant dose reduction was obtained compared with that in group B (P < .0001). The mean dose reduction was 20% in group A1 and 31% in group A2. Furthermore, CNRs were significantly higher in groups A1 and A2 than in group B (P < .0001). Despite the higher image noise in groups A1 and A2, the overall image quality was acceptable. Conclusion: Compared with the use of ATCM alone, the combined use of ATVS and ATCM allowed reduction of radiation exposure while maintaining good image quality at contrast-enhanced liver CT.
引用
收藏
页码:437 / 447
页数:11
相关论文
共 50 条
  • [1] Dose reduction in multidetector CT using attenuation-based online tube current modulation
    Tack, D
    De Maertelaer, V
    Gevenois, PA
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2003, 181 (02) : 331 - 334
  • [2] Evaluation of automated attenuation-based tube potential selection in combination with organ-specific dose reduction for contrast-enhanced chest CT examinations
    Schimmoeller, L.
    Lanzman, R. S.
    Dietrich, S.
    Boos, J.
    Heusch, P.
    Miese, F.
    Antoch, G.
    Kroepil, P.
    CLINICAL RADIOLOGY, 2014, 69 (07) : 721 - 726
  • [3] Dose reduction in CT examination of children by an attenuation-based on-line modulation of tube current (CARE Dose)
    Holger Greess
    Anton Nömayr
    Heiko Wolf
    Ulrich Baum
    Michael Lell
    Bernhard Böwing
    Willi Kalender
    Werner A. Bautz
    European Radiology, 2002, 12 : 1571 - 1576
  • [4] Dose reduction in CT examination of children by an attenuation-based on-line modulation of tube current (CARE Dose)
    Greess, H
    Nömayr, A
    Wolf, H
    Baum, U
    Lell, M
    Böwing, B
    Kalender, W
    Bautz, WA
    EUROPEAN RADIOLOGY, 2002, 12 (06) : 1571 - 1576
  • [5] Automatic Selection of Tube Potential for Radiation Dose Reduction in Vascular and Contrast-Enhanced Abdominopelvic CT
    Yu, Lifeng
    Fletcher, Joel G.
    Grant, Katharine L.
    Carter, Rickey E.
    Hough, David M.
    Barlow, John M.
    Vrtiska, Terri J.
    Williamson, Eric E.
    Young, Phillip M.
    Goss, Brian C.
    Shiung, Maria
    Leng, Shuai
    Raupach, Ranier
    Schmidt, Bernhard
    Flohr, Thomas
    McCollough, Cynthia H.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2013, 201 (02) : W297 - W306
  • [6] Dose Reduction Strategy Using Attenuation-Based Tube Current Modulation Method in CBCT for IGRT
    Son, K.
    Lee, H.
    Kim, C.
    Kim, J.
    Cho, S.
    MEDICAL PHYSICS, 2016, 43 (06) : 3414 - 3415
  • [7] Optimal dose reduction algorithm using an attenuation-based tube current modulation method for cone-beam CT imaging
    Son, Kihong
    Chang, Jieun
    Lee, Hoyeon
    Kim, Changhwan
    Lee, Taewon
    Cho, Seungryong
    Park, Sohyun
    Kim, Jin Sung
    PLOS ONE, 2018, 13 (02):
  • [8] Effect of Automated Attenuation-based Tube Voltage Selection on Radiation Dose at CT: An Observational Study on a Global Scale
    Spearman, James V.
    Schoepf, U. Joseph
    Rottenkolber, Marietta
    Driesser, Ivo
    Canstein, Christian
    Thierfelder, Kolja M.
    Krazinski, Aleksander W.
    De Cecco, Carlo N.
    Meinel, Felix G.
    RADIOLOGY, 2016, 279 (01) : 167 - 174
  • [9] Automated attenuation-based selection of tube voltage and tube current for coronary CT angiography: Reduction of radiation exposure versus a BMI-based strategy with an expert investigator
    Layritz, Christian
    Muschiol, Gerd
    Flohr, Thomas
    Bietau, Christian
    Marwan, Mohamed
    Schuhbaeck, Annika
    Schmid, Jasmin
    Ropers, Dieter
    Achenbach, Stephan
    Pflederer, Tobias
    JOURNAL OF CARDIOVASCULAR COMPUTED TOMOGRAPHY, 2013, 7 (05) : 303 - 310
  • [10] Potential for Radiation Dose Savings in Abdominal and Chest CT Using Automatic Tube Voltage Selection in Combination With Automatic Tube Current Modulation
    Mayer, Caroline
    Meyer, Mathias
    Fink, Christian
    Schmidt, Bernhard
    Sedlmair, Martin
    Schoenberg, Stefan O.
    Henzler, Thomas
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2014, 203 (02) : 292 - 299