Attenuation-based Automatic Tube Voltage Selection and Tube Current Modulation for Dose Reduction at Contrast-enhanced Liver CT

被引:78
|
作者
Lee, Kyung Hee [1 ]
Lee, Jeong Min [1 ]
Moon, Sung Kyoung [1 ]
Baek, Jee Hyun [1 ]
Park, Ji Hoon [1 ]
Flohr, Thomas G. [2 ]
Kim, Kyung Won [1 ]
Kim, Soo Jin [1 ]
Han, Joon Koo [1 ]
Choi, Byung Ihn [1 ]
机构
[1] Seoul Natl Univ, Coll Med, Inst Radiat Med, Dept Radiol,Div Abdominal Imaging, Seoul 110744, South Korea
[2] Siemens Med Solut, Forchheim, Germany
关键词
STATISTICAL ITERATIVE RECONSTRUCTION; INITIAL CLINICAL-EXPERIENCE; IMPROVED IMAGE QUALITY; MULTIDETECTOR ROW CT; ABDOMINAL CT; COMPUTED-TOMOGRAPHY; RADIATION-EXPOSURE; PATIENT SIZE; OPTIMIZATION; FEASIBILITY;
D O I
10.1148/radiol.12112434
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To retrospectively determine whether the combined use of automatic tube voltage selection (ATVS) and automatic tube current modulation (ATCM) can effectively reduce radiation dose at contrast material-enhanced liver computed tomography (CT) while maintaining acceptable image quality compared with the use of ATCM alone. Materials and Methods: This study was approved by an institutional review board, and informed consent was waived. Three hundred fourteen consecutive patients suspected of having liver disease were divided into three groups. In two groups, both ATVS and ATCM were used (group A1, n = 97; group A2, n = 101) but with different contrast gain settings; in one group, only ATCM with a fixed tube potential of 120 kV (group B, n = 116) was used. Weighted volume CT dose index and dose-length product, contrast-to-noise ratios (CNRs), and mean image noise were assessed. Qualitative analysis was performed by two board-certified radiologists and one radiology resident. Statistical analysis was performed by using the one-way analysis of variance test, two-tailed paired t test, Kruskal-Wallis test, and noninferiority test. Results: In groups A1 and A2, a significant dose reduction was obtained compared with that in group B (P < .0001). The mean dose reduction was 20% in group A1 and 31% in group A2. Furthermore, CNRs were significantly higher in groups A1 and A2 than in group B (P < .0001). Despite the higher image noise in groups A1 and A2, the overall image quality was acceptable. Conclusion: Compared with the use of ATCM alone, the combined use of ATVS and ATCM allowed reduction of radiation exposure while maintaining good image quality at contrast-enhanced liver CT.
引用
收藏
页码:437 / 447
页数:11
相关论文
共 50 条
  • [21] Attenuation-based size metric for estimating organ dose to patients undergoing tube current modulated CT exams
    Bostani, Maryam
    McMillan, Kyle
    Lu, Peiyun
    Kim, Hyun J.
    Cagnon, Chris H.
    DeMarco, John J.
    McNitt-Gray, Michael F.
    MEDICAL PHYSICS, 2015, 42 (02) : 958 - 968
  • [22] Generating Attenuation-Based Tube Current Modulation (TCM) Schema Based On Voxelized Patient Models to Be Used in Simulations Estimating Organ Dose From CT
    Angel, E.
    DeMarco, J.
    Cagnon, C.
    McNitt-Gray, M.
    MEDICAL PHYSICS, 2009, 36 (06)
  • [23] Automatic Tube Current Modulation and Tube Voltage Selection in Pediatric Computed Tomography A Phantom Study on Radiation Dose and Image Quality
    Papadalds, Antonios E.
    Damilakis, John
    INVESTIGATIVE RADIOLOGY, 2019, 54 (05) : 265 - 272
  • [24] Influence of CT automatic tube current modulation on uncertainty in effective dose
    Sookpeng, S.
    Martin, C. J.
    Gentle, D. J.
    RADIATION PROTECTION DOSIMETRY, 2016, 168 (01) : 46 - 54
  • [25] Bismuth Shielding, Organ-based Tube Current Modulation, and Global Reduction of Tube Current for Dose Reduction to the Eye at Head CT
    Wang, Jia
    Duan, Xinhui
    Christner, Jodie A.
    Leng, Shuai
    Grant, Katharine L.
    McCollough, Cynthia H.
    RADIOLOGY, 2012, 262 (01) : 191 - 198
  • [26] Reduced radiation dose and improved image quality at cardiovascular CT angiography by automated attenuation-based tube voltage selection: intra-individual comparison
    Krazinski, Aleksander W.
    Meinel, Felix G.
    Schoepf, U. Joseph
    Silverman, Justin R.
    Canstein, Christian
    De Cecco, Carlo N.
    Geyer, Lucas L.
    EUROPEAN RADIOLOGY, 2014, 24 (11) : 2677 - 2684
  • [27] Reduced radiation dose and improved image quality at cardiovascular CT angiography by automated attenuation-based tube voltage selection: intra-individual comparison
    Aleksander W. Krazinski
    Felix G. Meinel
    U. Joseph Schoepf
    Justin R. Silverman
    Christian Canstein
    Carlo N. De Cecco
    Lucas L. Geyer
    European Radiology, 2014, 24 : 2677 - 2684
  • [28] Pediatric cardiovascular CT angiography: Radiation dose reduction using automatic anatomic tube current modulation
    Herzog, Christopher
    Mulvihill, Denise M.
    Nguyen, Shaun A.
    Savino, Giancarlo
    Schmidt, Bernhard
    Costello, Philip
    Vogl, Thomas J.
    Schoepf, U. Joseph
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2008, 190 (05) : 1232 - 1240
  • [29] Reducing CT Dose During Routine Brain CT Using Attenuation Based Tube Current Modulation (TCM)
    Angel, E.
    Zhang, D.
    MEDICAL PHYSICS, 2012, 39 (06) : 3925 - 3925
  • [30] CT angiography of the neck: Value of contrast medium dose reduction with low tube voltage and high tube current in a 64-detector row CT
    Xia, W.
    Wu, J. T.
    Yin, X. R.
    Wang, Z. J.
    Wu, H. T.
    CLINICAL RADIOLOGY, 2014, 69 (04) : E183 - E189