Petal resonator surface coil with a circular envelope for high field magnetic resonance imaging

被引:0
|
作者
Hidalgo, S. S. [1 ]
Cuellar, G. [1 ]
Solis, S. E. [1 ]
Alejski, A.
Rodriguez, A. O. [1 ]
机构
[1] Univ Autonoma Metropolitana Iztapalapa, Ctr Invest & Instrumentac & Imagenol Med, Mexico City 09340, DF, Mexico
来源
MEDICAL PHYSICS | 2008年 / 1032卷
关键词
PERES coil; magnetic resonance imaging; electromagnetic simulation; high field;
D O I
暂无
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
A modified version of the petal resonator surface (PERES) coil is proposed in which the petal coils are located inside a circular-shaped coil as opposed to the previous PERES version which did not have a circular envelope. An 8-petal coil was simulated using a finite element method and the quasi-static approach to numerically investigate its electromagnetic characteristics. Numerical simulations showed that the field uniformity is not greatly affected by the use of circular-shaped coil inside in the circular envelope. A coil prototype was developed using the same coil configuration as above and tested on a commercial 3 T imager and a General Electric phantom. Phantom images also demonstrated that the mutual inductance between petals does not alter the coil performance. Standard pulse sequences were used to acquire phantom images with the petal resonator surface coil. It was shown that this prototype resonator coil is fully compatible with clinical high field MR imagers and clinical pulse sequences.
引用
收藏
页码:225 / 228
页数:4
相关论文
共 50 条
  • [41] Coupled microstrip line transverse electromagnetic resonator model for high-field magnetic resonance imaging
    Bogdanov, G
    Ludwig, R
    MAGNETIC RESONANCE IN MEDICINE, 2002, 47 (03) : 579 - 593
  • [42] High-Tc superconducting surface coil for 2 Tesla magnetic resonance imaging of small animals
    Wosik, J
    Wang, F
    Xie, LM
    Strikovski, M
    Kamel, M
    Nesteruk, K
    Bilgen, M
    Narayana, PA
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2001, 11 (01) : 681 - 684
  • [43] The magnetic field of a circular cylindrical coil.
    Farr, CC
    PHILOSOPHICAL MAGAZINE, 1931, 12 (77): : 603 - 603
  • [44] A micro planar coil for local high resolution Magnetic Resonance Imaging
    Takahashi, Hidetoshi
    Dohi, Tetsuji
    Matsumoto, Kiyoshi
    Shimoyama, Isao
    PROCEEDINGS OF THE IEEE TWENTIETH ANNUAL INTERNATIONAL CONFERENCE ON MICRO ELECTRO MECHANICAL SYSTEMS, VOLS 1 AND 2, 2007, : 786 - 789
  • [45] High-resolution ultra-low field magnetic resonance imaging with a high-sensitivity sensing coil
    Savukov, Igor
    Kim, Young Jin
    Newman, Shaun
    JOURNAL OF APPLIED PHYSICS, 2022, 132 (17)
  • [46] High Field Cardiac Magnetic Resonance Imaging A Case for Ultrahigh Field Cardiac Magnetic Resonance
    Niendorf, Thoralf
    Schulz-Menger, Jeanette
    Paul, Katharina
    Huelnhagen, Till
    Ferrari, Victor A.
    Hodge, Russell
    CIRCULATION-CARDIOVASCULAR IMAGING, 2017, 10 (06)
  • [47] A Proposal of Gradient Coil Configuration for Low-Field Magnetic Resonance Imaging
    Fedotov, Alexander
    Pugovkin, Vladimir
    Burov, Denis
    Hurshkainen, Anna
    Mirabal, Carlos Cabal
    APPLIED MAGNETIC RESONANCE, 2024, 55 (08) : 767 - 774
  • [48] Near-field imaging of magnetic resonance in terahertz dielectric resonator antennas
    Lee, Wendy S. L.
    Kaltenecker, Korbinian
    Nirantar, Shruti
    Withayachumnankul, Withawat
    Walther, Markus
    Bhaskaran, Madhu
    Fischer, Bernd M.
    Sriram, Sharath
    Fumeaux, Christophe
    2016 41ST INTERNATIONAL CONFERENCE ON INFRARED, MILLIMETER, AND TERAHERTZ WAVES (IRMMW-THZ), 2016,
  • [49] 6-slot surface coil for magnetic resonance imaging at 4 Tesla
    Solis, S. E.
    Tomasi, D.
    Wang, R. L.
    Rodriguez, A. O.
    MEDICAL PHYSICS, 2006, 854 : 75 - +
  • [50] Magnetic Resonance Imaging of Atherosclerotic Lesion with New Devised Animal Surface Coil
    赵炳辉
    李明华
    赵清
    程英升
    肖云峰
    赵家民
    Journal of Shanghai Jiaotong University(Science), 2008, 13 (06) : 748 - 752