Wear at the taper-trunnion junction of contemporary ceramic-on-ceramic hips shown in a multistation hip simulator

被引:13
|
作者
Bhalekar, Rohan M. [1 ]
Smith, Simon L. [1 ]
Joyce, Thomas J. [1 ]
机构
[1] Newcastle Univ, Sch Engn, Newcastle Upon Tyne NE1 7RU, Tyne & Wear, England
关键词
total hip replacement; ceramic-on-ceramic; hip simulator; taper-trunnion junction; wear; TITANIUM-ALLOY; IN-VITRO; METAL; FAILURE; HEAD; PROSTHESES; UHMWPE; DEBRIS; MICROSEPARATION; BEARINGS;
D O I
10.1002/jbm.b.34213
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Ceramic-on-ceramic (CoC) total hip replacements (THR) have a substantially lower wear rate than metal-on-polyethylene (MoP) hips, as shown by hip simulator testing. However, the revision rates of CoC and MoP hips are comparable. To try and explain this discrepancy the wear, at both the bearing surfaces and taper-trunnion interface of 36 mm BIOLOXdelta CoC THRs, mounted on 12/14 titanium (Ti6Al4V) trunnions was investigated using a hip simulator (n = 3) and a dynamically loaded CoC sample in a separate test station. Wear was assessed gravimetrically and surface roughness measurements of the articulating and taper surfaces taken at regular intervals. Scanning electron microscopy, surface roughness, and gravimetric measurements of trunnions was performed. After 5 million cycles, the mean total wear from the ceramic articulating surfaces was 0.25 mm(3) from the hip simulator test, and that from the titanium trunnions was 0.29 mm(3). This metal wear may provide an explanation for adverse reaction to metal debris found in contemporary CoC hip joints. It is therefore vital to consider taper-trunnion wear in preclinical testing of artificial hip joints. (c) 2018 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater 107B: 1199-1209, 2019.
引用
收藏
页码:1199 / 1209
页数:11
相关论文
共 50 条
  • [21] Improved friction and wear performance of micro dimpled ceramic-on-ceramic interface for hip joint arthroplasty
    Roy, Taposh
    Choudhury, Dipankar
    Ghosh, Subir
    Bin Maniat, Azuddin
    Pingguan-Murphy, Belinda
    CERAMICS INTERNATIONAL, 2015, 41 (01) : 681 - 690
  • [22] In situ measurements of local temperature and contact stress magnitude during wear of ceramic-on-ceramic hip joints
    Zhu, Wenliang
    Puppulin, Leonardo
    Leto, Andrea
    Takahashi, Yasuhito
    Sugano, Nobuhiko
    Pezzotti, Giuseppe
    JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2014, 31 : 68 - 76
  • [23] Combined wear and ageing of ceramic-on-ceramic bearings in total hip replacement under edge loading conditions
    Al-Hajjar, Mazen
    Gremillard, Laurent
    Begand, Sabine
    Oberbach, Thomas
    Hans, Karen
    Delfosse, Daniel
    Chevalier, Jerome
    Jennings, Louise M.
    JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2019, 98 : 40 - 47
  • [24] Wear simulator study of the tripolar all ceramic hip prosthesis
    Jennings, L. M.
    Fisher, J.
    Stewart, T. D.
    Masson, B.
    Lazennec, J. -Y.
    Bioceramics and Alternative Bearings in Joint Arthroplasty, 2006, : 307 - 308
  • [25] The Influence of Kinematic Conditions and Variations in Component Positioning on the Severity of Edge Loading and Wear of Ceramic-on-Ceramic Hip Bearings
    Ali, Murat
    Al-Hajjar, Mazen
    Fisher, John
    Jennings, Louise M.
    CERAMICS-SWITZERLAND, 2019, 2 (03): : 488 - 501
  • [26] Differences between the wear couples metal-on-polyethylene and ceramic-on-ceramic in the stability against dislocation of total hip replacement
    Rainer Bader
    Erwin Steinhauser
    Susanne Zimmermann
    Wolfram Mittelmeier
    Roger Scholz
    Raymonde Busch
    Journal of Materials Science: Materials in Medicine, 2004, 15 : 711 - 718
  • [27] Differences between the wear couples metal-on-polyethylene and ceramic-on-ceramic in the stability against dislocation of total hip replacement
    Bader, R
    Steinhauser, E
    Zimmermann, S
    Mittelmeier, W
    Scholz, R
    Busch, R
    JOURNAL OF MATERIALS SCIENCE-MATERIALS IN MEDICINE, 2004, 15 (06) : 711 - 718
  • [28] Severe Wear and multiple Pseudotumor formation due to revision for ceramic head breakage after ceramic-on-ceramic Total hip arthroplasty: a case report
    Xing, Dan
    Yang, Chaolei
    Li, Rujun
    Hou, Yunfei
    Kou, Bolong
    Li, Hu
    Lin, Jianhao
    BMC MUSCULOSKELETAL DISORDERS, 2019, 20 (1)
  • [29] Severe Wear and multiple Pseudotumor formation due to revision for ceramic head breakage after ceramic-on-ceramic Total hip arthroplasty: a case report
    Dan Xing
    Chaolei Yang
    Rujun Li
    Yunfei Hou
    Bolong Kou
    Hu Li
    Jianhao Lin
    BMC Musculoskeletal Disorders, 20
  • [30] Uncemented third-generation ceramic-on-ceramic total hip arthroplasty using metal acetabular shell with direct taper locking liner
    Sachiyuki Tsukada
    Motohiro Wakui
    Munenori Matsueda
    Archives of Orthopaedic and Trauma Surgery, 2013, 133 : 861 - 868