Wear at the taper-trunnion junction of contemporary ceramic-on-ceramic hips shown in a multistation hip simulator

被引:13
|
作者
Bhalekar, Rohan M. [1 ]
Smith, Simon L. [1 ]
Joyce, Thomas J. [1 ]
机构
[1] Newcastle Univ, Sch Engn, Newcastle Upon Tyne NE1 7RU, Tyne & Wear, England
关键词
total hip replacement; ceramic-on-ceramic; hip simulator; taper-trunnion junction; wear; TITANIUM-ALLOY; IN-VITRO; METAL; FAILURE; HEAD; PROSTHESES; UHMWPE; DEBRIS; MICROSEPARATION; BEARINGS;
D O I
10.1002/jbm.b.34213
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Ceramic-on-ceramic (CoC) total hip replacements (THR) have a substantially lower wear rate than metal-on-polyethylene (MoP) hips, as shown by hip simulator testing. However, the revision rates of CoC and MoP hips are comparable. To try and explain this discrepancy the wear, at both the bearing surfaces and taper-trunnion interface of 36 mm BIOLOXdelta CoC THRs, mounted on 12/14 titanium (Ti6Al4V) trunnions was investigated using a hip simulator (n = 3) and a dynamically loaded CoC sample in a separate test station. Wear was assessed gravimetrically and surface roughness measurements of the articulating and taper surfaces taken at regular intervals. Scanning electron microscopy, surface roughness, and gravimetric measurements of trunnions was performed. After 5 million cycles, the mean total wear from the ceramic articulating surfaces was 0.25 mm(3) from the hip simulator test, and that from the titanium trunnions was 0.29 mm(3). This metal wear may provide an explanation for adverse reaction to metal debris found in contemporary CoC hip joints. It is therefore vital to consider taper-trunnion wear in preclinical testing of artificial hip joints. (c) 2018 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater 107B: 1199-1209, 2019.
引用
收藏
页码:1199 / 1209
页数:11
相关论文
共 50 条
  • [1] Squeaking in Ceramic-On-Ceramic Hips: No Evidence of Contribution From the Trunnion Morse Taper
    Roffe, Lloyd
    FitzPatrick, Anthony J.
    Rodgers, Geoffrey W.
    Woodfield, Tim B. F.
    Hooper, Gary J.
    JOURNAL OF ORTHOPAEDIC RESEARCH, 2017, 35 (08) : 1793 - 1798
  • [2] Hip simulator testing of the taper-trunnion junction and bearing surfaces of contemporary metal-on-cross-linked-polyethylene hip prostheses
    Bhalekar, Rohan M.
    Smith, Simon L.
    Joyce, Thomas J.
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2020, 108 (01) : 156 - 166
  • [3] Pushing Ceramic-on-Ceramic in the most extreme wear conditions: A hip simulator study
    De Fine, Marcello
    Terrando, Silvio
    Hintner, Martin
    Porporati, Alessandro Alan
    Pignatti, Giovanni
    ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2021, 107 (01)
  • [4] A comparison of metal/metal and ceramic/metal taper-trunnion modular connections in explanted total hip replacements
    Cadel, Eileen S.
    Topoleski, L. D. Timmie
    Vesnovsky, Oleg
    Anderson, Charles R.
    Hopper, Robert H., Jr.
    Engh, Charles A., Jr.
    Di Prima, Matthew A.
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2022, 110 (01) : 135 - 143
  • [5] Wear characteristics of ceramic-on-ceramic for hip endoprostheses
    Fruh, HJ
    Willmann, G
    Pfaff, HG
    BIOMATERIALS, 1997, 18 (12) : 873 - 876
  • [6] Wear Prediction of Ceramic-on-Ceramic Artificial Hip Joints
    Askari, E.
    Flores, P.
    Dabirrahmani, D.
    Appleyard, R.
    NEW TRENDS IN MECHANISM AND MACHINE SCIENCE: FROM FUNDAMENTALS TO INDUSTRIAL APPLICATIONS, 2015, 24 : 463 - 470
  • [7] Wear of novel ceramic-on-ceramic bearings under adverse and clinically relevant hip simulator conditions
    Al-Hajjar, Mazen
    Jennings, Louise M.
    Begand, Sabine
    Oberbach, Thomas
    Delfosse, Daniel
    Fisher, John
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2013, 101 (08) : 1456 - 1462
  • [8] Hip simulator wear comparison of metal-on-metal, ceramic-on-ceramic and crosslinked UHMWPE bearings
    Essner, A
    Sutton, K
    Wang, A
    WEAR, 2005, 259 : 992 - 995
  • [9] Revision Hip Arthroplasty for Ceramic-on-Ceramic Squeaking Hips Does Not Compromise the Results
    Matar, Wadih Y.
    Restrepo, Camilo
    Parvizi, Javad
    Kurtz, Steven M.
    Hozack, William J.
    JOURNAL OF ARTHROPLASTY, 2010, 25 (06): : 81 - 86
  • [10] Clinical failure analysis of contemporary ceramic-on-ceramic total hip replacements
    Brandt, Jan-M
    Gascoyne, Trevor C.
    Guenther, Leah E.
    Allen, Andrew
    Hedden, David R.
    Turgeon, Thomas R.
    Bohm, Eric R.
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2013, 227 (08) : 833 - 846