Evaluation of the i-STAT analyzer for determination of ionized calcium concentrations in bovine blood

被引:18
|
作者
Yilmaz, Oznur [1 ]
Karapinar, Tolga [1 ]
机构
[1] Firat Univ, Dept Internal Med, Fac Vet Med, Elazig, Turkey
关键词
calcium; cattle; cow; point-of-care; sensitivity; PORTABLE CLINICAL ANALYZER; GAS PARTIAL PRESSURES; SUBCLINICAL HYPOCALCEMIA; ELECTROLYTE CONCENTRATIONS; NEUTROPHIL FUNCTION; DAIRY; PREVALENCE; PREDICTION; PLASMA; DEVICE;
D O I
10.1111/vcp.12705
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Background Information regarding the performance of the i-STAT, a point-of-care analyzer, to determine ionized calcium (iCa(2+)) concentration in bovine blood is limited. Objective The primary goal of this study was to evaluate the agreement of the i-STAT against a traditional bench-top blood gas analyzer (Radiometer ABL800 FLEX) in the measurement of iCa(2+) in the blood of cattle with various disorders. Methods In this study, 121 cattle with various diseases were used. Venous blood samples were run on the i-STAT and the Radiometer ABL800 FLEX for determination of blood iCa(2+) concentrations. Data were subjected to Passing-Bablok regression and Bland-Altman plots. The Radiometer ABL800 FLEX was the reference method and the i-STAT was the test method. The i-STAT performance was evaluated at two different cutoff values for iCa(2+) concentrations (<1.00 and <1.18 mmol/L) using the receiver operating characteristic curve. Results The Passing-Bablok regression equation that fit the iCa(2+) concentration as measured with the i-STAT (Y) and the Radiometer ABL800 FLEX (X), was Y = -0.108276 [-0.15/-0.06] + 1.103448 [1.06/1.14] x X, with a residual deviation of 0.02. The deviation from the identity line was not significant (P = 0.64). The Bland-Altman difference plot revealed no bias between the i-STAT and the Radiometer ABL 800 FLEX. The i-STAT had a sensitivity of 100% (95% CI, 87.7-100.0) and 94.8% (95% CI, 87.2-98.6) at the iCa(2+) concentration cutoff values of <1.00 and <1.18 mmol/L, respectively. Conclusions iCa(2+) concentrations in bovine blood measured with the i-STAT agreed with those of the Radiometer ABL800 FLEX. In particular, the i-STAT predicted hypocalcemia with a high sensitivity.
引用
收藏
页码:31 / 35
页数:5
相关论文
共 50 条
  • [41] Validation of the i-STAT system for the analysis of blood parameters in fish
    Harter, T. S.
    Shartau, R. B.
    Brauner, C. J.
    Farrell, A. P.
    CONSERVATION PHYSIOLOGY, 2014, 2 (01):
  • [42] AN EVALUATION OF THE CHEMPRO-1000 ANALYZER FOR THE DETERMINATION OF POTASSIUM, SODIUM, PH AND IONIZED CALCIUM
    HARDING, RS
    CLINICAL CHEMISTRY, 1986, 32 (06) : 1105 - 1106
  • [43] EVALUATION OF THE NOVA-7 ANALYZER FOR THE SIMULTANEOUS DETERMINATION OF PH, IONIZED CALCIUM AND TOTAL CALCIUM IN SERUM
    OSTERMANN, H
    WIDDOWSON, G
    JOURNAL OF CLINICAL CHEMISTRY AND CLINICAL BIOCHEMISTRY, 1985, 23 (09): : 603 - 603
  • [44] EVALUATION OF THE NOVA-7 ANALYZER FOR THE SIMULTANEOUS DETERMINATION OF PH, IONIZED CALCIUM AND TOTAL CALCIUM IN SERUM
    WIDDOWSON, G
    ANNALES DE BIOLOGIE CLINIQUE, 1985, 43 (04) : 672 - 672
  • [45] Evaluation of the blood-saving Kaolin i-STAT ACT technique in paediatric cardiac anaesthesia
    Ulmer, F.
    Bettex, D.
    Baulig, W.
    Buerki, C.
    Weiss, M.
    SWISS MEDICAL WEEKLY, 2008, 138 (41-42) : 11S - 11S
  • [46] EVALUATION OF A NEW CALCIUM ANALYZER AS A STAT INSTRUMENT
    VANDREAL, PA
    BLACKBUR.R
    CLINICAL CHEMISTRY, 1974, 20 (07) : 908 - 908
  • [47] i-STAT hand-held point-of-care analyzer for dialysis units
    Gault, MH
    Harding, CE
    Duffett, S
    Longerich, L
    NEPHRON, 1998, 80 (03): : 344 - 348
  • [48] Evaluation of the i-STAT Alinity v in a veterinary clinical setting
    Burke, Jasper E.
    Nguyen, Thuy Hien T.
    Davis, Tanya
    Koenig, Amie
    Lane, Selena L.
    Good, Jennifer
    Brainard, Benjamin M.
    JOURNAL OF VETERINARY DIAGNOSTIC INVESTIGATION, 2021, 33 (04) : 703 - 710
  • [49] EVALUATION OF THE I-STAT(TM) SYSTEM - A PORTABLE CHEMISTRY ANALYZER FOR THE MEASUREMENT OF SODIUM, POTASSIUM, CHLORIDE, UREA, GLUCOSE, AND HEMATOCRIT
    MOCK, T
    MORRISON, D
    YATSCOFF, R
    CLINICAL BIOCHEMISTRY, 1995, 28 (02) : 187 - 192
  • [50] MEASUREMENT OF SERUM AND PLASMA IONIC CALCIUM WITH SPACE-STAT 20 IONIZED CALCIUM ANALYZER
    HUSDAN, H
    LEUNG, M
    OREOPOULOS, D
    RAPOPORT, A
    CLINICAL CHEMISTRY, 1977, 23 (09) : 1775 - 1777