Evaluation of the i-STAT analyzer for determination of ionized calcium concentrations in bovine blood

被引:18
|
作者
Yilmaz, Oznur [1 ]
Karapinar, Tolga [1 ]
机构
[1] Firat Univ, Dept Internal Med, Fac Vet Med, Elazig, Turkey
关键词
calcium; cattle; cow; point-of-care; sensitivity; PORTABLE CLINICAL ANALYZER; GAS PARTIAL PRESSURES; SUBCLINICAL HYPOCALCEMIA; ELECTROLYTE CONCENTRATIONS; NEUTROPHIL FUNCTION; DAIRY; PREVALENCE; PREDICTION; PLASMA; DEVICE;
D O I
10.1111/vcp.12705
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Background Information regarding the performance of the i-STAT, a point-of-care analyzer, to determine ionized calcium (iCa(2+)) concentration in bovine blood is limited. Objective The primary goal of this study was to evaluate the agreement of the i-STAT against a traditional bench-top blood gas analyzer (Radiometer ABL800 FLEX) in the measurement of iCa(2+) in the blood of cattle with various disorders. Methods In this study, 121 cattle with various diseases were used. Venous blood samples were run on the i-STAT and the Radiometer ABL800 FLEX for determination of blood iCa(2+) concentrations. Data were subjected to Passing-Bablok regression and Bland-Altman plots. The Radiometer ABL800 FLEX was the reference method and the i-STAT was the test method. The i-STAT performance was evaluated at two different cutoff values for iCa(2+) concentrations (<1.00 and <1.18 mmol/L) using the receiver operating characteristic curve. Results The Passing-Bablok regression equation that fit the iCa(2+) concentration as measured with the i-STAT (Y) and the Radiometer ABL800 FLEX (X), was Y = -0.108276 [-0.15/-0.06] + 1.103448 [1.06/1.14] x X, with a residual deviation of 0.02. The deviation from the identity line was not significant (P = 0.64). The Bland-Altman difference plot revealed no bias between the i-STAT and the Radiometer ABL 800 FLEX. The i-STAT had a sensitivity of 100% (95% CI, 87.7-100.0) and 94.8% (95% CI, 87.2-98.6) at the iCa(2+) concentration cutoff values of <1.00 and <1.18 mmol/L, respectively. Conclusions iCa(2+) concentrations in bovine blood measured with the i-STAT agreed with those of the Radiometer ABL800 FLEX. In particular, the i-STAT predicted hypocalcemia with a high sensitivity.
引用
收藏
页码:31 / 35
页数:5
相关论文
共 50 条
  • [31] ANALYTICAL EVALUATION OF I-STAT PORTABLE CLINICAL ANALYZER AND USE BY NONLABORATORY HEALTH-CARE PROFESSIONALS
    JACOBS, E
    VADASDI, E
    SARKOZI, L
    COLMAN, N
    CLINICAL CHEMISTRY, 1993, 39 (06) : 1069 - 1074
  • [32] EVALUATION OF AN IONIZED CALCIUM-PH ANALYZER
    KROUWER, JS
    OMALLEY, J
    CLINICAL CHEMISTRY, 1988, 34 (06) : 1234 - 1234
  • [33] Effect of emicizumab on activated clotting time performed on i-STAT Alinity analyzer
    Seyve, Landry
    Prigent, Jean Baptiste
    Lo Presti, Caroline
    Bedague, Damien
    Szymanski, Gautier
    Bulabois, Benedicte
    Barro, Claire
    Marlu, Raphael
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2024, 46 (06) : 1132 - 1135
  • [34] Analytical Performances of the Novel i-STAT Alinity Point-of-Care Analyzer
    Larcher, Romaric
    Lottelier, Maxence
    Badiou, Stephanie
    Dupuy, Anne-Marie
    Bargnoux, Anne-Sophie
    Cristol, Jean-Paul
    DIAGNOSTICS, 2023, 13 (02)
  • [35] Selected Blood Chemistry and Gas Reference Ranges for Broiler Breeders Using the i-STAT® Handheld Clinical Analyzer
    Martin, Michael P.
    Wineland, Michael
    Barnes, H. John
    AVIAN DISEASES, 2010, 54 (03) : 1016 - 1020
  • [36] Evaluation of the IRMA TRUpoint and i-STAT creatinine assays
    Nichols, James H.
    Bartholomew, Cathy
    Bonzagi, Arianna
    Garb, Jane L.
    Jin, Long
    CLINICA CHIMICA ACTA, 2007, 377 (1-2) : 201 - 205
  • [37] Improvised Method for Increasing the Temperature of an i-STAT Analyzer and Cartridge in Cold Environments
    Radler, David
    Wetschler, Matthew
    Christensen, Mark
    Lipman, Grant S.
    WILDERNESS & ENVIRONMENTAL MEDICINE, 2015, 26 (02) : 280 - 282
  • [38] The portable laboratory:: an evaluation of the accuracy and reproducibility of i-STAT©
    Bingham, D
    Kendall, J
    Clancy, M
    ANNALS OF CLINICAL BIOCHEMISTRY, 1999, 36 : 66 - 71
  • [39] Selected Blood Chemistry Values in Mobility-Impaired Broiler Breeder Hens with Suspected Calcium Tetany Using the i-STAT® Handheld Clinical Analyzer
    Martin, Michael P.
    Wineland, Michael
    Fletcher, Oscar J.
    Barnes, H. John
    AVIAN DISEASES, 2011, 55 (03) : 340 - 345
  • [40] Evaluation of the i-STAT portable clinical analyzer for point-of-care blood testing in the intensive care units of a university children's hospital
    Papadea, C
    Foster, J
    Grant, S
    Ballard, SA
    Cate, JC
    Southgate, M
    Purohit, DM
    ANNALS OF CLINICAL AND LABORATORY SCIENCE, 2002, 32 (03): : 231 - 243