Evaluation of the i-STAT Alinity v in a veterinary clinical setting

被引:3
|
作者
Burke, Jasper E. [1 ]
Nguyen, Thuy Hien T. [2 ]
Davis, Tanya [2 ]
Koenig, Amie [1 ]
Lane, Selena L. [1 ]
Good, Jennifer [1 ]
Brainard, Benjamin M. [1 ]
机构
[1] Univ Georgia, Coll Vet Med, Dept Small Anim Med & Surg, Athens, GA USA
[2] Zoetis Inc, Parsippany, NJ USA
关键词
blood gas; cats; dogs; electrolytes; emergency; point-of-care; BLOOD-GASES; ANALYZER; PH; ELECTROLYTES; GUIDELINES; ACCURACY;
D O I
10.1177/10406387211019710
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Many point-of-care (POC) analyzers are available for the measurement of electrolytes and acid-base status in animals. We assessed the precision of the i-STAT Alinity v, a recently introduced POC analyzer, and compared it to 2 commonly used and previously validated POC analyzers (i-STAT 1, Stat Profile pHOx Ultra). Precision was evaluated by performing multiple analyses of whole blood samples from healthy dogs, cats, and horses on multiple i-STAT Alinity v analyzers. For comparison between analyzers, whole blood samples from dogs and cats presented to the emergency room were run concurrently on all 3 POC instruments. Reported values were compared by species (dogs and cats only) using Pearson correlation, and all values from all species were analyzed together for the Bland-Altman analysis. Results suggested that the i-STAT Alinity v precision was very good, with median coefficients of variability <2.5% for all measured parameters (except the anion gap), with variable ranges of coefficients of variation. In addition, good-to-excellent correlation was observed between the i-STAT Alinity v and i-STAT 1, and between the i-STAT Alinity v and Stat Profile pHOx Ultra for all parameters in both cats and dogs, respectively. In this cohort, the i-STAT Alinity v had clinically acceptable bias compared to the currently marketed analyzers and can be used for monitoring measured analytes in cats and dogs, although serial measurements in a single animal should be performed on the same analyzer whenever possible.
引用
收藏
页码:703 / 710
页数:8
相关论文
共 50 条
  • [1] Evaluation of the i-STAT Alinity Point-of-Care Analyzer
    Lee, Mee-Yin
    Lim, Sian-Foong
    Lam, Leslie
    POINT OF CARE, 2019, 18 (02): : 48 - 55
  • [2] Evaluation of the i-STAT Alinity point-of-care analyzer in pediatric patient populations
    Di Meo, Ashley
    Lee, Elsie
    Quach, Kim
    Nemec, Rose
    Zappitelli, Michael
    Licht, Christoph
    Adeli, Khosrow
    CLINICAL BIOCHEMISTRY, 2022, 109 : 104 - 104
  • [3] Evaluation of the i-STAT Portable Clinical Analyzer in dogs
    Matburger, C
    Henke, J
    Hirschberger, J
    Matburger, S
    Erhardt, W
    TIERARZTLICHE PRAXIS AUSGABE KLEINTIERE HEIMTIERE, 2000, 28 (02): : 132 - 137
  • [4] Effect of emicizumab on activated clotting time performed on i-STAT Alinity analyzer
    Seyve, Landry
    Prigent, Jean Baptiste
    Lo Presti, Caroline
    Bedague, Damien
    Szymanski, Gautier
    Bulabois, Benedicte
    Barro, Claire
    Marlu, Raphael
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2024, 46 (06) : 1132 - 1135
  • [5] Analytical Performances of the Novel i-STAT Alinity Point-of-Care Analyzer
    Larcher, Romaric
    Lottelier, Maxence
    Badiou, Stephanie
    Dupuy, Anne-Marie
    Bargnoux, Anne-Sophie
    Cristol, Jean-Paul
    DIAGNOSTICS, 2023, 13 (02)
  • [6] Evaluation of the i-STAT portable clinical analyzer for use in a helicopter
    Burritt, MF
    Santrach, PJ
    Hankins, DG
    Herr, D
    Newton, NC
    SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 1996, 56 : 121 - 128
  • [7] Evaluation of i-STAT portable clinical analyzer in a hemodialysis unit
    Gault, MH
    Harding, CE
    CLINICAL BIOCHEMISTRY, 1996, 29 (02) : 117 - 124
  • [8] Evaluation of i-STAT creatinine assay
    Gault, MH
    Seymour, ME
    Howell, WE
    NEPHRON, 2001, 88 (02): : 178 - 182
  • [9] Precision Evaluation of the i-STAT® Glucose Test in the i-STAT G Cartridge using the i-STAT 1 System for Venous, Arterial and Capillary Whole Blood Specimens in Clinical Settings
    Thibodeau, L.
    Aguirre, M. Gonzalez
    Hahn, S.
    Zhang, H.
    CLINICAL CHEMISTRY, 2024, 70 : I129 - I129
  • [10] Evaluation of the i-STAT portable clinical analyzer in chickens (Gallus gallus)
    Steinmetz, Hanspeter W.
    Vogt, Rainer
    Kastner, Sabine
    Riond, Barbara
    Hatt, Jean-Michel
    JOURNAL OF VETERINARY DIAGNOSTIC INVESTIGATION, 2007, 19 (04) : 382 - 388