Biomechanical analysis of a novel hook-screw technique for C1-2 stabilization Technical note

被引:4
|
作者
Reis, Marco Tulio [1 ]
Nottmeier, Eric W. [2 ]
Reyes, Phillip M. [1 ]
Baek, Seungwon [1 ]
Crawford, Neil R. [1 ]
机构
[1] St Josephs Hosp, Barrow Neurol Inst, Spinal Biomech Lab, Dept Neurosurg Res, Phoenix, AZ 85013 USA
[2] Mayo Clin, Dept Neurosurg, Jacksonville, FL 32224 USA
关键词
atlantoaxial instability; cervical spine; stabilization; craniovertebral junction; cervical laminar hook; posterior stabilization; posterior cervical surgery; POSTERIOR FIXATION TECHNIQUES; ATLANTOAXIAL FIXATION; OCCIPITOCERVICAL FUSION; POLYAXIAL SCREW; IN-VITRO; INSTRUMENTATION; SPINE; STABILITY; PEDICLE; PLATE;
D O I
10.3171/2012.5.SPINE1242
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
The Food and Drug Administration has not cleared the following medical devices for the use described in this study. The following medical devices are being discussed for an off-label use: cervical lateral mass screws. Object. As an alternative for cases in which the anatomy and spatial relationship between C-2 and a vertebral artery precludes insertion of C-2 pedicle/pars or C1-2 transarticular screws, a technique that includes opposing laminar hooks (claw) at C-2 combined with C-1 lateral mass screws may be used. The biomechanical stability of this alternate technique was compared with that of a standard screw-rod technique in vitro. Methods. Flexibility tests were performed in 7 specimens (occiput to C-3) in the following 6 different conditions: 1) intact; 2) after creating instability and attaching a posterior cable/graft at C1-2; 3) after removing the graft and attaching a construct comprising C-1 lateral mass screws and C-2 laminar claws; 4) after reattaching the posterior cable-graft at C1-2 (posterior hardware still in place); 5) after removing the posterior cable-graft and laminar hooks and placing C-2 pedicle screws interconnected to C-1 lateral mass screws via rod; and 6) after reattaching the posterior cable-graft at C1-2 (screw-rod construct still in place). Results. All types of stabilization significantly reduced the range of motion, lax zone, and stiff zone compared with the intact condition. There was no significant biomechanical difference in terms of range of motion or lax zone between the screw-rod construct and the screw-claw-rod construct in any direction of loading. Conclusions. The screw-claw-rod technique restricts motion much like the standard Harms technique, making it an acceptable alternative technique when aberrant arterial anatomy precludes the placement of C-2 pars/pedicle screws or C1-2 transarticular screws. (http :thejns.org/doi/abs/10.3171/2012.5.SPINE1242)
引用
收藏
页码:220 / 226
页数:7
相关论文
共 50 条
  • [1] Biomechanical analysis of Goel technique for C1-2 fusion
    Park, Jon
    Scheer, Justin K.
    Lim, T. Jesse
    Deviren, Vedat
    Ames, Christopher P.
    JOURNAL OF NEUROSURGERY-SPINE, 2011, 14 (05) : 639 - 646
  • [2] Biomechanical Study of Horizontal Screw-screw Crosslink in C1-2 Pedicle Screw-rod Fixation
    Ouyang, Beiping
    Zou, Xiaobao
    Luo, Chunshan
    Lu, Tingsheng
    Ma, Xiangyang
    CLINICAL SPINE SURGERY, 2023, 36 (10): : E519 - E523
  • [3] OPERATIVE STABILIZATION IN C1-2 FRACTURES
    GROB, D
    MAGERL, F
    ORTHOPADE, 1987, 16 (01): : 46 - 54
  • [4] Biomechanical characteristics of C1-2 cable fixations
    Dickman, CA
    Crawford, NR
    Paramore, CG
    JOURNAL OF NEUROSURGERY, 1996, 85 (02) : 316 - 322
  • [5] ANTERIOR SCREW FIXATION FOR C1-2 INSTABILITY
    APFELBAUM, RI
    JOURNAL OF NEUROSURGERY, 1989, 70 (02) : A316 - A316
  • [6] Placement of baclofen pump catheter through a C1-2 puncture: technical note
    Aljuboori, Zaid
    Archer, Jacob
    Huff, Wei
    Moreno, Amee
    Jea, Andrew
    JOURNAL OF NEUROSURGERY-PEDIATRICS, 2018, 21 (04) : 389 - 394
  • [7] Biomechanical Assessment of Bilateral C1 Laminar Hook and C1-2 Transarticular Screws and Bone Graft for Atlantoaxial Instability
    Guo, Xiang
    Ni, Bin
    Zhao, Weidong
    Wang, Mingfei
    Zhou, Fengjin
    Li, Songkai
    Ren, Zhongwu
    JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2009, 22 (08): : 578 - 585
  • [8] Placement of unilateral lag screw through the lateral mass of C-1: description of a novel technique Technical note
    Tabbosha, Monir
    Dowdy, Justin
    Pait, T. Glenn
    JOURNAL OF NEUROSURGERY-SPINE, 2013, 19 (01) : 128 - 132
  • [9] Surgical nuances and construct patterns influence construct stiffness in C1-2 stabilizations: a biomechanical study of C1-2 gapping and advanced C1-2 fixation
    Heiko Koller
    Sebastian Hartmann
    Gmeiner Raphael
    Werner Schmölz
    Christoph Orban
    Claudius Thome
    European Spine Journal, 2021, 30 : 1596 - 1606
  • [10] Surgical nuances and construct patterns influence construct stiffness in C1-2 stabilizations: a biomechanical study of C1-2 gapping and advanced C1-2 fixation
    Koller, Heiko
    Hartmann, Sebastian
    Raphael, Gmeiner
    Schmoelz, Werner
    Orban, Christoph
    Thome, Claudius
    EUROPEAN SPINE JOURNAL, 2021, 30 (06) : 1596 - 1606