Comparison of Accuracy and Safety of the SEVEN and the Navigator Continuous Glucose Monitoring Systems

被引:54
|
作者
Garg, Satish K. [1 ,2 ,3 ]
Smith, James [1 ]
Beatson, Christie [1 ]
Lopez-Baca, Benita [1 ]
Voelmle, Mary [1 ]
Gottlieb, Peter A. [1 ,2 ,3 ]
机构
[1] Univ Colorado Denver, Sch Med, Barbara Davis Ctr Childhood Diabet, Aurora, CO 80045 USA
[2] Univ Colorado Denver, Sch Med, Dept Internal Med, Aurora, CO USA
[3] Univ Colorado Denver, Sch Med, Dept Pediat, Aurora, CO USA
关键词
BLOOD-GLUCOSE;
D O I
10.1089/dia.2008.0109
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: This study evaluated the accuracy and safety of two continuous glucose monitoring (CGM) systems, the SEVEN (R) (DexCom (TM), San Diego, CA) and the Navigator (R) (Abbott Diabetes Care, Alameda, CA), with the YSI laboratory measurements of blood glucose (blood glucose meter manufactured by YSI, Yellow Springs, OH), when worn concurrently in adults with type 1 diabetes. Design and Methods: Fourteen subjects with type 1 diabetes, 33 +/- 6 (mean +/- SD) years old, were enrolled in this study. All subjects wore both sensors concurrently over three consecutive 5-day CGM sessions (15-day wear). On Days 5, 10, and 15, subjects participated in an 8-h in-clinic session where measurements from the CGM systems were collected and compared with YSI measurements every 15 min. At the end of Day 5 and 10 in-clinic sessions, the sensors were removed, and new sensors were inserted for the following CGM session despite the SEVEN system's recommended use for up to 7 days. Results: The mean absolute relative difference (ARD) for the two CGM devices versus YSI was not different: 16.8% and 16.1% for SEVEN and Navigator, respectively (P = 0.38). In the hypoglycemic region (YSI value <80 mg/dL), the mean ARD for SEVEN was lower than for Navigator (21.5% vs. 29.8%, respectively; P = 0.001). The data analyses were similar when compared with self-monitoring of blood glucose (SMBG) values. Thirteen additional Navigator replacement devices were issued compared to two for the SEVEN. A total of three versus 14 skin reactions were reported with the SEVEN and Navigator insertion area, respectively. Conclusions: Glucose measurements with the SEVEN and Navigator were found to be similar compared with YSI and SMBG measurements, with the exception of the hypoglycemic range where the SEVEN performed better. However, the Navigator caused more skin area reactions.
引用
收藏
页码:65 / 72
页数:8
相关论文
共 50 条
  • [21] Detection of hypoglycemic events with the FreeStyle navigator® continuous glucose monitoring system
    McGarraugh, Geoffrey
    DIABETES, 2007, 56 : A110 - A110
  • [22] Home use evaluation of the freestyle navigator continuous glucose monitoring system
    Bugler, J. R.
    Bergenstal, R. M.
    Bode, B. W.
    Weinstein, R. L.
    Schwartz, S. L.
    Bernstein, R. M.
    ElDeiry, S. S.
    DIABETOLOGIA, 2006, 49 : 582 - 582
  • [23] Comparison of Accuracy of Freestyle Libre Pro and Medtronic iPro2 Continuous Glucose Monitoring Systems in Pregnancy
    Golla, Kiran Kumar
    Gupta, Yashdeep
    Goyal, Alpesh
    Kalaivani, Mani
    Kachhawa, Garima
    Kulshrestha, Vidushi
    Sharma, Aparna. K. K.
    Meena, Jyoti
    Bharti, Juhi
    Sharma, Jai. B. B.
    Dadhwal, Vatsla
    Malhotra, Neena
    Bhatla, Neerja
    Tandon, Nikhil
    DIABETES TECHNOLOGY & THERAPEUTICS, 2023, 25 (08) : 538 - 542
  • [24] CONTINUOUS MONITORING SYSTEMS OF GLUCOSE
    Vidal, Merce
    Jansa, Margarida
    REVISTA ROL DE ENFERMERIA, 2013, 36 (04): : 275 - 278
  • [25] Accuracy and Safety of the 15-Day CareSens Air Continuous Glucose Monitoring System
    Kim, Kyung-Soo
    Lee, Seung-Hwan
    Yoo, Won Sang
    Park, Cheol-Young
    DIABETES TECHNOLOGY & THERAPEUTICS, 2024, 26 (04) : 222 - 228
  • [26] Accuracy and Safety of Dexcom G7 Continuous Glucose Monitoring in Adults with Diabetes
    Garg, Satish K.
    Kipnes, Mark
    Castorino, Kristin
    Bailey, Timothy S.
    Akturk, Halis Kaan
    Welsh, John B.
    Christiansen, Mark P.
    Balo, Andrew K.
    Brown, Sue A.
    Reid, Jennifer L.
    Beck, Stayce E.
    DIABETES TECHNOLOGY & THERAPEUTICS, 2022, 24 (06) : 373 - 380
  • [27] Efficacy and safety comparison of continuous glucose monitoring and self-monitoring of blood glucose in type 1 diabetes
    Wojciechowski, Piotr
    Rys, Przemyslaw
    Lipowska, Anna
    Gaweska, Magdalena
    Malecki, Maciej T.
    POLSKIE ARCHIWUM MEDYCYNY WEWNETRZNEJ-POLISH ARCHIVES OF INTERNAL MEDICINE, 2011, 121 (10): : 333 - 343
  • [28] Comparison of the accuracy of seven home blood glucose meters
    Koulas, Georgios
    Sampanis, Christos
    Tziomalos, Konstantinos
    Semertzidis, Panagiotis
    Athanasiadou, Zoe
    Dogramatzi, Fani
    Machera, Konstantina
    Panagiotidou, Dimitra
    Zamboulis, Chrysanthos
    DIABETES, 2006, 55 : A96 - A96
  • [29] Hypoglycemia Reduction and Accuracy of Continuous Glucose Monitoring
    Kovatchev, Boris P.
    DIABETES TECHNOLOGY & THERAPEUTICS, 2015, 17 (08) : 530 - 533
  • [30] Clarke Error Grid and continuous Clarke Error Grid Analysis of the accuracy of the 5-day FreeStyle Navigator Continuous Glucose Monitoring System
    Weinstein, R. L.
    Schwartz, S. L.
    Brazg, R. L.
    Bugler, J. R.
    Peyser, T. A.
    McGarraugh, G. V.
    DIABETOLOGIA, 2006, 49 : 583 - 583