Comparison of Accuracy and Safety of the SEVEN and the Navigator Continuous Glucose Monitoring Systems

被引:54
|
作者
Garg, Satish K. [1 ,2 ,3 ]
Smith, James [1 ]
Beatson, Christie [1 ]
Lopez-Baca, Benita [1 ]
Voelmle, Mary [1 ]
Gottlieb, Peter A. [1 ,2 ,3 ]
机构
[1] Univ Colorado Denver, Sch Med, Barbara Davis Ctr Childhood Diabet, Aurora, CO 80045 USA
[2] Univ Colorado Denver, Sch Med, Dept Internal Med, Aurora, CO USA
[3] Univ Colorado Denver, Sch Med, Dept Pediat, Aurora, CO USA
关键词
BLOOD-GLUCOSE;
D O I
10.1089/dia.2008.0109
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: This study evaluated the accuracy and safety of two continuous glucose monitoring (CGM) systems, the SEVEN (R) (DexCom (TM), San Diego, CA) and the Navigator (R) (Abbott Diabetes Care, Alameda, CA), with the YSI laboratory measurements of blood glucose (blood glucose meter manufactured by YSI, Yellow Springs, OH), when worn concurrently in adults with type 1 diabetes. Design and Methods: Fourteen subjects with type 1 diabetes, 33 +/- 6 (mean +/- SD) years old, were enrolled in this study. All subjects wore both sensors concurrently over three consecutive 5-day CGM sessions (15-day wear). On Days 5, 10, and 15, subjects participated in an 8-h in-clinic session where measurements from the CGM systems were collected and compared with YSI measurements every 15 min. At the end of Day 5 and 10 in-clinic sessions, the sensors were removed, and new sensors were inserted for the following CGM session despite the SEVEN system's recommended use for up to 7 days. Results: The mean absolute relative difference (ARD) for the two CGM devices versus YSI was not different: 16.8% and 16.1% for SEVEN and Navigator, respectively (P = 0.38). In the hypoglycemic region (YSI value <80 mg/dL), the mean ARD for SEVEN was lower than for Navigator (21.5% vs. 29.8%, respectively; P = 0.001). The data analyses were similar when compared with self-monitoring of blood glucose (SMBG) values. Thirteen additional Navigator replacement devices were issued compared to two for the SEVEN. A total of three versus 14 skin reactions were reported with the SEVEN and Navigator insertion area, respectively. Conclusions: Glucose measurements with the SEVEN and Navigator were found to be similar compared with YSI and SMBG measurements, with the exception of the hypoglycemic range where the SEVEN performed better. However, the Navigator caused more skin area reactions.
引用
收藏
页码:65 / 72
页数:8
相关论文
共 50 条
  • [1] Comparison of Accuracy and Safety of the SEVEN and the Navigator Continuous Glucose Monitoring Systems (vol 11, pg 65, 2009)
    Garg, S. K.
    Smith, J.
    Beatson, C.
    Lopez-Baca, B.
    Voelmle, M.
    Gottlieb, P. A.
    DIABETES TECHNOLOGY & THERAPEUTICS, 2009, 11 (03) : 205 - 205
  • [2] Safety and accuracy of neonatal continuous glucose monitoring
    Ohishi, Akira
    Ueno, Daizo
    Fujita, Tomoka
    Segawa, Yuki
    Yamamoto, Takuya
    Fujisawa, Yasuko
    Iijima, Shigeo
    PEDIATRICS INTERNATIONAL, 2023, 65 (01)
  • [4] Accuracy and Reliability of Continuous Glucose Monitoring Systems: A Head-to-Head Comparison
    Luijf, Yoeri M.
    Mader, Julia K.
    Doll, Werner
    Pieber, Thomas
    Farret, Anne
    Place, Jerome
    Renard, Eric
    Bruttomesso, Daniela
    Filippi, Alessio
    Avogaro, Angelo
    Arnolds, Sabine
    Benesch, Carsten
    Heinemann, Lutz
    DeVries, J. Hans
    DIABETES TECHNOLOGY & THERAPEUTICS, 2013, 15 (08) : 721 - 726
  • [5] Accuracy and Reliability of Current Continuous Glucose Monitoring (CGM) systems; a Direct Comparison
    Y.M. Luijf
    J.H. DeVries
    J.K. Mader
    W. Doll
    J. Place
    E. Renard
    D. Bruttomesso
    A. Filippi
    C. Benesch
    L. Heinemann
    Nederlands Tijdschrift voor Diabetologie, 2012, 10 (3) : 157 - 157
  • [6] The accuracy of the FreeStyle navigator continuous glucose monitoring system in children with type 1 diabetes
    Wilson, Darrell M.
    Beck, Roy W.
    Tamborlane, William V.
    Dontchev, Mariya J.
    Kollman, Craig
    Chase, Peter
    Fox, Larry A.
    Ruedy, Katrina J.
    Tsalikian, Eva
    Weinzimer, Stuart A.
    DIABETES CARE, 2007, 30 (01) : 59 - 64
  • [7] Accuracy of the 5-day freestyle navigator continuous glucose monitoring system - Comparison with frequent laboratory reference measurements
    Weinstein, Richard L.
    Bugler, Jolyon R.
    Schwartz, Sherwyn L.
    Peyser, Thomas A.
    Brazg, Ronald L.
    McGarraugh, Geoffrey V.
    DIABETES CARE, 2007, 30 (05) : 1125 - 1130
  • [8] Comparison of Point Accuracy Between Two Widely Used Continuous Glucose Monitoring Systems
    Hanson, Kevin
    Kipnes, Mark
    Tran, Hien
    JOURNAL OF DIABETES SCIENCE AND TECHNOLOGY, 2024, 18 (03): : 598 - 607
  • [9] Accuracy of the Freestyle Navigator® Continuous Glucose Monitoring System in children with T1DM
    Fox, Larry
    Beck, Roy
    Weinzimer, Stuart
    Ruedy, Katrina
    Kollman, Craig
    Xing, Dongyuan
    Tamborlane, William
    Wilson, Darrell
    Chase, Peter
    Tansey, Michael
    DIABETES, 2006, 55 : A91 - A92
  • [10] Performance of FreeStyle navigator against standards for continuous glucose monitoring systems and ISO15197 system accuracy blood glucose meter guidelines
    Bugler, J. R.
    Pryor, H.
    Harris, C.
    DIABETOLOGIA, 2007, 50 : S419 - S419