A frequency domain compensation scheme for multi-slice helical CT reconstruction with tilted gantry

被引:1
|
作者
Hsieh, J [1 ]
Hu, H [1 ]
机构
[1] GE Co, Med Syst, Milwaukee, WI 53201 USA
关键词
multi-slice computed tomography; tilted helical scan; tomographic reconstruction; image artifact;
D O I
10.1117/12.430998
中图分类号
TP18 [人工智能理论];
学科分类号
081104 ; 0812 ; 0835 ; 1405 ;
摘要
Multi-slice CT (MCT) is one of the most recent technical advancements in computed tomography. MCT offers high volume coverage, faster scan speed, and reduced x-ray tube loading. When combined with the helical scan mode, MCT provides even higher volume coverage as a result of the elimination of the inter-scan delays. Similar to the single slice CT, the projection data collected in MCT is inherently inconsistent.. To combat projection inconsistency, many image reconstruction algorithms were investigated and developed. Recent studies, have shown, however, that the image quality of MCT can be significantly degraded when the data is collected in a tilted helical mode. The degradation can be so severe that it is unacceptable for routine clinical usage. In this paper, we present a detailed investigation to, show that the root cause of the image quality degradation is the deviation of the reconstruction iso-center from the gantry iso-center. An analytical model is derived to establish a quantitative relationship between the iso-center shift and other system parameters. The model serves as the mathematical basis for the derivation of a frequency-domain correction algorithm. Detailed performance evaluation is provided in terms of spatial resolution, noise, computational efficiency, and image artifacts.
引用
收藏
页码:105 / 112
页数:4
相关论文
共 50 条
  • [21] Oblique surface reconstruction to approximate cone-beam helical data in multi-slice CT
    Chen, LG
    Heuscher, DJ
    Liang, Y
    IMAGE RECONSTRUCTION FROM INCOMPLETE DATA, 2000, 4123 : 279 - 284
  • [22] Analytical models for multi-slice helical CT performance parameters
    Hsieh, J
    MEDICAL PHYSICS, 2003, 30 (02) : 169 - 178
  • [23] Pulmonary vein varix: diagnosis with multi-slice helical CT
    Vanherreweghe, E
    Rigauts, H
    Bogaerts, Y
    Meeus, L
    EUROPEAN RADIOLOGY, 2000, 10 (08) : 1315 - 1317
  • [24] Pulmonary vein varix: diagnosis with multi-slice helical CT
    E. Vanherreweghe
    H. Rigauts
    Y. Bogaerts
    L. Meeus
    European Radiology, 2000, 10 : 1315 - 1317
  • [25] Multi-slice helical CT: Principles, imaging characteristics, and performance
    Hu, H
    He, HD
    Fox, S
    Ackelsberg, S
    Strong, G
    Seidenschnur, G
    PROCEEDINGS OF THE 20TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOL 20, PTS 1-6: BIOMEDICAL ENGINEERING TOWARDS THE YEAR 2000 AND BEYOND, 1998, 20 : 637 - 639
  • [26] Correction of a new type of artifact in helical multi-slice CT
    Zmora, I
    Taguchi, K
    Silver, MD
    Han, KS
    RADIOLOGY, 1998, 209P : 434 - 435
  • [27] Multi-sector reconstruction for multi-slice, cardiac CT
    Pan, T
    Shen, Y
    RADIOLOGY, 2000, 217 : 161 - 161
  • [28] Fourier crosstalk analysis of multi-slice and conebeam helical CT
    La Rivière, PJ
    MEDICAL IMAGING 2004: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2004, 5368 : 19 - 28
  • [29] Optimization of Image Acquisition and Reconstruction in Multi-Slice CT
    Yu, L.
    McCollough, C.
    Leng, S.
    Kofler, J.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [30] Analytical noise and slice-sensitivity-profile models for multi-slice helical CT
    Hsieh, J
    MEDICAL IMAGING 2003: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2003, 5030 : 112 - 119