A frequency domain compensation scheme for multi-slice helical CT reconstruction with tilted gantry

被引:1
|
作者
Hsieh, J [1 ]
Hu, H [1 ]
机构
[1] GE Co, Med Syst, Milwaukee, WI 53201 USA
关键词
multi-slice computed tomography; tilted helical scan; tomographic reconstruction; image artifact;
D O I
10.1117/12.430998
中图分类号
TP18 [人工智能理论];
学科分类号
081104 ; 0812 ; 0835 ; 1405 ;
摘要
Multi-slice CT (MCT) is one of the most recent technical advancements in computed tomography. MCT offers high volume coverage, faster scan speed, and reduced x-ray tube loading. When combined with the helical scan mode, MCT provides even higher volume coverage as a result of the elimination of the inter-scan delays. Similar to the single slice CT, the projection data collected in MCT is inherently inconsistent.. To combat projection inconsistency, many image reconstruction algorithms were investigated and developed. Recent studies, have shown, however, that the image quality of MCT can be significantly degraded when the data is collected in a tilted helical mode. The degradation can be so severe that it is unacceptable for routine clinical usage. In this paper, we present a detailed investigation to, show that the root cause of the image quality degradation is the deviation of the reconstruction iso-center from the gantry iso-center. An analytical model is derived to establish a quantitative relationship between the iso-center shift and other system parameters. The model serves as the mathematical basis for the derivation of a frequency-domain correction algorithm. Detailed performance evaluation is provided in terms of spatial resolution, noise, computational efficiency, and image artifacts.
引用
收藏
页码:105 / 112
页数:4
相关论文
共 50 条
  • [1] Gantry tilted helical reconstruction in 16-slice CT
    Hein, IA
    Taguchi, K
    Mori, I
    Kazama, M
    Silver, MA
    RADIOLOGY, 2002, 225 : 254 - 254
  • [2] Multi-slice helical CT: Scan and reconstruction
    Hu, H
    MEDICAL PHYSICS, 1999, 26 (01) : 5 - 18
  • [3] A generalized helical reconstruction algorithm for multi-slice CT
    Hsieh, J
    Toth, TL
    Simoni, P
    Grekowicz, B
    Slack, CC
    Seidenschnur, GE
    RADIOLOGY, 2001, 221 : 217 - 217
  • [4] Algorithm for image reconstruction in multi-slice helical CT
    Taguchi, K
    Aradate, H
    MEDICAL PHYSICS, 1998, 25 (04) : 550 - 561
  • [5] A multi-slice helical CT reconstruction with generalized weighting
    Hsieh, J
    Toth, T
    Simoni, P
    Slack, C
    Grekowicz, B
    Seidenschnur, G
    Wang, S
    2001 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORDS, VOLS 1-4, 2002, : 1732 - 1736
  • [6] Convolution reconstruction algorithm for multi-slice helical CT
    Hsieh, J
    Grekowicz, B
    Simoni, P
    Thibault, JB
    Joshi, MC
    Dutta, S
    Williams, EC
    Shaughnessy, C
    Sainath, P
    MEDICAL IMAGING 2003: IMAGE PROCESSING, PTS 1-3, 2003, 5032 : 716 - 723
  • [7] Adaptive interpolation approach for multi-slice helical CT reconstruction
    Hsieh, J
    MEDICAL IMAGING 2003: IMAGE PROCESSING, PTS 1-3, 2003, 5032 : 1876 - 1883
  • [8] Optimization of view weighting in tilted-plane-based reconstruction algorithms to minimize helical artifacts in multi-slice helical CT
    Tang, XY
    MEDICAL IMAGING 2003: IMAGE PROCESSING, PTS 1-3, 2003, 5032 : 1868 - 1875
  • [9] A new algorithm and evaluation for image reconstruction in multi-slice helical CT
    Taguchi, K
    Aradate, H
    RADIOLOGY, 1997, 205 : 1053 - 1053
  • [10] L0 constrained sparse reconstruction for multi-slice helical CT reconstruction
    Hu, Yining
    Xie, Lizhe
    Luo, Limin
    Nunes, Jean Claude
    Toumoulin, Christine
    PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (04): : 1173 - 1189