Performance of the St. Jude Medical Riata Leads

被引:47
|
作者
Epstein, Andrew E. [1 ]
Baker, James H. [2 ]
Beau, Scott L. [3 ]
Deering, Thomas F. [4 ]
Greenberg, Steven M. [5 ]
Goldman, Daniel S. [6 ]
机构
[1] Univ Alabama Birmingham, Dept Med, Div Cardiovasc Dis, Birmingham, AL 35294 USA
[2] St Thomas Hosp, Nashville, TN USA
[3] Arkansas Heart Hosp, Little Rock, AR USA
[4] Piedmont Hosp, Atlanta, GA USA
[5] St Francis Hosp, Roslyn, NY USA
[6] Ctr Cardiol, Delray Beach, FL USA
关键词
Implantable cardioverter-defibrillator; Lead complications; Perforation; IMPLANTABLE CARDIOVERTER-DEFIBRILLATORS; SURGICAL REVISION; ICD LEAD; PERFORATION; COMPLICATIONS; RECOMMENDATIONS; FAILURE;
D O I
10.1016/j.hrthm.2008.10.030
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BACKGROUND Recent attention has been focused on the performance of high-vottage implantable cardioverter-defibrillator (ICD) Leads. Lead-related adverse events can be caused by several factors, including patient characteristics, concomitant therapies, implantation technique, access site, and lead design. Few studies have included the number of patients and the follow-up duration necessary to characterize accurately the Lead performance and the incidence of Lead-related adverse events. OBJECTIVE This study sought to define the incidence of lead-related adverse events including conductor fracture, insulation damage, perforation, and dislodgement of St. Jude Medical Riata ICD leads. METHODS Data from the Advancements in ICD Therapy, OPTIM Lead InsUlation Material (OPTIMUM), Resynchronization HemodYnamic Treatment for Heart Failure Management (RHYTHM), and Post Approval Study (PAS) registries provided data from 7,497 patients with Riata leads. Patients were followed up for a median of 22 months. Adverse events, defined as those that required Lead revision, extraction, or replacement, were recorded. RESULTS Conductor fracture occurred in 0.09%, insulation damage in 0.13%, dislodgement in 0.88%, and perforation in 0.31% of patients, respectively. CONCLUSION During more than 22 months of follow-up of over 7,000 Riata leads in 4 prospective registries and clinical trials, the incidence of tead-related adverse events including conductor fracture, insulation damage, dislodgement, and perforation necessitating lead revision, extraction, or replacement is rare.
引用
收藏
页码:204 / 209
页数:6
相关论文
共 50 条
  • [31] Industry viewpoint: St. Jude Medical: Pacemakers, ICDs and MRI
    Levine, PA
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2005, 28 (04): : 266 - 267
  • [32] St. Jude medical trifecta ™ aortic bioprosthesis - the training perspective
    Sepehripour, A. H.
    McCormack, D. J.
    Lall, K. S.
    BRITISH JOURNAL OF SURGERY, 2013, 100 : 165 - 165
  • [33] Stored electrograms in pacemakers and ICDs from St. Jude Medical
    Israel C.W.
    Herzschrittmachertherapie + Elektrophysiologie, 2010, 21 (1) : 53 - 70
  • [34] Subacute Perforation of the St. Jude Riata Implantable Cardioverter-Defibrillator Lead: A Report of Two Pediatric Cases
    Ben Morrison, T.
    Ackerman, Michael J.
    Rea, Robert F.
    PEDIATRIC CARDIOLOGY, 2009, 30 (06) : 834 - 836
  • [35] Hemodynamics and early clinical performance of the St. Jude Medical Regent mechanical aortic valve - Discussion
    Fullerton, D
    Emery
    Laas, J
    ANNALS OF THORACIC SURGERY, 2002, 74 (06): : 2009 - 2009
  • [36] The new St. Jude Medical Regent™ mechanical heart valve:: Laboratory measurements of hydrodynamic performance
    Walker, DK
    Brendzel, AM
    Scotten, LN
    JOURNAL OF HEART VALVE DISEASE, 1999, 8 (06): : 687 - 696
  • [37] Long-term performance of the St Jude Riata 1580–1582 ICD lead family
    S. D. A. Valk
    D. A. M. J. Theuns
    L. Jordaens
    Netherlands Heart Journal, 2013, 21 : 127 - 134
  • [38] Hemodynamic study of the elliptic St. Jude Medical valve: A computational study
    Jahandardoost, Mehdi
    Fradet, Guy
    Mohammadi, Hadi
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2016, 230 (02) : 85 - 96
  • [39] Long-term experience with the St. Jude Medical valve prosthesis
    Zellner, JL
    Kratz, JM
    Crumbley, AJ
    Stroud, MR
    Bradley, SM
    Sade, RM
    Crawford, FA
    ANNALS OF THORACIC SURGERY, 1999, 68 (04): : 1210 - 1218
  • [40] Early calcific degeneration of the St. Jude Medical Epic aortic bioprosthesis
    Izzat, Mohammad Bashar
    Sabbagh, Nada
    Aijasem, Hazem
    CLINICAL CASE REPORTS, 2020, 8 (02): : 387 - 388