Performance of the St. Jude Medical Riata Leads

被引:47
|
作者
Epstein, Andrew E. [1 ]
Baker, James H. [2 ]
Beau, Scott L. [3 ]
Deering, Thomas F. [4 ]
Greenberg, Steven M. [5 ]
Goldman, Daniel S. [6 ]
机构
[1] Univ Alabama Birmingham, Dept Med, Div Cardiovasc Dis, Birmingham, AL 35294 USA
[2] St Thomas Hosp, Nashville, TN USA
[3] Arkansas Heart Hosp, Little Rock, AR USA
[4] Piedmont Hosp, Atlanta, GA USA
[5] St Francis Hosp, Roslyn, NY USA
[6] Ctr Cardiol, Delray Beach, FL USA
关键词
Implantable cardioverter-defibrillator; Lead complications; Perforation; IMPLANTABLE CARDIOVERTER-DEFIBRILLATORS; SURGICAL REVISION; ICD LEAD; PERFORATION; COMPLICATIONS; RECOMMENDATIONS; FAILURE;
D O I
10.1016/j.hrthm.2008.10.030
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BACKGROUND Recent attention has been focused on the performance of high-vottage implantable cardioverter-defibrillator (ICD) Leads. Lead-related adverse events can be caused by several factors, including patient characteristics, concomitant therapies, implantation technique, access site, and lead design. Few studies have included the number of patients and the follow-up duration necessary to characterize accurately the Lead performance and the incidence of Lead-related adverse events. OBJECTIVE This study sought to define the incidence of lead-related adverse events including conductor fracture, insulation damage, perforation, and dislodgement of St. Jude Medical Riata ICD leads. METHODS Data from the Advancements in ICD Therapy, OPTIM Lead InsUlation Material (OPTIMUM), Resynchronization HemodYnamic Treatment for Heart Failure Management (RHYTHM), and Post Approval Study (PAS) registries provided data from 7,497 patients with Riata leads. Patients were followed up for a median of 22 months. Adverse events, defined as those that required Lead revision, extraction, or replacement, were recorded. RESULTS Conductor fracture occurred in 0.09%, insulation damage in 0.13%, dislodgement in 0.88%, and perforation in 0.31% of patients, respectively. CONCLUSION During more than 22 months of follow-up of over 7,000 Riata leads in 4 prospective registries and clinical trials, the incidence of tead-related adverse events including conductor fracture, insulation damage, dislodgement, and perforation necessitating lead revision, extraction, or replacement is rare.
引用
收藏
页码:204 / 209
页数:6
相关论文
共 50 条
  • [21] A Candle for St. Jude
    Willis, Katherine Tappert
    LIBRARY JOURNAL, 1948, 73 (12) : 944 - 944
  • [22] Subacute Perforation of the St. Jude Riata Implantable Cardioverter–Defibrillator Lead: A Report of Two Pediatric Cases
    T. Ben Morrison
    Michael J. Ackerman
    Robert F. Rea
    Pediatric Cardiology, 2009, 30 : 834 - 836
  • [23] The Gospel of St. Jude
    Plisch, Uwe-Karsten
    ZEITSCHRIFT FUR ANTIKES CHRISTENTUM-JOURNAL OF ANCIENT CHRISTIANITY, 2006, 10 (01): : 5 - 14
  • [24] Hemodynamics and early clinical performance of the St. Jude Medical Regent mechanical aortic valve
    Bach, DS
    Sakwa, MP
    Goldbach, M
    Petracek, MR
    Emery, RW
    Mohr, FW
    ANNALS OF THORACIC SURGERY, 2002, 74 (06): : 2003 - 2009
  • [25] Detection and discrimination of tachycardia in ICDs manufactured by St. Jude Medical; [Erkennung und Klassifizierung von Tachykardien bei ICDs von St. Jude Medical]
    Zdarek J.
    Israel C.W.
    Herzschrittmachertherapie + Elektrophysiologie, 2016, 27 (3) : 226 - 239
  • [26] Acute mechanical valve thrombosis of the St. Jude medical prosthesis
    Katircioglu, SF
    Ulus, AT
    Yamak, B
    Özsöyler, I
    Birincioglu, L
    Tasdemir, O
    JOURNAL OF CARDIAC SURGERY, 1999, 14 (03) : 164 - 168
  • [27] Electrical abnormalities with St. Jude Medical/Abbott pacing leads: Let's not call it lead failure yet
    Joglar, Jose A.
    HEART RHYTHM, 2021, 18 (12) : 2070 - 2071
  • [28] Silzone™-coated St. Jude Medical valve:: A safe valve
    Houël, R
    Kirsch, M
    Hillion, ML
    Loisance, D
    JOURNAL OF HEART VALVE DISEASE, 2001, 10 (06): : 724 - 727
  • [29] Integrity failure in Riata and Riata ST defibrillator leads: An ongoing problem
    Madeira, Joao
    Parreira, Leonor
    Amador, Pedro
    Soares, Luis
    REVISTA PORTUGUESA DE CARDIOLOGIA, 2013, 32 (10) : 823 - 826
  • [30] Implantable loop recorder of the Confirm family (St. Jude Medical)
    Pujdak K.
    Herzschrittmachertherapie + Elektrophysiologie, 2016, 27 (4) : 341 - 344