Performance of the St. Jude Medical Riata Leads

被引:47
|
作者
Epstein, Andrew E. [1 ]
Baker, James H. [2 ]
Beau, Scott L. [3 ]
Deering, Thomas F. [4 ]
Greenberg, Steven M. [5 ]
Goldman, Daniel S. [6 ]
机构
[1] Univ Alabama Birmingham, Dept Med, Div Cardiovasc Dis, Birmingham, AL 35294 USA
[2] St Thomas Hosp, Nashville, TN USA
[3] Arkansas Heart Hosp, Little Rock, AR USA
[4] Piedmont Hosp, Atlanta, GA USA
[5] St Francis Hosp, Roslyn, NY USA
[6] Ctr Cardiol, Delray Beach, FL USA
关键词
Implantable cardioverter-defibrillator; Lead complications; Perforation; IMPLANTABLE CARDIOVERTER-DEFIBRILLATORS; SURGICAL REVISION; ICD LEAD; PERFORATION; COMPLICATIONS; RECOMMENDATIONS; FAILURE;
D O I
10.1016/j.hrthm.2008.10.030
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BACKGROUND Recent attention has been focused on the performance of high-vottage implantable cardioverter-defibrillator (ICD) Leads. Lead-related adverse events can be caused by several factors, including patient characteristics, concomitant therapies, implantation technique, access site, and lead design. Few studies have included the number of patients and the follow-up duration necessary to characterize accurately the Lead performance and the incidence of Lead-related adverse events. OBJECTIVE This study sought to define the incidence of lead-related adverse events including conductor fracture, insulation damage, perforation, and dislodgement of St. Jude Medical Riata ICD leads. METHODS Data from the Advancements in ICD Therapy, OPTIM Lead InsUlation Material (OPTIMUM), Resynchronization HemodYnamic Treatment for Heart Failure Management (RHYTHM), and Post Approval Study (PAS) registries provided data from 7,497 patients with Riata leads. Patients were followed up for a median of 22 months. Adverse events, defined as those that required Lead revision, extraction, or replacement, were recorded. RESULTS Conductor fracture occurred in 0.09%, insulation damage in 0.13%, dislodgement in 0.88%, and perforation in 0.31% of patients, respectively. CONCLUSION During more than 22 months of follow-up of over 7,000 Riata leads in 4 prospective registries and clinical trials, the incidence of tead-related adverse events including conductor fracture, insulation damage, dislodgement, and perforation necessitating lead revision, extraction, or replacement is rare.
引用
收藏
页码:204 / 209
页数:6
相关论文
共 50 条
  • [1] National Consultant in Cardiology Experts' Group Guidelines on dealing with patients implanted with some St. Jude Medical Riata and Riata ST leads
    Mitkowski, Przemyslaw
    Grabowski, Marcin
    Kowalski, Oskar
    Kutarski, Andrzej
    Mojkowski, Wlodzimierz
    Przybylski, Andrzej
    Sterlinski, Maciej
    Trusz-Gluza, Maria
    Opolski, Grzegorz
    KARDIOLOGIA POLSKA, 2014, 72 (06) : 576 - 582
  • [2] Concerns about the Riata® ST (St. Jude medical) ICD lead
    Vlay, Stephen C.
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2008, 31 (01): : 1 - 2
  • [3] Clinical Performance of the St. Jude Medical Riata Defibrillation Lead in a Large Patient Population
    Porterfield, James G.
    Porterfield, Linda M.
    Kuck, Karl H.
    Corbisiero, Raffaele
    Greenberg, Steven M.
    Hindricks, Gerhard
    Wazni, Oussama
    Beau, Scott L.
    Herre, John M.
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2010, 21 (05) : 551 - 556
  • [4] Extraction of St. Jude Riata Leads: Clinical Experience and Outcomes at a High Volume Extraction Center
    Sager, Solomon J.
    Carrillo, Roger G.
    CIRCULATION, 2012, 126 (21)
  • [5] Low Prevalence of Electrical Failure in St. Jude Riata Family ICD Leads with Externalized Cables
    Kumar, Vineet
    Laminack, Stacy
    Yamada, Takumi
    Kay, Neal
    Plumb, Vance
    CIRCULATION, 2012, 126 (21)
  • [6] St. Jude Medical introduces ultrathin ICD leads to Japan
    不详
    ADVANCED MATERIALS & PROCESSES, 2006, 164 (01): : 75 - 75
  • [7] High prevalence of insulation failure with externalized cables in St. Jude Medical Riata family ICD leads: Fluoroscopic grading scale and correlation to extracted leads
    Parvathaneni, Sunthosh V.
    Ellis, Christopher R.
    Rottman, Jeffrey N.
    HEART RHYTHM, 2012, 9 (08) : 1218 - 1224
  • [8] Thrombogenicity and Reoperation of the St. Jude Medical Silzone® valve:: A comparison with the conventional St. Jude medical valve
    Stalenhoef, JE
    Mellema, EC
    Veeger, NJGM
    Ebels, T
    JOURNAL OF HEART VALVE DISEASE, 2003, 12 (05): : 635 - 639
  • [9] Comparison of lead failure manifestation of Biotronik Linox with St. Jude Medical Riata and Medtronic Sprint Fidelis lead
    Anna Lam
    Stefan Buehler
    Eleni Goulouti
    Romy Sweda
    Andreas Haeberlin
    Argelia Medeiros-Domingo
    Helge Servatius
    Jens Seiler
    Samuel Baldinger
    Fabian Noti
    Hildegard Tanner
    Laurent Roten
    Journal of Interventional Cardiac Electrophysiology, 2019, 54 : 161 - 170
  • [10] Hemodynamic performance of the St. Jude Medical Hemodynamic Plus valve
    Second Department of Surgery, Kurume University Hospital, Kurume, Japan
    不详
    ARTIF. ORGANS, 8 (916-921):