A longitudinal study of electrical stimulation levels and electrode impedance in children using the Clarion cochlear implant

被引:36
|
作者
Henkin, Y [1 ]
Kaplan-Neeman, R
Kronenberg, J
Migirov, L
Hildesheimer, M
Muchnik, C
机构
[1] Tel Aviv Univ, Sackler Fac Med, Chaim Sheba Med Ctr, Dept Commun Disorders, IL-52621 Tel Hashomer, Israel
[2] Chaim Sheba Med Ctr, Speech & Hearing Ctr, Tel Hashomer, Israel
[3] Chaim Sheba Med Ctr, Dept Otolaryngol Head & Neck Surg, Tel Hashomer, Israel
关键词
Clarion cochlear implant; children; threshold (T) levels; comfortable (M) levels; dynamic range; electrode impedance;
D O I
10.1080/00016480500443391
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
Conclusions. Electrical stimulation levels and electrode impedance values (EIVs) in children using the Clarion cochlear implant (CI) programmed with CIS strategy stabilized after 3 months of implant use. The data presented here may be useful as a general guideline for the programming of infants and young children and may further be of help for the identification of patients who fall outside the "average'' range. Objectives. The purpose of the present study was to evaluate changes in electrical stimulation levels, i.e. threshold (T) levels, comfortable (M) levels, dynamic range (DR), and EIVs during the first 18 months of implant use, in children using the Clarion CI. Materials and methods. The maps of 18 pre-lingual children (mean age at implantation 4.2 years; range 1-8), using the Enhanced Bipolar 1.2 or Bipolar standard electrode with the S-Series speech processor programmed with CIS strategy, were examined at five time points: connection, and 3, 6, 12, and 18 months post-initial stimulation. T levels, M levels, DR and EIVs were analyzed according to four cochlear segments: apical, apical-medial, medial-basal, and basal. Results. During the first 3 months of implant use T levels increased to some extent, whereas M levels and DR increased significantly. From 3 months and through the entire follow-up, T and M levels as well as DR were stable. EIVs of current carrying electrodes decreased significantly from connection to the 3-month visit; thereafter a stabilization of values was evident. Electrical stimulation levels and EIVs did not differ among the cochlear segments during the entire follow-up.
引用
收藏
页码:581 / 586
页数:6
相关论文
共 50 条
  • [41] Electrode Impedance Fluctuations and Sudden Decline in Cochlear-Implant Benefit
    Pitt, Cache
    Nagaraj, Naveen K.
    Salisbury, Anna
    OTOLOGY & NEUROTOLOGY, 2022, 43 (09) : E1008 - E1012
  • [42] MEASURING THE IMPEDANCE OF THE ACTIVE ELECTRODE OF A SINGLE CHANNEL COCHLEAR IMPLANT INSITU
    HRUBY, J
    KLIER, E
    PICKA, J
    SEDLAK, S
    BETKA, J
    VALVODA, M
    JOURNAL OF BIOMEDICAL ENGINEERING, 1988, 10 (05): : 470 - 471
  • [43] Detection of Translocation of Cochlear Implant Electrode Arrays by Intracochlear Impedance Measurements
    Dong, Yu
    Briaire, Jeroen J.
    Siebrecht, Michael
    Stronks, H. Christiaan
    Frijns, Johan H. M.
    EAR AND HEARING, 2021, 42 (05): : 1397 - 1404
  • [44] Electropermeabilization of Adherent Cells with Cochlear Implant Electrical Stimulation in vitro
    Newbold, Carrie
    Farrington, Andrew
    Peters, Lawson
    Cowan, Robert
    Needham, Karina
    AUDIOLOGY AND NEURO-OTOLOGY, 2014, 19 (05) : 283 - 292
  • [45] Comparison of the electrical stimulation performance of four cochlear implant electrodes
    Choi, CTM
    Lai, WD
    Chen, YB
    IEEE TRANSACTIONS ON MAGNETICS, 2005, 41 (05) : 1920 - 1923
  • [46] STAPEDIUS REFLEX TO ELECTRICAL-STIMULATION IN A PATIENT WITH A COCHLEAR IMPLANT
    JERGER, J
    JENKINS, H
    FIFER, R
    MECKLENBURG, D
    ANNALS OF OTOLOGY RHINOLOGY AND LARYNGOLOGY, 1986, 95 (02): : 151 - 157
  • [47] SUPPRESSION OF TINNITUS BY ELECTRICAL-STIMULATION IN COCHLEAR IMPLANT PATIENTS
    BATTMER, RD
    HEERMANN, R
    LASZIG, R
    HNO, 1989, 37 (04) : 148 - 152
  • [48] Modelling of the electrical stimulation delivered by the DIGISONIC Multichannel cochlear implant
    Gallego, S.
    Luu, B.L.
    Berger-Vachon, Ch.
    Advances in Modeling and Analysis B, 1998, 39 (01): : 39 - 53
  • [49] Auditory cortex activation by direct cochlear implant electrical stimulation
    Lazeyras, F
    Boëx, C
    Sigrist, A
    Cosendai, G
    Terrier, F
    Pelizzone, M
    NEUROIMAGE, 2001, 13 (06) : S905 - S905
  • [50] The phenomenon of nystagmus upon electrical stimulation in a cochlear implant patient
    Wong, ECM
    See, HKP
    Yu, HC
    UPDATES IN COCHLEAR IMPLANTATION, 2000, 57 : 189 - 191