Comparison of an Implantable Middle Ear Microphone and Conventional External Microphone for Cochlear Implants: A Clinical Feasibility Study

被引:1
|
作者
Craddock, Louise C. [1 ]
Hodson, James [1 ,2 ]
Gosling, Amy [1 ]
Cooper, Stacey [1 ]
Morse, Robert P. [3 ]
Begg, Philip [1 ,4 ]
Prokopiou, Andreas [3 ]
Irving, Richard M. [1 ]
机构
[1] Univ Hosp Birmingham NHS Fdn Trust, Birmingham, England
[2] Univ Hosp Birmingham NHS Fdn Trust, Inst Translat Med, Health Data Sci Team, Res Dev & Innovat, Birmingham, England
[3] Univ Warwick, Sch Engn, Coventry, England
[4] Univ Kentucky, Lexington, KY USA
关键词
Audiology; Cochlear implants; Implantable technology; Research trial; Surgical trial;
D O I
10.1097/MAO.0000000000003713
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
ObjectivesAll commercially available cochlear implant (CI) systems use an external microphone and sound processor; however, external equipment carries lifestyle limitations. Although totally implantable devices using subcutaneous microphones have been developed, these are compromised by problems with soft tissue sound attenuation, feedback, and intrusive body noise. This in vivo pilot study evaluates a middle ear microphone (MEM) that aims to overcome these issues and compares hearing performance with that of an external CI microphone.DesignSix adult participants with an existing CI were implanted with a temporary MEM in the contralateral ear. Signals from the MEM were routed via a percutaneous plug and cable to the CI sound processor. Testing was performed in the CI microphone and MEM conditions using a range of audiometric assessments, which were repeated across four visits.ResultsPerformance of the MEM did not differ significantly from that of the CI on the assessments of Auditory Speech Sounds Evaluation loudness scaling at either 250 or 1000 Hz, or in the accuracy of repeating keywords presented at 70 dB. However, the MEM had significantly poorer aided sound-field thresholds, particularly at higher frequencies (>= 4000 Hz), and significantly poorer performance on Arthur Boothroyd words presented at 55 dB, compared with the CI.ConclusionIn this pilot study, the MEM showed comparable performance to that of an external CI microphone across some audiometric assessments. However, performance with the MEM was poorer than the CI in soft-level speech (55 dB) and at higher frequencies. As such, the benefits of MEM need to be considered against the compromises in hearing performance. However, with future development, MEM is a potentially promising technology.
引用
收藏
页码:1162 / 1169
页数:8
相关论文
共 43 条
  • [21] Creation of an incus recess for a middle-ear microphone using a drill or laser ablation: a comparison of equivalent noise level and middle ear transfer function
    Morse, Robert P.
    Mitchell-Innes, Alistair
    Prokopiou, Andreas N.
    Irving, Richard M.
    Begg, Philip A.
    EUROPEAN ARCHIVES OF OTO-RHINO-LARYNGOLOGY, 2023, 280 (02) : 661 - 669
  • [22] An Implantable Piezofilm Middle Ear Microphone: Performance in Human Cadaveric Temporal Bones (Jan, 10.1007/s10162-024-00927-4, 2023)
    Zhang, John Z.
    Graf, Lukas
    Banerjee, Annesya
    Yeiser, Aaron
    McHugh, Christopher I.
    Kymissis, Ioannis
    Lang, Jeffrey H.
    Olson, Elizabeth S.
    Nakajima, Hideko Heidi
    JARO-JOURNAL OF THE ASSOCIATION FOR RESEARCH IN OTOLARYNGOLOGY, 2024, 25 (01): : 89 - 89
  • [23] Comparison of hearing performance in patients with borderline indication for active middle ear or cochlear implants: clinical outcomes to guide preoperative counseling and decision making
    Herr, Constanze
    Stoever, Timo
    Baumann, Uwe
    Weissgerber, Tobias
    EUROPEAN ARCHIVES OF OTO-RHINO-LARYNGOLOGY, 2024, 281 (07) : 3569 - 3575
  • [24] Fully Implantable Active Middle Ear Implants After Subtotal Petrosectomy With Fat Obliteration: Preliminary Clinical Results
    Lasurashvili, Nikoloz
    Lailach, Susen
    Seidler, Hannes
    Bornitz, Matthias
    Neudert, Marcus
    Zahnert, Thomas
    OTOLOGY & NEUROTOLOGY, 2020, 41 (07) : E912 - E920
  • [25] Sheep as a Large Animal Model for Middle and Inner Ear Implantable Hearing Devices: A Feasibility Study in Cadavers
    Schnabl, Johannes
    Glueckert, Rudolf
    Feuchtner, Gudrun
    Recheis, Wolfgang
    Potrusil, Thomas
    Kuhn, Volker
    Wolf-Magele, Astrid
    Riechelmann, Herbert
    Sprinzl, Georg M.
    OTOLOGY & NEUROTOLOGY, 2012, 33 (03) : 481 - 489
  • [26] Feasibility study of round window stimulation with the MET-V implantable middle ear device: A temporal bone study
    Holland, N. Julian
    Koka, Kanthaiah
    Lupo, J. Eric
    Tollin, Daniel J.
    Jenkins, Herman A.
    HEARING RESEARCH, 2010, 263 (1-2) : 251 - 251
  • [27] A new implantable middle ear hearing device for mixed hearing loss: A feasibility study in human temporal bones
    Huber, Alexander M.
    Ball, Geoffrey R.
    Veraguth, Dorothe
    Dillier, Norbert
    Bodmer, Daniel
    Sequeira, Damien
    OTOLOGY & NEUROTOLOGY, 2006, 27 (08) : 1104 - 1109
  • [28] UNILATERAL DUPLICATION OF THE CEREBELLAR HEMISPHERE AND INTERNAL, MIDDLE, AND EXTERNAL EAR - A CLINICAL CASE-STUDY
    JACKSON, JM
    SADOVE, AM
    WEAVER, DD
    EDWARDS, MK
    BULL, MJ
    PLASTIC AND RECONSTRUCTIVE SURGERY, 1990, 86 (03) : 550 - 553
  • [29] EPITHELIOMAS OF THE MIDDLE-EAR AND THE EXTERNAL AUDITORY-CANAL - A RADIO-TOMOGRAPHIC AND CLINICAL-STUDY
    HOTTON, F
    BOLLAERT, A
    KLEINER, S
    TAELMAN, D
    DEGANDT, JB
    JOURNAL BELGE DE RADIOLOGIE, 1980, 63 (2-3): : 313 - 320
  • [30] Phase III clinical trial results with the Vibrant Soundbridge implantable middle ear hearing device: A prospective controlled multicenter study
    Luetje, CM
    Brackman, D
    Balkany, TJ
    Maw, J
    Baker, RS
    Kelsall, D
    Backous, D
    Miyamoto, R
    Parisier, S
    Arts, A
    OTOLARYNGOLOGY-HEAD AND NECK SURGERY, 2002, 126 (02) : 97 - 107