Creation of an incus recess for a middle-ear microphone using a drill or laser ablation: a comparison of equivalent noise level and middle ear transfer function

被引:0
|
作者
Morse, Robert P. [1 ]
Mitchell-Innes, Alistair [2 ]
Prokopiou, Andreas N. [1 ]
Irving, Richard M. [1 ,3 ]
Begg, Philip A. [3 ]
机构
[1] Univ Warwick, Sch Engn, Coventry CV4 7AL, W Midlands, England
[2] Univ Hosp Birmingham, ENT Dept, Birmingham B15 2GW, W Midlands, England
[3] Inst Translat Med, Birmingham B15 2TH, W Midlands, England
基金
美国国家卫生研究院;
关键词
Laser Doppler vibrometry; Noise-induced hearing loss; Middle ear microphone; Equivalent noise level; Middle ear transfer function; STAPES SURGERY;
D O I
10.1007/s00405-022-07532-2
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
Purpose Studies have assessed the trauma and change in hearing function from the use of otological drills on the ossicular chain, but not the effects of partial laser ablation of the incus. A study of the effectiveness of a novel middle-ear microphone for a cochlear implant, which required an incus recess for the microphone balltip, provided an opportunity to compare methods and inform a feasibility study of the microphone with patients. Methods We used laser Doppler vibrometry with an insert earphone and probe microphone in 23 ears from 14 fresh-frozen cadavers to measure the equivalent noise level at the tympanic membrane that would have led to the same stapes velocity as the creation of the incus recess. Results Drilling on the incus with a diamond burr created peak noise levels equivalent to 125.1-155.0 dB SPL at the tympanic membrane, whilst using the laser generated equivalent noise levels barely above the baseline level. The change in middle ear transfer function following drilling showed greater variability at high frequencies, but the change was not statistically significant in the three frequency bands tested. Conclusions Whilst drilling resulted in substantially higher equivalent noise, we considered that the recess created by laser ablation was more likely to lead to movement of the microphone balltip, and therefore decrease performance or result in malfunction over time. For patients with greatly reduced residual hearing, the greater consistency from drilling the incus recess may outweigh the potential benefits of hearing preservation with laser ablation.
引用
收藏
页码:661 / 669
页数:9
相关论文
共 25 条
  • [1] Creation of an incus recess for a middle-ear microphone using a drill or laser ablation: a comparison of equivalent noise level and middle ear transfer function
    Robert P. Morse
    Alistair Mitchell-Innes
    Andreas N. Prokopiou
    Richard M. Irving
    Philip A. Begg
    [J]. European Archives of Oto-Rhino-Laryngology, 2023, 280 : 661 - 669
  • [2] Middle-ear velocity transfer function, cochlear input immittance, and middle-ear efficiency in chinchilla
    Ravicz, Michael E.
    Rosowski, John J.
    [J]. JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2013, 134 (04): : 2852 - 2865
  • [3] MIDDLE-EAR TRANSFER-FUNCTION OF ANESTHETIZED MONGOLIAN GERBIL
    SCHMIEDT, RA
    SMITH, RL
    [J]. JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1974, 56 : S4 - S4
  • [4] A prediction of the minke whale (Balaenoptera acutorostrata) middle-ear transfer function
    Tubelli, Andrew A.
    Zosuls, Aleks
    Ketten, Darlene R.
    Yamato, Maya
    Mountain, David C.
    [J]. JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2012, 132 (05): : 3263 - 3272
  • [5] Human middle-ear sound transfer function and cochlear input impedance
    Aibara, R
    Welsh, JT
    Puria, S
    Goode, RL
    [J]. HEARING RESEARCH, 2001, 152 (1-2) : 100 - 109
  • [6] Coupling of an active middle-ear implant to the long process of the incus using an elastic clip attachment
    Schraven, Sebastian P.
    Mlynski, Robert
    Dalhoff, Ernst
    Heyd, Andreas
    Wildenstein, Daniela
    Rak, Kristen
    Radeloff, Andreas
    Hagen, Rudolf
    Gummer, Anthony W.
    [J]. HEARING RESEARCH, 2016, 340 : 179 - 184
  • [7] Human middle-ear nonlinearity measurements using laser Doppler vibrometry
    Gladine, Kilian
    Muyshondt, Pieter G. G.
    Dirckx, Joris J. J.
    [J]. OPTICS AND LASERS IN ENGINEERING, 2017, 99 : 98 - 102
  • [8] Effects of preloads on middle-ear transfer function and acoustic reflex in ossiculoplasty with PORP
    Schar, Merlin
    Dobrev, Ivo
    Roosli, Christof
    Huber, Alexander M.
    Sim, Jae Hoon
    [J]. HEARING RESEARCH, 2023, 430
  • [9] Middle-ear function in the chinchilla: Circuit models and comparison with other mammalian species
    Lemons, Charlsie
    Meaud, Julien
    [J]. JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2016, 140 (04): : 2735 - 2753
  • [10] Reverse middle-ear transfer function in the guinea pig measured with cubic difference tones
    Magnan, P
    Avan, P
    Dancer, A
    Smurzynski, J
    Probst, R
    [J]. HEARING RESEARCH, 1997, 107 (1-2) : 41 - 45