A Low-Noise Charge Sensitive Preamplifier for Ge Spectroscopy operating at Cryogenic Temperature in the GERDA Experiment

被引:0
|
作者
Riboldi, Stefano [1 ,2 ]
Zocca, Francesca [1 ,2 ]
Pullia, Alberto [1 ,2 ]
Barnabe-Heider, Marik [4 ]
Budjas, Dusan [4 ]
D'Andragora, Alessio [3 ]
Cattadori, Carla [2 ]
机构
[1] Univ Milan, Dept Phys, I-20133 Milan, Italy
[2] INFN, I-17323 Milan, Italy
[3] INFN, LNGS, Milan, Italy
[4] MPIK, Heidelberg, Germany
关键词
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Front-end electronics for ionizing radiation detectors signal read-out must sometimes operate at cryogenic temperatures. The classical solution based on cold input transistor and subsequent warm electronics is sometimes not suitable due to the physical constraints of the experimental setup, so the entire front-end circuit has to operate at cryogenic temperature. The proposed multi-channel charge sensitive preamplifiers ("CC2") have been tested in the framework of the GERDA experiment (GERmanium Detector Array), in connection to High-Purity Germanium (HPGe) detectors and are able to operate at cryogenic (liquid Nitrogen) temperature. The results obtained during preliminary tests with an encapsulated Germanium detector and subsequently with naked Germanium detectors are all within the specifications of the GERDA experiment for Phase-I.
引用
收藏
页码:1386 / 1388
页数:3
相关论文
共 50 条
  • [42] A low noise CMOS charge sensitive preamplifier for MEMS capacitive accelerometer readout
    Chen, Jianghua
    Ni, Xuewen
    Mo, Bangxian
    ASICON 2007: 2007 7TH INTERNATIONAL CONFERENCE ON ASIC, VOLS 1 AND 2, PROCEEDINGS, 2007, : 490 - 493
  • [43] WIDE BAND LOW NOISE VOLTAGE SENSITIVE PREAMPLIFIER WITH TEMPERATURE COMPENSATION
    KOPP, MK
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1971, 42 (05): : 714 - &
  • [44] LOW NOISE CHARGE-SENSITIVE PREAMPLIFIER EMPLOYING FIELD EFFECT TRANSISTORS
    PODYACHE.VN
    SHATUNOVSKII, RM
    INSTRUMENTS AND EXPERIMENTAL TECHNIQUES-USSR, 1970, (04): : 1042 - +
  • [45] Cryogenic Front-End ASICs for Low-Noise Readout of Charge Signals
    Mukim, Prashansa
    Carini, Gabriella A.
    Chen, Hucheng
    Deptuch, Grzegorz W.
    Gao, Shanshan
    De Geronimo, Gianluigi
    Mandal, Soumyajit
    Manyam, Venkata Narasimha
    Radeka, Veljko
    Rescia, Sergio
    IEEE TRANSACTIONS ON CIRCUITS AND SYSTEMS I-REGULAR PAPERS, 2024,
  • [46] A RAD HARD, LOW-NOISE, HIGH-SPEED, BIFET CHARGE PREAMPLIFIER FOR THE SUPERCONDUCTING SUPERCOLLIDER
    LING, KY
    VANPETEGHEM, PM
    LEE, SY
    LIU, HH
    DIBITONTO, D
    PROCEEDINGS OF THE 1989 BIPOLAR CIRCUITS AND TECHNOLOGY MEETING, 1989, : 124 - 127
  • [47] Low-noise variable-temperature preamplifier for piezoelectric tuning fork force sensors
    Patil, NG
    Levy, J
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2002, 73 (02): : 486 - 487
  • [48] LOW-NOISE GALLIUM-ARSENIDE CHARGE-SENSITIVE PREAMPLIFIERS FOR LOW-TEMPERATURE PARTICLE DETECTORS
    ALESSANDRELLO, A
    BROFFERIO, C
    CAMIN, DV
    GIULIANI, A
    PESSINA, G
    PREVITALI, E
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1990, 37 (03) : 1242 - 1247
  • [49] Low-noise rotating sample holder for ultrafast transient spectroscopy at cryogenic temperatures
    Fanciulli, R.
    Cerjak, I.
    Herek, J. L.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2007, 78 (05):
  • [50] A hybrid low-noise charge sensitive amplifier with fast discharge mechanism
    Bassini, R
    Boiano, C
    Brambilla, S
    Pullia, A
    2001 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORDS, VOLS 1-4, 2002, : 1014 - 1017