Validation of PRIMO Monte Carlo Model of Clinac®iX 6MV Photon Beam

被引:9
|
作者
Sarin, B. [1 ,2 ]
Bindhu, B. [1 ]
Saju, B. [2 ]
Nair, Raguram K. [2 ]
机构
[1] Noorul Islam Ctr Higher Educ, Dept Phys, Kanyakumari, Tamil Nadu, India
[2] Reg Canc Ctr, Div Radiat Phys, C Block,Room 3, Thiruvananthapuram 695011, Kerala, India
关键词
Clinac iX; Monte Carlo simulation; PENELOPE; PRIMO; DOSE-CALCULATION; DOSIMETRIC VALIDATION; SIMULATION; ALGORITHM; PENELOPE; CODE; XB;
D O I
10.4103/jmp.JMP_75_19
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: This study aims to model 6MV photon of Clinac (R) iX linear accelerator using PRIMO Monte Carlo (MC) code and to assess PRIMO as an independent MC-based dose verification and quality assurance tool. Materials and Methods: The modeling of Clinac (R) iX linear accelerator has been carried out by using PRIMO simulation software (Version 0.3.1.1681). The simulated beam parameters were compared against the measured beam data of the Clinac (R) iX machine. The PRIMO simulation model of Clinac (R) iX was also validated against Eclipse (R) Acuros XB dose calculations in the case of both homogenous and inhomogeneous mediums. The gamma analysis method with the acceptance criteria of 2%, 2 mm was used for the comparison of dose distributions. Results: Gamma analysis shows a minimum pass percentage of 99% for depth dose curves and 95.4% for beam profiles. The beam quality index and output factors and absolute point dose show good agreement with measurements. The validation of PRIMO dose calculations, in both homogeneous and inhomogeneous medium, against Acuros (R) XB shows a minimum gamma analysis pass rate of 99%. Conclusions: This study shows that the research software PRIMO can be used as a treatment planning system-independent quality assurance and dose verification tool in daily clinical practice. Further validation will be performed with different energies, complex multileaf collimators fields, and with dynamic treatment fields.
引用
收藏
页码:24 / 35
页数:12
相关论文
共 50 条
  • [31] IMRT treatment of prostate cancer using 6MV photon beam
    Huang, J
    Katz, A
    Haas, J
    Urrutia, T
    Lauritano, J
    MacMelville, W
    MEDICAL PHYSICS, 2002, 29 (06) : 1254 - 1254
  • [32] PERIPHERAL DOSIMETRY BY MONTE CARLO FOR 6MV AND CO-60 BEAMS
    Joosten, A.
    Levi, F.
    Baechler, S.
    Mirimanoff, R. O.
    Bochud, F.
    Moeckli, R.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2009, 185 (09) : 636 - 636
  • [33] Monte Carlo simulation of Varian Linac for 6 MV photon beam with BEAMnrc code
    Mohammed, Maged
    El Bardouni, T.
    Chakir, E.
    Boukhal, H.
    Saeed, M.
    Ahmed, Abdul-Aziz
    RADIATION PHYSICS AND CHEMISTRY, 2018, 144 : 69 - 75
  • [34] Monte Carlo study of the influence of a novel transmission detector on 6 MV photon beam
    Asuni, G.
    McCurdy, B.
    Jensen, M.
    MEDICAL PHYSICS, 2009, 36 (09) : 4309 - 4309
  • [35] Monte Carlo dose calculations for a 6-MV photon beam in a thorax phantom
    Farajollahi A.
    Mesbahi A.
    Radiation Medicine, 2006, 24 (4): : 269 - 276
  • [36] Monte Carlo model of the Elekta SLiplus accelerator:: validation of a new MLC component module in BEAM for a 6 MV beam
    Van de Walle, J
    Martens, C
    Reynaert, N
    Palmans, H
    Coghe, M
    De Neve, W
    De Wagter, C
    Thierens, H
    PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (03): : 371 - 385
  • [37] Development and Validation of a Monte Carlo Model for the Varian Clinac IX linear Accelerator in Electron Mode Using GATE
    Mutanga, T.
    Wang, Y.
    Wu, X.
    Nielsen, M.
    MEDICAL PHYSICS, 2018, 45 (06) : E699 - E699
  • [38] Out-of-field beam characteristics of a 6 MV photon beam: Results of a Monte Carlo study
    Atarod, M.
    Shokrani, P.
    Azarnoosh, A.
    APPLIED RADIATION AND ISOTOPES, 2013, 72 : 182 - 194
  • [39] Validation of Monte carlo Geant4 multithreading code for a 6 MV photon beam of varian linac on the grid computing
    Assalmi, Mustapha
    Diaf, El Yamani
    Mansour, Najim
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2020, 25 (06) : 1001 - 1010
  • [40] Monte Carlo Simulation of a 6MV Varian Truebeam Without Flattening Filter Linac
    Song, T.
    Zhou, L.
    Jiang, S.
    MEDICAL PHYSICS, 2012, 39 (06) : 3819 - 3819