Accuracy of Different Surgical Guide Designs for Static Computer-Assisted Implant Surgery: An In Vitro Study

被引:4
|
作者
Wu, Yu Tsung [1 ]
Papaspyridakos, Panos [1 ]
Kang, Kiho [1 ]
Finkelman, Matthew [2 ]
Kudara, Yukio
De Souza, Andre B. [1 ]
机构
[1] Tufts Univ, Sch Dent Med, Dept Prosthodont, Boston, MA 02111 USA
[2] Tufts Univ, Sch Dent Med, Dept Publ Hlth & Community Serv, Boston, MA USA
关键词
dental implants; digital workflow; guided implant surgery; accuracy; computer-aided; computed-manufactured; TECHNOLOGY APPLICATIONS; DENTAL PROSTHESES; COMPLICATIONS; DENTISTRY; PRECISION; SURVIVAL; RATES;
D O I
10.1563/aaid-joi-D-21-00055
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
The aims of this study were to evaluate the effect of (1) the different surgical guide designs and (2) implant placement location on the accuracy of fully guided implant placement in single edentulous sites using an in vitro study model. Forty-five partially edentulous models were scanned and divided into 3 groups: group 1, tooth-supported full-arch surgical guide; group 2, 3 different tooth-supported shortened surgical guides (SSGs); and group 3, tooth-supported full-arch surgical guide with a crossbar. All surgical guides were printed and used for fully guided implant placement. A total of 180 implants (60 per group) were placed, and scan bodies were positioned on all models, and postoperative surface scan files (STL) files were obtained. Superimposition of preoperative and postoperative STL files was performed, and the accuracy of implant position was evaluated. The interaction between group and implant location was statistically significant for angle, 3D offset at the base, and at the tip (P < .001). The post-hoc tests showed a statistically significantly higher deviation for group 2 compared to group 3 for all outcomes for implants #4 (P < .05) and #7 (P < .05). There was also a statistically significant difference in all outcomes between groups 1 and 3 for implant #7 (P < .05). All surgical guide designs presented satisfactory performance with clinically acceptable levels of deviation. However, SSGs presented higher accuracy for guided implant placement in a single -edentulous site, whereas a full-arch surgical guide with a crossbar presented superior outcomes when 2 or more guided implants were placed simultaneously.
引用
收藏
页码:351 / 357
页数:7
相关论文
共 50 条
  • [11] Does implant drill design influence the accuracy of dental implant placement using static computer-assisted implant surgery? An in vitro study
    Anna Takács
    Gyula Marada
    Kinga Turzó
    Ákos Nagy
    Orsolya Németh
    Eitan Mijiritsky
    Márton Kivovics
    Attila Mühl
    BMC Oral Health, 23
  • [12] Does implant drill design influence the accuracy of dental implant placement using static computer-assisted implant surgery? An in vitro study
    Takacs, Anna
    Marada, Gyula
    Turzo, Kinga
    Nagy, Akos
    Nemeth, Orsolya
    Mijiritsky, Eitan
    Kivovics, Marton
    Muhl, Attila
    BMC ORAL HEALTH, 2023, 23 (01)
  • [13] Accuracy of dynamic computer-assisted implant surgery in fully edentulous patients: An in vitro study
    Ruiz-Romero, Victor
    Jorba-Garcia, Adria
    Camps-Font, Octavi
    Figueiredo, Rui
    Valmaseda-Castellon, Eduard
    JOURNAL OF DENTISTRY, 2024, 149
  • [14] Positional accuracy during the sequence of static computer-assisted implant surgery in three alveolar ridge morphologies: An in vitro study
    Raabe, Clemens
    Abou-Ayash, Samir
    Yilmaz, Burak
    Surbek, Fiona Jennifer
    Chappuis, Vivianne
    Couso-Queiruga, Emilio
    JOURNAL OF PROSTHODONTICS-IMPLANT ESTHETIC AND RECONSTRUCTIVE DENTISTRY, 2025, 34 (01): : 78 - 85
  • [15] Influence of guide support on the accuracy of static Computer-Assisted Implant Surgery (sCAIS) in partially edentulous cases using a keyless guiding system: an in vitro study
    David Kasradze
    Ricardas Kubilius
    BMC Oral Health, 25 (1)
  • [16] Influence of implant macrodesign and insertion connection technology on the accuracy of static computer-assisted implant surgery
    El Kholy, Karim
    Ebenezer, Supriya
    Wittneben, Julia-Gabriela
    Lazarin, Rafael
    Rousson, Dominique
    Buser, Daniel
    CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2019, 21 (05) : 1073 - 1079
  • [17] Are Sleeves Necessary in Static Computer-Assisted Implant Surgery? A Comparative In Vitro Analysis
    Hang, Jenna
    Guentsch, Arndt
    CLINICAL ORAL IMPLANTS RESEARCH, 2025, 36 (01) : 117 - 126
  • [18] In vitro and in vivo accuracy of autonomous robotic vs. fully guided static computer-assisted implant surgery
    He, Jin
    Zhang, Qinmeng
    Wang, Xueting
    Fu, Mengdie
    Zhang, Hui
    Song, Luyao
    Pu, Rui
    Jiang, Zhiwei
    Yang, Guoli
    CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2024, 26 (02) : 385 - 401
  • [19] Comparison of the accuracy of dental implant placement using static and dynamic computer-assisted systems: an in vitro study
    Zhou, Miao
    Zhou, Hui
    Li, Shu-Yi
    Zhu, Yi-Bo
    Geng, Yuan-Ming
    JOURNAL OF STOMATOLOGY ORAL AND MAXILLOFACIAL SURGERY, 2021, 122 (04) : 343 - 348
  • [20] Comparison of the accuracy of implant placement using different drilling systems for static computer-assisted implant surgery: A simulation-based experimental study
    Sittikornpaiboon, Paknisa
    Arunjaroensuk, Sirida
    Kaboosaya, Boosana
    Subbalekha, Keskanya
    Mattheos, Nikos
    Pimkhaokham, Atiphan
    CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2021, 23 (04) : 635 - 643