Accuracy of 14 intraoral scanners for the All-on-4 treatment concept: a comparative in vitro study

被引:5
|
作者
Kaya, Gozde [1 ]
Bilmenoglu, Caglar [1 ,2 ]
机构
[1] Trakya Univ, Fac Dent, Dept Prosthodont, Edirne, Turkey
[2] Trakya Univ, Fac Dent, Dept Prosthodont, TR-22100 Edirne, Turkey
来源
JOURNAL OF ADVANCED PROSTHODONTICS | 2022年 / 14卷 / 06期
关键词
Intraoral scanner; Trueness; Precision; Accuracy; Digital impression; IMPRESSION; TRUENESS; IMPACT;
D O I
10.4047/jap.2022.14.6.388
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
PURPOSE. This in vitro study aimed to evaluate the accuracy of 14 different intraoral scanners for the All-on-4 treatment concept. MATERIALS AND METHODS. Four implants were placed in regions 13, 16, 23, and 26 of an edentulous maxillary model that was poured with scannable Type 4 gypsum to imitate the All-on-4 concept. The cast was scanned 10 times for each of 14 intraoral scanners (Primescan, iTero 2, iTero 5D, Virtuo Vivo, Trios 3, Trios 4, CS3600, CS3700, Emerald, Emerald S, Medit i500, BenQ BIS-I, Heron IOS, and Aadva IOS 100P) after the polyether ether ketone scanbody was placed. For the control group, the gypsum model was scanned 10 times with an industrial scanner. The first of the 10 virtual models obtained from the industrial model was chosen as the reference model. For trueness, the data of the 14 dental scanners were superimposed with the reference model; for precision, the data of all 14 scanners were superimposed within the groups. Statistical analyses were performed using the Kolmogorov-Smirnov, Shapiro-Wilks, and Dunn's tests.RESULTS. Primescan showed the highest trueness and precision values (P < .005), followed by the iTero 5D scanner (P < .005). CONCLUSION. Some of these digital scanners can be used to make impressions within the All-on-4 concept. However, the possibility of data loss due to artifacts, reflections, and the inability to combine the data should be considered. [J Adv Prosthodont 2022;14:388-98]
引用
收藏
页码:388 / 398
页数:11
相关论文
共 50 条
  • [1] Accuracy of four intraoral scanners in oral implantology: a comparative in vitro study
    Imburgia, Mario
    Logozzo, Silvia
    Hauschild, Uli
    Veronesi, Giovanni
    Mangano, Carlo
    Mangano, Francesco Guido
    [J]. BMC ORAL HEALTH, 2017, 17
  • [2] Accuracy of four intraoral scanners in oral implantology: a comparative in vitro study
    Mario Imburgia
    Silvia Logozzo
    Uli Hauschild
    Giovanni Veronesi
    Carlo Mangano
    Francesco Guido Mangano
    [J]. BMC Oral Health, 17
  • [3] Accuracy of Five Intraoral Scanners and Two Laboratory Scanners for a Complete Arch: A Comparative In Vitro Study
    Kang, Byung-hyun
    Son, Keunbada
    Lee, Kyu-bok
    [J]. APPLIED SCIENCES-BASEL, 2020, 10 (01):
  • [4] The time efficiency of intraoral scanners An in vitro comparative study
    Patzelt, Sebastian B. M.
    Lamprinos, Christos
    Stampf, Susanne
    Att, Wael
    [J]. JOURNAL OF THE AMERICAN DENTAL ASSOCIATION, 2014, 145 (06): : 542 - 551
  • [5] Treatment of the edentulous mandible by the All-on-4 concept
    Quantius, Bernd
    [J]. IMPLANTOLOGIE, 2012, 20 (01): : 33 - 42
  • [6] Accuracy evaluation of two different intraoral scanners in implant prosthodontics. A comparative in vitro study
    Verniani, G.
    Casucci, A.
    Nosrati, N.
    D'Arienzo, L. F.
    Val, M.
    Cagidiaco, E. Ferrari
    [J]. JOURNAL OF OSSEOINTEGRATION, 2024, 16 (01) : 61 - 64
  • [7] Comparative Analysis of Four Different Intraoral Scanners: An In Vitro Study
    Ciocan, Lucian Toma
    Vasilescu, Vlad Gabriel
    Rauta, Sabina-Ana
    Pantea, Mihaela
    Pituru, Silviu-Mirel
    Imre, Marina
    [J]. DIAGNOSTICS, 2024, 14 (13)
  • [8] Trueness and Precision of Two Intraoral Scanners: A Comparative In Vitro Study
    Rotar, Raul Nicolae
    Jivanescu, Anca
    Ille, Codruta
    Podariu, Angela Codruta
    Jumanca, Daniela Elisabeta
    Matichescu, Ana-Maria
    Balean, Octavia
    Rusu, Laura Cristina
    [J]. SCANNING, 2019, 2019
  • [9] Complete-Arch Accuracy of Four Intraoral Scanners: An In Vitro Study
    Celeghin, Giordano
    Franceschetti, Giulio
    Mobilio, Nicola
    Fasiol, Alberto
    Catapano, Santo
    Corsalini, Massimo
    Grande, Francesco
    [J]. HEALTHCARE, 2021, 9 (03)
  • [10] Influence of Liquid on the Tooth Surface on the Accuracy of Intraoral Scanners: An In Vitro Study
    Chen, Yuming
    Zhai, Zhihao
    Li, Hefei
    Yamada, Shuhei
    Matsuoka, Takashi
    Ono, Shinji
    Nakano, Tamaki
    [J]. JOURNAL OF PROSTHODONTICS-IMPLANT ESTHETIC AND RECONSTRUCTIVE DENTISTRY, 2022, 31 (01): : 59 - 64