A new optical method for determining deformation and volumetric wear in retrieved polyethylene acetabular cups

被引:2
|
作者
Thomsen, M
Burandt, C
Görtz, A
Dathe, H
Meesenburg, DK
Spiering, S
Görtz, W
Näger, H
机构
[1] Univ Klin Heidelberg, Stiftung Orthopad, Heidelberg, Germany
[2] Univ Gottingen, Inst Phys 4, D-3400 Gottingen, Germany
[3] Univ Gottingen, Abt Kieferorthopadie, D-3400 Gottingen, Germany
[4] Nanofilm NFT Gottingen, Gottingen, Germany
来源
BIOMEDIZINISCHE TECHNIK | 1999年 / 44卷 / 09期
关键词
new optical measuring method; polyethylene cups; total hip replacement;
D O I
10.1515/bmte.1999.44.9.247
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Aim: Most methods used for the determination of volumetric wear of polyethylene cups are based on the assumption that the head of the prosthesis penetrates the cup in "cylindrical" fashion. The new accurate optical method is independent of this disputable assumption. Method: The articulating surface of the cup is scanned with light and a data set of 60 000 pixels obtained in this way is stored in a computer. Data obtained from used cups were compared with those obtained from unused cups. The volumetric wear was calculated directly by threefold integration. To assess the changes in surface shape, the data are fitted by an ellipsoid whose long axis defines the mean direction of load. A total of 18 retrieved and 3 unused cups of different types were studied. Results: The unused acetabular cups deviated only slightly from ideal hemispheres. The surfaces showed rotational symmetry, and an undulation having an amplitude of 0.1 mm between dome and equator. For all explanted cups, the assumption of cylindrical penetration of the head into the polyethylene was shown not to represent the true situation. The cup expands in all directions, and the volumetric wear is underestimated by 50 % with the traditional methods. The data suggest that long-term survival may be jeopardized when the main direction of loading is centered on the dome of the cup. Ceramic heads were associated with smaller rates of volumetric wear. Conclusion: The new optical method is characterised by short measuring times, precision and simple application. Analysis of the wear patterns of polyethylene components using this technique may contribute to a further understanding of the complex mechanisms of aseptic loosening.
引用
收藏
页码:247 / 254
页数:8
相关论文
共 50 条
  • [21] Backsurface wear and deformation in polyethylene tibial inserts retrieved postmortem
    Surace, MF
    Berzins, A
    Urban, RM
    Jacobs, JJ
    Berger, RA
    Natarajan, RN
    Andriacchi, TP
    Galante, JO
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2002, (404) : 14 - 23
  • [22] Highly Crosslinked Polyethylene Improves Wear But Not Surface Damage in Retrieved Acetabular Liners
    Pang, Hee-Nee
    Naudie, Douglas D. R.
    McCalden, Richard W.
    MacDonald, Steven J.
    Teeter, Matthew G.
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2015, 473 (02) : 463 - 468
  • [23] Wear polyethylene cups in total hip arthroplasty - A study of specimens retrieved post mortem
    Sychterz, CJ
    Moon, KH
    Hashimoto, Y
    Terefenko, KM
    Engh, CA
    Bauer, TW
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1996, 78A (08): : 1193 - 1200
  • [24] Validation of a micro-CT technique for measuring volumetric wear in retrieved acetabular liners
    Bowden, AE
    Kurtz, SM
    Edidin, AA
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2005, 75B (01) : 205 - 209
  • [25] Wear of gamma-crosslinked polyethylene acetabular cups against roughened femoral balls
    McKellop, H
    Shen, FW
    DiMaio, W
    Lancaster, JG
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1999, (369) : 73 - 82
  • [26] Relationship of surface damage appearance and volumetric wear in retrieved TKR polyethylene liners
    Knowlton, Christopher B.
    Bhutani, Priyanka
    Wimmer, Markus A.
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2017, 105 (07) : 2053 - 2059
  • [27] A finite element method approach to compare the wear of acetabular cups in polyethylene according to their lateral tilt in relation to the coronal plane
    Queiroz, R. D.
    Oliveira, A. L. L.
    Trigo, F. C.
    Lopes, J. A.
    WEAR, 2013, 298 : 8 - 13
  • [28] Effect of femoral head surface roughness on the wear of ultrahigh molecular weight polyethylene acetabular cups
    Wang, A
    Polineni, VK
    Stark, C
    Dumbleton, JH
    JOURNAL OF ARTHROPLASTY, 1998, 13 (06): : 615 - 620
  • [29] Wear and surface cracking in early retrieved highly cross-linked polyethylene acetabular liners
    Bradford, L
    Baker, DA
    Graham, J
    Chawan, A
    Ries, MD
    Pruitt, LA
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2004, 86A (06): : 1271 - 1282
  • [30] Dual mobility retentive acetabular liners and wear: Surface analysis of 40 retrieved polyethylene implants
    Adam, P.
    Farizon, F.
    Fessy, M. -H.
    ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2014, 100 (01) : 85 - 91