Reproducibility of thrombelastometry (ROTEM®): Point-of-care versus hospital laboratory performance

被引:35
|
作者
Haas, Thorsten [1 ]
Spielmann, Nelly [1 ]
Mauch, Jacqueline [1 ]
Speer, Oliver [2 ]
Schmugge, Markus [2 ]
Weiss, Markus [1 ]
机构
[1] Univ Childrens Hosp Zurich, Dept Anaesthesia, CH-8032 Zurich, Switzerland
[2] Univ Childrens Hosp Zurich, Dept Haematol, CH-8032 Zurich, Switzerland
关键词
Anaesthesiology; point-of-care testing; internal quality control; secondary haemostasis; FIBRINOGEN;
D O I
10.3109/00365513.2012.665474
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
Thrombelastometry (ROTEM (R)) has gained wide acceptance in detecting and tailoring acquired hemostatic changes in adults and children. We investigated in this observational trial whether the reproducibility of this point-of-care testing was influenced by performance at the bedside or in the hospital laboratory. In addition, difference in time of performance between both measurements was compared. Perioperative blood samples obtained during major pediatric surgery were run in duplicate on two different ROTEM (R) devices located in the OR and in the hospital laboratory. The Bland-Altman test was used to compare differences of both measurements. ROTEM (R) measurements of 90 blood samples obtained from 24 children showed no overall clinically meaningful differences, whether they were performed bedside or in the hospital laboratory. Minor differences were found for the InTEM clot formation time (CFT) showing a mean bias of 10.79 seconds. Time saving was 11 minutes (8-16 minutes) if ROTEM (R) measurements were performed bedside (p < 0.001). In conclusion, there were minimal effects on ROTEM (R) measurements irrespective of whether they were performed in the hospital laboratory or at the bedside by a single trained staff member, while the latter saved valuable time.
引用
收藏
页码:313 / 317
页数:5
相关论文
共 50 条
  • [1] NT-proBNP on Cobas h 232 in point-of-care testing: Performance in the primary health care versus in the hospital laboratory
    Gils, Charlotte
    Ramanathan, Ramshanker
    Breindahl, Torben
    Brokner, Mette
    Christiansen, Anne L.
    Eng, Oyvin
    Hammer, Ida J.
    Herrera, Catherine B.
    Jansen, Aina
    Langsjoen, Eva C.
    Lokkebo, Elisabeth S.
    Osestad, Toril
    Schroder, Anne D.
    Walther, Lisa
    [J]. SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 2015, 75 (07): : 602 - 609
  • [2] Point-of-care versus laboratory measurement of the International Normalized Ratio
    Pierce, MT
    Crain, L
    Smith, J
    Mehta, V
    [J]. AMERICAN JOURNAL OF HEALTH-SYSTEM PHARMACY, 2000, 57 (24) : 2271 - 2274
  • [3] ROTEM/TEG® as a point-of-care coagulation monitoring device: pro
    Schoechl, H.
    [J]. ANASTHESIOLOGIE & INTENSIVMEDIZIN, 2007, 48 : S174 - S176
  • [4] COMPARING TEG® AND ROTEM® AS POINT-OF-CARE DEVICES IN TRAUMA PATIENTS
    Hagemo, J. S.
    Gaarder, C.
    Naess, P. A.
    Viksmoen, Y.
    Heier, H. E.
    [J]. VOX SANGUINIS, 2010, 99 : 437 - 437
  • [5] Variability of point-of-care testing blood glucometers versus the laboratory reference method in a tertiary teaching hospital
    Jamaluddin, Farhi Ain
    Gunavathy, M.
    Yean, Chew Yee
    Thevarajah, Malathi
    [J]. ASIAN BIOMEDICINE, 2012, 6 (01) : 67 - 74
  • [6] Point-of-care Laboratory Testing
    Gauglitz, Guenter
    Luppa, Peter B.
    [J]. CHEMIE IN UNSERER ZEIT, 2009, 43 (05) : 308 - 318
  • [7] Point-of-care laboratory instruments
    Wallen, Roy D.
    [J]. Biomedical Instrumentation and Technology, 2007, 41 (01): : 62 - 64
  • [8] Comparability of Point-of-Care versus Central Laboratory Hemoglobin Determination in Emergency Patients at a Supra-Maximal Care Hospital
    Dolscheid-Pommerich, Ramona C.
    Dolscheid, Sarah
    Grigutsch, Daniel
    Stoffel-Wagner, Birgit
    Graeff, Ingo
    [J]. PLOS ONE, 2016, 11 (11):
  • [9] Point-of-care aPTT testing versus established laboratory based systems
    Risch, A
    Biedler, A
    Mertzlufft, F
    [J]. ANESTHESIOLOGY, 1996, 85 (3A) : A454 - A454
  • [10] Point-of-care versus central laboratory testing of INR in acute stroke
    Zenlander, R.
    von Euler, M.
    Antovic, J.
    Berglund, A.
    [J]. ACTA NEUROLOGICA SCANDINAVICA, 2018, 137 (02): : 252 - 255