Reproducibility of thrombelastometry (ROTEM®): Point-of-care versus hospital laboratory performance

被引:35
|
作者
Haas, Thorsten [1 ]
Spielmann, Nelly [1 ]
Mauch, Jacqueline [1 ]
Speer, Oliver [2 ]
Schmugge, Markus [2 ]
Weiss, Markus [1 ]
机构
[1] Univ Childrens Hosp Zurich, Dept Anaesthesia, CH-8032 Zurich, Switzerland
[2] Univ Childrens Hosp Zurich, Dept Haematol, CH-8032 Zurich, Switzerland
关键词
Anaesthesiology; point-of-care testing; internal quality control; secondary haemostasis; FIBRINOGEN;
D O I
10.3109/00365513.2012.665474
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
Thrombelastometry (ROTEM (R)) has gained wide acceptance in detecting and tailoring acquired hemostatic changes in adults and children. We investigated in this observational trial whether the reproducibility of this point-of-care testing was influenced by performance at the bedside or in the hospital laboratory. In addition, difference in time of performance between both measurements was compared. Perioperative blood samples obtained during major pediatric surgery were run in duplicate on two different ROTEM (R) devices located in the OR and in the hospital laboratory. The Bland-Altman test was used to compare differences of both measurements. ROTEM (R) measurements of 90 blood samples obtained from 24 children showed no overall clinically meaningful differences, whether they were performed bedside or in the hospital laboratory. Minor differences were found for the InTEM clot formation time (CFT) showing a mean bias of 10.79 seconds. Time saving was 11 minutes (8-16 minutes) if ROTEM (R) measurements were performed bedside (p < 0.001). In conclusion, there were minimal effects on ROTEM (R) measurements irrespective of whether they were performed in the hospital laboratory or at the bedside by a single trained staff member, while the latter saved valuable time.
引用
收藏
页码:313 / 317
页数:5
相关论文
共 50 条
  • [31] Point-of-Care Testing Evolving at Tawam Hospital
    Dhatt, Gurdeep Singh
    Sahni, Sumedha
    Van Jaarsveld, Theresa
    James, David
    Matarelli, Steven
    [J]. POINT OF CARE, 2008, 7 (04): : 248 - 252
  • [32] The Equine Neonatal Intensive Care Laboratory: Point-of-Care Testing
    Wilkins, Pamela A.
    [J]. CLINICS IN LABORATORY MEDICINE, 2011, 31 (01) : 125 - +
  • [33] Ecarin modified rotational thrombelastometry: a point-of-care applicable alternative to monitor the direct thrombin inhibitor argatroban
    Schaden, Eva
    Schober, Andreas
    Hacker, Stefan
    Kozek-Langenecker, Sibylle
    [J]. WIENER KLINISCHE WOCHENSCHRIFT, 2013, 125 (5-6) : 156 - 159
  • [34] The impact of the introduction of a point-of-care haematology analyser in a New Zealand rural hospital with no onsite laboratory
    Blattnerl, Katharina
    Beazley, Catherine J.
    Nixon, Arry
    Herd, Geoffrey
    Wigglesworth, John
    Rogers-Koroheke, Marara Grace
    [J]. RURAL AND REMOTE HEALTH, 2019, 19 (02):
  • [35] Diagnostic performance of point-of-care and central laboratory cardiac troponin assays in an emergency department
    Wilke, Petra
    Masuch, Annette
    Fahron, Oliver
    Zylla, Stephanie
    Leipold, Tobias
    Petersmann, Astrid
    [J]. PLOS ONE, 2017, 12 (11):
  • [36] Thrombolysis in stroke patients: Comparability of point-of-care versus central laboratory international normalized ratio
    Dolscheid-Pommerich, Ramona C.
    Dolscheid, Sarah
    Eichhorn, Lars
    Stoffel-Wagner, Birgit
    Graeff, Ingo
    [J]. PLOS ONE, 2018, 13 (01):
  • [37] DIFFERENCES IN POINT-OF-CARE VERSUS CENTRAL LABORATORY HEMOGLOBIN LEVEL DUE TO EXTREME LEUKOCYTOSIS.
    Huisman, Albert
    de Vooght, Karen
    [J]. INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2012, 34 : 119 - 119
  • [38] Comparison of point-of-care HbA1c test versus standardized laboratory testing
    Sicard, DA
    Taylor, JR
    [J]. ANNALS OF PHARMACOTHERAPY, 2005, 39 (06) : 1024 - 1028
  • [39] Point-of-Care Testing: The Good, the Bad, and the Laboratory Oversight
    Shaw, Julie L. V.
    [J]. JOURNAL OF APPLIED LABORATORY MEDICINE, 2021, 6 (04): : 1090 - 1093
  • [40] THE SATELLITE LABORATORY AND POINT-OF-CARE TESTING - INTEGRATION OF INFORMATION
    JACOBS, E
    LAUDIN, AG
    [J]. AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1995, 104 (04) : S33 - S39