Reproducibility of thrombelastometry (ROTEM®): Point-of-care versus hospital laboratory performance

被引:35
|
作者
Haas, Thorsten [1 ]
Spielmann, Nelly [1 ]
Mauch, Jacqueline [1 ]
Speer, Oliver [2 ]
Schmugge, Markus [2 ]
Weiss, Markus [1 ]
机构
[1] Univ Childrens Hosp Zurich, Dept Anaesthesia, CH-8032 Zurich, Switzerland
[2] Univ Childrens Hosp Zurich, Dept Haematol, CH-8032 Zurich, Switzerland
关键词
Anaesthesiology; point-of-care testing; internal quality control; secondary haemostasis; FIBRINOGEN;
D O I
10.3109/00365513.2012.665474
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
Thrombelastometry (ROTEM (R)) has gained wide acceptance in detecting and tailoring acquired hemostatic changes in adults and children. We investigated in this observational trial whether the reproducibility of this point-of-care testing was influenced by performance at the bedside or in the hospital laboratory. In addition, difference in time of performance between both measurements was compared. Perioperative blood samples obtained during major pediatric surgery were run in duplicate on two different ROTEM (R) devices located in the OR and in the hospital laboratory. The Bland-Altman test was used to compare differences of both measurements. ROTEM (R) measurements of 90 blood samples obtained from 24 children showed no overall clinically meaningful differences, whether they were performed bedside or in the hospital laboratory. Minor differences were found for the InTEM clot formation time (CFT) showing a mean bias of 10.79 seconds. Time saving was 11 minutes (8-16 minutes) if ROTEM (R) measurements were performed bedside (p < 0.001). In conclusion, there were minimal effects on ROTEM (R) measurements irrespective of whether they were performed in the hospital laboratory or at the bedside by a single trained staff member, while the latter saved valuable time.
引用
收藏
页码:313 / 317
页数:5
相关论文
共 50 条
  • [41] The portable laboratory and emergency department point-of-care testing
    van Heyningen, C
    Watson, ID
    [J]. ANNALS OF CLINICAL BIOCHEMISTRY, 1999, 36 : 531 - 531
  • [42] The laboratory-clinical interface - Point-of-care testing
    Kost, GJ
    Ehrmeyer, SS
    Chernow, B
    Winkelman, JW
    Zaloga, GP
    Dellinger, RR
    Shirey, T
    [J]. CHEST, 1999, 115 (04) : 1140 - 1154
  • [43] Comparison of point-of-care and Laboratory troponin I assays
    Markota, Andrej
    Bernhardt, Matej
    Palfy, Miroslav
    [J]. ZDRAVNISKI VESTNIK-SLOVENIAN MEDICAL JOURNAL, 2011, 80 (12): : 905 - 908
  • [44] Point-of-Care Test Equipment for Flexible Laboratory Automation
    You, Won Suk
    Park, Jae Jun
    Jin, Sung Moon
    Ryew, Sung Moo
    Choi, Hyouk Ryeol
    [J]. JALA, 2014, 19 (04): : 403 - 412
  • [45] Point-of-Care Testing, a Critical Component of Laboratory Medicine
    Hammett-Stabler, Catherine A.
    Nichols, James H.
    [J]. CLINICAL BIOCHEMISTRY, 2009, 42 (03) : 135 - 135
  • [46] Hospital point-of-care: Collection and integration of test results
    Mappes, RP
    Byrdy, PC
    Stephans, EJ
    [J]. LABORATORY MEDICINE, 2002, 33 (01) : 33 - +
  • [47] Point-of-care testing management protocol in a Paris hospital
    Pernet, P.
    Beneteau-Burnat, B.
    Vaubourdolle, M.
    [J]. CLINICAL CHEMISTRY, 2008, 54 (06) : A194 - A194
  • [48] Use of point-of-care testing for respiratory viruses in hospital
    Hayward, Andrew C.
    [J]. LANCET RESPIRATORY MEDICINE, 2021, 9 (04): : 324 - 326
  • [49] Fulfilling the Potential of Point-of-Care Ultrasound in Hospital Medicine
    Wyman, Molly F.
    Yocum, Matthew D.
    Schnobrich, Daniel J.
    [J]. JOURNAL OF HOSPITAL MEDICINE, 2020, 15 (03) : 190 - 191
  • [50] How to compare point-of-care testing results with hospital laboratory results: Recommendations for primary healthcare laboratories
    Stavelin, A.
    Fossum, A. L.
    Binder, S.
    Aker, C.
    Ronneseth, E.
    Nguyen, M.
    Sandberg, S.
    [J]. CLINICA CHIMICA ACTA, 2024, 558