Quality indicators and technique for analyzing permanent I-125 prostate seed implants: Seven years postimplant dosimetry evaluation

被引:6
|
作者
Marcu, Loredana G. [1 ,2 ,3 ]
Lawson, John M. [1 ]
Rutten, Thomas [1 ]
Gowda, Raghu [4 ]
机构
[1] Royal Adelaide Hosp, Dept Med Phys, Adelaide, SA 5000, Australia
[2] Univ Oradea, Fac Sci, Oradea 10087, Romania
[3] Univ Adelaide, Sch Chem & Phys, Adelaide, SA 5000, Australia
[4] Royal Adelaide Hosp, Dept Radiat Oncol, Adelaide, SA 5000, Australia
关键词
image fusion; dose volume histogram; natural dose ratio; low dose rate brachytherapy; organs at risk; ESTRO/EAU/EORTC RECOMMENDATIONS; BRACHYTHERAPY; CANCER; IMPACT; MONOTHERAPY; EXPERIENCE; VOLUMES; EDEMA;
D O I
10.1118/1.4725173
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The roles of postimplant dosimetry (PID) after permanent I-125 implant are to identify and rectify inadequate implants, assess the dosimetric quality indicators, and evaluate dose to the organs at risk. The aim of the current work was to assess the progress of prostate implant quality via postimplant dosimetry over seven years. Methods: The following factors were investigated to assess the PID results obtained over seven years: the improvement in implant technique, the computed tomography (CT) delineation-based PID versus ultrasound-CT (US-CT) fusion-based PID, and the evolution of parameters such as D90 and NDR (natural dose ratio). The correlation between dosimetric parameters and clinical outcomes were also evaluated. Results: The seven years PID learning curve shows clear changes in dosimetric trend for the 265 patients studied. Manual target contouring on CT was shown to overestimate the prostate volume when compared to ultrasound data, translating to CT-based D90 values being lower than US-CT D90. It was found that NDR does not contribute with additional dosimetric information to postimplant dosimetry evaluation. Patient follow-up data show that 4.7% patients have relapsed, and urinary retention was reported in 2.7% of the patients. Conclusions: CT-based PID was found less reliable than US-CT fusion-based PID due to target volume overestimation. This result shows the biased interpretation of low D90 values based on CT-based targeting, providing unreliable correlations between D90 and relapse probability. The low urinary retention statistics are in accordance with the PID data for the organ, as only 5.2% of patients had their PID D10 > 218 Gy, i.e., above the recommended GEC-ESTRO guidelines. Besides the "learning" component, the PID D90 curve is influenced by the PID technique. (C) 2012 American Association of Physicists in Medicine. [http://dx.doi.org/10.1118/1.4725173]
引用
收藏
页码:4123 / 4131
页数:9
相关论文
共 50 条
  • [31] Influence of source strength on outcomes in patients undergoing prostate I-125 seed implants for prostate carcinoma
    Mahadevan, A
    Reddy, C
    Ulchaker, J
    Angermeier, K
    Djemil, T
    Wilkinson, W
    Ciezki, J
    Klein, E
    [J]. RADIOTHERAPY AND ONCOLOGY, 2004, 71 : S44 - S44
  • [32] Comprehensive I-125 mutti-seed comparison for prostate brachytherapy: Dosimetry and visibility analysis
    Al-Qaisieh, Bashar
    Smith, David W.
    Brearley, Elizabeth
    Clair, Shaun St.
    Bownes, Peter
    [J]. RADIOTHERAPY AND ONCOLOGY, 2007, 84 (02) : 140 - 147
  • [33] An evaluation of the AAPM-TG43 dosimetry protocol for I-125 brachytherapy seed
    Mobit, P
    Badragan, I
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (14): : 3161 - 3170
  • [34] Limitations of the minimum peripheral dose as a parameter for dose specification in permanent I-125 prostate implants
    Yu, Y
    Waterman, FM
    Suntharalingam, N
    Schulsinger, A
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1996, 34 (03): : 717 - 725
  • [35] Preplan Dosimetry as a Predictor of Clinical Outcome for 125I Permanent Prostate Implants (PPI)
    Alexander, J.
    Weinberg, V.
    Nash, M.
    Pickett, B.
    Gottschalk, A.
    Hsu, I.
    Shinohara, K.
    Roach, M.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2012, 84 (03): : S185 - S185
  • [36] Improved homogeneity with fourth generation prostate brachytherapy implants - A revised I-125 seed nomogram
    Kudchadker, R.
    Johnson, J.
    Bruno, T.
    Frank, S.
    [J]. MEDICAL PHYSICS, 2007, 34 (06) : 2459 - 2459
  • [37] DOSIMETRIC COMPARISON BETWEEN LINKED AND STRAND SEEDS FOR I-125 PERMANENT SEED PROSTATE BRACHYTHERAPY
    Jakstas, Karolis
    Sniureviciute, Migle
    [J]. MEDICAL PHYSICS IN THE BALTIC STATES, 2010, (02): : 112 - +
  • [38] The effect of seed orientation deviations on the quality of 125I prostate implants
    Corbett, JF
    Jezioranski, JJ
    Crook, J
    Tran, T
    Yeung, IWT
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (11): : 2785 - 2800
  • [39] Seed Cloud Volume as Surrogate for Prostate Volume for Dosimetry of 125I Permanent Brachytherapy
    Ravindran, P.
    Sloboda, R.
    [J]. MEDICAL PHYSICS, 2010, 37 (06) : 3204 - +
  • [40] Seed displacements after permanent I-125 brachytherapy for prostate cancer - Comparison of seeds in different prostate areas
    Pinkawa, M.
    Gagel, B.
    Asadpour, B.
    Piroth, M. D.
    Klotz, J.
    Borchers, H.
    Jakse, G.
    Eble, M. J.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 72 (01): : S347 - S347