Accurate determination of ultra-trace levels of Ti in blood serum using ICP-MS/MS

被引:84
|
作者
Balcaen, Lieve [1 ]
Bolea-Fernandez, Eduardo [1 ]
Resano, Martin [2 ]
Vanhaecke, Frank [1 ]
机构
[1] Univ Ghent, Dept Analyt Chem, B-9000 Ghent, Belgium
[2] Univ Zaragoza, Dept Analyt Chem, E-50009 Zaragoza, Spain
关键词
Blood serum; Titanium; Reaction cell inductively coupled plasma mass spectrometry; Implants; TOTAL HIP-ARTHROPLASTY; TITANIUM LEVELS; METAL RELEASE; MS; SPECTROMETRY; EXCRETION; IMPLANTS; VANADIUM; URINE; ALLOY;
D O I
10.1016/j.aca.2013.10.017
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Ti is frequently used in implants and prostheses and it has been shown before that the presence of these in the human body can lead to elevated Ti concentrations in body fluids such as serum and urine. As identification of the exact mechanisms responsible for this increase in Ti concentrations, and the risks associated with it, are not fully understood, it is important to have sound analytical methods that enable straightforward quantification of Ti levels in body fluids (for both implanted and non-implanted individuals). Until now, only double-focusing sector field ICP-mass spectrometry (SF-ICP-MS) offered limits of detection that are good enough to deal with the very low basal levels of Ti in human serum. This work reports on the development of a novel method for the accurate and precise determination of trace levels of Ti in human serum samples, based on the use of ICP-MS/MS. O-2 and NH3/He have been compared as reaction gases. While the use of O-2 did not enable to overcome all spectral interferences, it has been shown that conversion of Ti+ ions into Ti(NH3)(6)(+) cluster ions by using NH3/He as a reaction gas in an ICP-QQQ-MS system, operated in MS/MS mode, provided interference-free conditions and sufficiently low limits of detection, down to 3 ng L-1 (instrumental detection limit obtained for the most abundant Ti isotope). The accuracy of the method proposed was evaluated by analysis of a Seronorm Trace Elements Serum L-1 reference material and by comparing the results obtained with those achieved by means of SF-ICP-MS. As a proof-of-concept, the newly developed method was successfully applied to the determination of Ti in serum samples obtained from individuals with and without Ti-based implants. All results were found to be in good agreement with those obtained by means of SF-ICP-MS. The typical basal Ti level in human serum was found to be < 1 mu g L-1, while values in the range of 2-6 mu g L-1 were observed for implanted patients. (C) 2013 Elsevier B.V. All rights reserved.
引用
收藏
页码:1 / 8
页数:8
相关论文
共 50 条
  • [21] An efficient method for the removal of Pb prior to the ultra-trace Am measurement by ICP-MS
    Wang, Yufeng
    Fan, Jinlong
    Zhai, Xiufang
    Li, Zhiming
    Yu, Qingjiang
    Zhang, Weichao
    Liu, Zhichao
    RADIOCHIMICA ACTA, 2022, 110 (10) : 825 - 831
  • [22] Automated separation of ultra-trace actinides from urine by ion chromatography ICP-MS
    Jones, Robert L.
    Caldwell, Kathleen L.
    Gonzales, Edward R.
    Peterson, Dominic S.
    Walter, Peter J.
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2006, 231
  • [23] Surface cleaning techniques: ultra-trace ICP-MS sample preparation and assay of HDPE
    N. R. Overman
    E. W. Hoppe
    R. S. Addleman
    Journal of Radioanalytical and Nuclear Chemistry, 2013, 296 : 1303 - 1310
  • [24] Elimination of oxide interferences and determination of ultra-trace silver in soils by ICP-MS with ion-molecule reactions
    Guo, Wei
    Hu, Shenghong
    Zhang, Jiangyi
    Zhang, Hongfei
    SCIENCE OF THE TOTAL ENVIRONMENT, 2011, 409 (15) : 2981 - 2986
  • [25] Determination of Ultra-trace Levels of 238U, 232Th, and 40K in Polymethyl Methacrylate by ICP-MS
    Feng, Shunqing
    Li, Mengting
    Wang, Wenqin
    Wang, Xiangyue
    Huang, Chunhao
    Liu, Hongtao
    Tanga, Jian
    ATOMIC SPECTROSCOPY, 2025, 46 (01)
  • [26] Determination of ultra-trace elements in human plasma or serum by ICP-MS using sodium in the presence of carbon as a single calibration matrix-match component
    Gajek, Ryszard
    Choe, Key-Young
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2015, 30 (05) : 1142 - 1153
  • [27] Optimization of analytical methods for origin assessment of orange juices .2. ICP-MS determination of trace and ultra-trace elements
    Martin, GJ
    Fournier, JB
    Allain, P
    Mauras, Y
    Aguile, L
    ANALUSIS, 1997, 25 (01) : 7 - 13
  • [28] A quick and simple method for the determination of six trace elements in mammalian serum samples using ICP-MS/MS
    Kopp, Johannes Florian
    Mueller, Sandra Marie
    Pohl, Gabriele
    Lossow, Kristina
    Kipp, Anna Patricia
    Schwerdtle, Tanja
    JOURNAL OF TRACE ELEMENTS IN MEDICINE AND BIOLOGY, 2019, 54 : 221 - 225
  • [29] Trace element determination in lichens by ICP-MS
    Bettinelli, M
    Spezia, S
    Bizzarri, G
    ATOMIC SPECTROSCOPY, 1996, 17 (03) : 133 - 141
  • [30] ICP-MS determination of trace elements in fish
    Chamberlain, I
    Adams, K
    Le, S
    ATOMIC SPECTROSCOPY, 2000, 21 (04) : 118 - 122