Ion recombination corrections of ionization chambers in flattening filter-free photon radiation

被引:29
|
作者
Wang, Yuenan [1 ]
Easterling, Stephen B. [1 ]
Ting, Joseph Y. [1 ]
机构
[1] Melbourne Canc Ctr, Melbourne, FL 32901 USA
来源
关键词
flattening filter free; ion recombination; two-voltage technique; P-ion; ionization chamber; MODULATED ARC THERAPY; STEREOTACTIC RADIOTHERAPY; REFERENCE DOSIMETRY; ELECTRON-BEAMS; ACCELERATOR;
D O I
10.1120/jacmp.v13i5.3758
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The flattening filter free (FFF) X-rays can provide much higher dose rate at the treatment target compared to the conventional flattened X-rays. However, the substantial increase of dose rate for FFF beams may affect the ion recombination correction factor, which is required for accurate measurements using ionization chambers in clinical dosimetry. The purpose of this work is to investigate the ion recombination of three types of commonly used ion chambers (Farmer, PinPoint and plane-parallel) in the FFF photon radiation. Both 6 MV and 10 MV flattened and FFF beams were fully commissioned on a Varian TrueBeam linear accelerator. The ion recombination correction factor, P-ion, was determined using the two-voltage technique for a 0.6 cc Farmer chamber, a 0.015 cc PinPoint chamber, and a 0.02 cc parallel-plate chamber at different source-to-axis distances (SAD) in a solid water phantom or water tank phantom at a depth of 10 cm in a 10 x 10 cm(2) field. Good repeatability of measurements was demonstrated. Less than 1% difference in Pion between the flattened and FFF photons for all three ion chambers was observed. At a SAD of 100 cm and a depth of 10 cm for a 10 x 10 cm(2) field, P-ion for the Farmer chamber was 1.004 and 1.008 for the 6 MV flattened and FFF beams, respectively. At the same setup using the Farmer chamber, P-ion was 1.002 and 1.015 for the 10 MV flattened and FFF beams, respectively. All P-ion results for the Farmer, PinPoint, or parallel plate chamber in the 6 MV and 10 MV flattened and FFF beams were within 2% from the unity (1 <= P-ion < 1.02). The Pion ratio of the FFF to flattened beams was 0.99 similar to 1.01 for both 6 MV and 10 MV photons. The ion recombination effect of the Farmer, PinPoint, and plane-parallel chamber in the FFF beams is not substantially different from that in the conventional flattened beams.
引用
下载
收藏
页码:262 / 268
页数:7
相关论文
共 50 条
  • [11] Use and Choice of Ionization Chambers for the Commissioning of Flattened and Flattening-Filter-Free Photon Beams: Determination of Recombination Correction Factor (ks)
    Stucchi, C.
    Bresolin, A.
    Mongioj, V.
    Carrara, M.
    Pignoli, E.
    Bonfantini, F.
    MEDICAL PHYSICS, 2014, 41 (06) : 371 - 371
  • [12] ION RECOMBINATION CORRECTIONS FOR PLANE-PARALLEL AND THIMBLE CHAMBERS IN ELECTRON AND PHOTON RADIATION
    HAVERCROFT, JM
    KLEVENHAGEN, SC
    PHYSICS IN MEDICINE AND BIOLOGY, 1993, 38 (01): : 25 - 38
  • [13] On flattening filter-free portal dosimetry
    Pardo, Eduardo
    Castro Novais, Juan
    Molina Lopez, Maria Yolanda
    Ruiz Maqueda, Sheila
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2016, 17 (04): : 132 - 145
  • [14] The impact of flattening filter free (FFF) photon beams to ion recombination correction factor
    Hentihu, F. K.
    Ryangga, D.
    Pawiro, S. A.
    18TH ASIA-OCEANIA CONGRESS OF MEDICAL PHYSICS (AOCMP) & 16TH SOUTH-EAST ASIA CONGRESS OF MEDICAL PHYSICS (SEACOMP), 2019, 1248
  • [15] The comparison of properties for radiotherapy with flattening filter-free and flattening filter beam
    Gu, J. H.
    Won, H. S.
    Hong, J. W.
    Chang, N. J.
    Park, J. H.
    RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S978 - S978
  • [16] Surface Dose Evaluation of the Equivalent Quality Flattening Filter-Free Photon Beams
    Meshram, M.
    Pramanik, S.
    Ranjith, C. P.
    Meshram, M.
    MEDICAL PHYSICS, 2015, 42 (06) : 3502 - 3502
  • [17] Dosimetric Evaluation of Flattening Filter-free (FFF) Photon Beams for Lung SBRT
    Chen, Q.
    Cai, J.
    Wu, Q.
    Yin, F.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2012, 84 (03): : S609 - S609
  • [18] COMMENTS ON ION RECOMBINATION CORRECTIONS FOR PLANE-PARALLEL AND THIMBLE CHAMBERS IN ELECTRON AND PHOTON RADIATION
    BURNS, JE
    BURNS, DT
    PHYSICS IN MEDICINE AND BIOLOGY, 1993, 38 (12): : 1986 - 1988
  • [19] ION RECOMBINATION CORRECTIONS FOR PLANE-PARALLEL AND THIMBLE CHAMBERS IN ELECTRON AND PHOTON RADIATION - REPLY
    KLEVENHAGEN, SC
    HAVERCROFT, JM
    PHYSICS IN MEDICINE AND BIOLOGY, 1993, 38 (12): : 1988 - 1991
  • [20] Investigation of neutron contamination of flattening filter and flattening filter-free 10-MV photon beams in Elekta InfinityTM accelerator
    Yani S.
    Budiansah I.
    Lestari F.P.
    Tursinah R.
    Rhani M.F.
    Haryanto F.