Experimental investigation of design parameters on dry powder inhaler performance

被引:16
|
作者
Nguyen Thi Quynh Ngoc [1 ]
Chang, Lusi [1 ]
Jia, Xinli [1 ]
Lau, Raymond [1 ]
机构
[1] Nanyang Technol Univ, Sch Chem & Biomed Engn, Singapore 639798, Singapore
关键词
Dry powder inhaler; Flow field; Turbulence intensity; In vitro deposition; Emitted dose; Fine particle fraction; COMPUTATIONAL FLUID-DYNAMICS; POLLEN-SHAPE CARRIER; AIR-FLOW; INHALATION PERFORMANCE; DRUG-DELIVERY; PARTICLE-SIZE; DEVICE DESIGN; FORMULATION; DISPERSION; DEPOSITION;
D O I
10.1016/j.ijpharm.2013.08.072
中图分类号
R9 [药学];
学科分类号
1007 ;
摘要
The study aims to investigate the impact of various design parameters of a dry powder inhaler on the turbulence intensities generated and the performance of the dry powder inhaler. The flow fields and turbulence intensities in the dry powder inhaler are measured using particle image velocimetry (PIV) techniques. In vitro aerosolization and deposition a blend of budesonide and lactose are measured using an Andersen Cascade Impactor. Design parameters such as inhaler grid hole diameter, grid voidage and chamber length are considered. The experimental results reveal that the hole diameter on the grid has negligible impact on the turbulence intensity generated in the chamber. On the other hand, hole diameters smaller than a critical size can lead to performance degradation due to excessive particle-grid collisions. An increase in grid voidage can improve the inhaler performance but the effect diminishes at high grid voidage. An increase in the chamber length can enhance the turbulence intensity generated but also increases the powder adhesion on the inhaler wall. (C) 2013 Elsevier B.V. All rights reserved.
引用
收藏
页码:92 / 100
页数:9
相关论文
共 50 条
  • [21] Effect of turbulent kinetic energy on dry powder inhaler performance
    Suwandecha, Tan
    Wongpoowarak, Wibul
    Maliwan, Kittinan
    Srichana, Teerapol
    POWDER TECHNOLOGY, 2014, 267 : 381 - 391
  • [22] Lactose fluidisation properties and their relationship to dry powder inhaler performance
    Kinnunen, H.
    Shur, J.
    Hebbink, G.
    Muresan, A. S.
    Price, R.
    JOURNAL OF PHARMACY AND PHARMACOLOGY, 2010, 62 (10) : 1342 - 1343
  • [23] Effect of carrier particle shape on dry powder inhaler performance
    Kaialy, Waseem
    Alhalaweh, Amjad
    Velaga, Sitaram P.
    Nokhodchi, Ali
    INTERNATIONAL JOURNAL OF PHARMACEUTICS, 2011, 421 (01) : 12 - 23
  • [24] Drug delivery performance of the mometasone furoate dry powder inhaler
    Yang, TT
    Li, S
    Wyka, B
    Kenyon, D
    JOURNAL OF AEROSOL MEDICINE-DEPOSITION CLEARANCE AND EFFECTS IN THE LUNG, 2001, 14 (04): : 487 - 494
  • [25] An In Vitro Examination of the Effects of Altitude on Dry Powder Inhaler Performance
    Ruzycki, Conor A.
    Martin, Andrew R.
    Vehring, Reinhard
    Finlay, Warren H.
    JOURNAL OF AEROSOL MEDICINE AND PULMONARY DRUG DELIVERY, 2018, 31 (04) : 221 - 236
  • [26] Performance of salbutamol sulphate and budesonide in a novel dry powder inhaler
    Harjunen, P
    Lehto, VP
    Lankinen, T
    Järvinen, K
    EUROPEAN JOURNAL OF PHARMACEUTICAL SCIENCES, 2003, 19 : S35 - S35
  • [27] Dry powder inhaler device influence on carrier particle performance
    Donovan, Martin J.
    Kim, Sin Hyen
    Raman, Venkatramanan
    Smyth, Hugh D.
    JOURNAL OF PHARMACEUTICAL SCIENCES, 2012, 101 (03) : 1097 - 1107
  • [28] IS THE CAPSULE FILLING PROCESS ALTERING DRY POWDER INHALER PERFORMANCE?
    Kinnunen, H.
    Hebbink, G. A.
    Shur, J.
    Price, R.
    JOURNAL OF AEROSOL MEDICINE AND PULMONARY DRUG DELIVERY, 2013, 26 (05) : A24 - A24
  • [29] The clinical relevance of dry powder inhaler performance for drug delivery
    Demoly, Pascal
    Hagedoorn, Paul
    de Boer, Anne H.
    Frijlink, Henderik W.
    RESPIRATORY MEDICINE, 2014, 108 (08) : 1195 - 1203
  • [30] A CFD-DEM investigation of powder transport and aerosolization in ELLIPTA® dry powder inhaler
    Sulaiman, Mostafa
    Liu, Xiaoyu
    Sundaresan, Sankaran
    POWDER TECHNOLOGY, 2022, 409