Experimental investigation of design parameters on dry powder inhaler performance

被引:16
|
作者
Nguyen Thi Quynh Ngoc [1 ]
Chang, Lusi [1 ]
Jia, Xinli [1 ]
Lau, Raymond [1 ]
机构
[1] Nanyang Technol Univ, Sch Chem & Biomed Engn, Singapore 639798, Singapore
关键词
Dry powder inhaler; Flow field; Turbulence intensity; In vitro deposition; Emitted dose; Fine particle fraction; COMPUTATIONAL FLUID-DYNAMICS; POLLEN-SHAPE CARRIER; AIR-FLOW; INHALATION PERFORMANCE; DRUG-DELIVERY; PARTICLE-SIZE; DEVICE DESIGN; FORMULATION; DISPERSION; DEPOSITION;
D O I
10.1016/j.ijpharm.2013.08.072
中图分类号
R9 [药学];
学科分类号
1007 ;
摘要
The study aims to investigate the impact of various design parameters of a dry powder inhaler on the turbulence intensities generated and the performance of the dry powder inhaler. The flow fields and turbulence intensities in the dry powder inhaler are measured using particle image velocimetry (PIV) techniques. In vitro aerosolization and deposition a blend of budesonide and lactose are measured using an Andersen Cascade Impactor. Design parameters such as inhaler grid hole diameter, grid voidage and chamber length are considered. The experimental results reveal that the hole diameter on the grid has negligible impact on the turbulence intensity generated in the chamber. On the other hand, hole diameters smaller than a critical size can lead to performance degradation due to excessive particle-grid collisions. An increase in grid voidage can improve the inhaler performance but the effect diminishes at high grid voidage. An increase in the chamber length can enhance the turbulence intensity generated but also increases the powder adhesion on the inhaler wall. (C) 2013 Elsevier B.V. All rights reserved.
引用
收藏
页码:92 / 100
页数:9
相关论文
共 50 条
  • [1] Evolution of dry powder inhaler design, formulation, and performance
    Newman, SP
    Busse, WW
    RESPIRATORY MEDICINE, 2002, 96 (05) : 293 - 304
  • [2] Characterization of dry powder inhaler performance through experimental methods
    Ruzycki, Conor A.
    Tavernini, Scott
    Martin, Andrew R.
    Finlay, Warren H.
    ADVANCED DRUG DELIVERY REVIEWS, 2022, 189
  • [3] A novel dry powder inhaler: Effect of device design on dispersion performance
    Selvam, Parthiban
    McNair, Doug
    Truman, Randall
    Smyth, Hugh D. C.
    INTERNATIONAL JOURNAL OF PHARMACEUTICS, 2010, 401 (1-2) : 1 - 6
  • [4] The Role of Capsule on the Performance of a Dry Powder Inhaler Using Computational and Experimental Analyses
    Matthew S. Coates
    David F. Fletcher
    Hak-Kim Chan
    Judy A. Raper
    Pharmaceutical Research, 2005, 22 : 923 - 932
  • [5] The role of capsule on the performance of a dry powder inhaler using computational and experimental analyses
    Coates, MS
    Fletcher, DF
    Chan, HK
    Raper, JA
    PHARMACEUTICAL RESEARCH, 2005, 22 (06) : 923 - 932
  • [6] Experimental observations of dry powder inhaler dose fluidisation
    Tuley, Rob
    Shrimpton, John
    Jones, Matthew D.
    Price, Rob
    Palmer, Mark
    Prime, Dave
    INTERNATIONAL JOURNAL OF PHARMACEUTICS, 2008, 358 (1-2) : 238 - 247
  • [7] Rational design of a dry powder inhaler: device design and optimisation
    Friebel, Christian
    Steckel, Hartwig
    Mueller, Bernd W.
    JOURNAL OF PHARMACY AND PHARMACOLOGY, 2012, 64 (09) : 1303 - 1315
  • [8] Influence of Air Flow on the Performance of a Dry Powder Inhaler Using Computational and Experimental Analyses
    Matthew S. Coates
    Hak-Kim Chan
    David F. Fletcher
    Judy A. Raper
    Pharmaceutical Research, 2005, 22 : 1445 - 1453
  • [9] Influence of air flow on the performance of a dry powder inhaler using computational and experimental analyses
    Coates, MS
    Chan, HK
    Fletcher, DF
    Raper, JA
    PHARMACEUTICAL RESEARCH, 2005, 22 (09) : 1445 - 1453