Potential thermal artifacts in hip joint wear simulators

被引:1
|
作者
Lu, Z [1 ]
McKellop, H
Liao, P
Benya, P
机构
[1] Los Angeles Orthopaed Hosp, J Vernon Luck Orthopaed Res Ctr, Los Angeles, CA USA
[2] Univ So Calif, Sch Med, Dept Orthopaed, Los Angeles, CA 90033 USA
[3] Univ So Calif, Dept Biomed Engn, Los Angeles, CA 90089 USA
来源
关键词
wear; hip prosthesis; polyethylene; frictional heat; hip simulator;
D O I
10.1002/(SICI)1097-4636(1999)48:4<458::AID-JBM9>3.3.CO;2-2
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Frictional heat was monitored during wear tests of ultrahigh molecular weight polyethylene acetabular cups bearing against femoral balls of metal or ceramic in a hip simulator, using bovine serum as a lubricant. About 1 to 2 h of continuous cycling were required for the temperature in the zone of contact between the cup and ball to rise to its maximum steady value, and this equilibrium temperature was markedly higher with increased load and/or cycling rate, Frictional heating caused substantial precipitation of the proteins from the serum and, in some of the tests running at 1.5 or 2 Hz, an adherent proteinaceous layer was observed attached to the surface of the balls. The maximum temperature was also substantially higher in tests run with the cup mounted above the ball rather than below. Surprisingly, the tests running at higher frictional torque and temperature (i.e., those with the most protein precipitation and/or adherent layers) produced the least wear of. the polyethylene. This might have been due to the solid proteins that formed a protective layer between the ball and cup. Because patients with hip prostheses typically do not walk for hours without rest, the maximum temperatures irt vivo are likely to be much lower than those reached in the hip simulator, Therefore, the affects of protein precipitation on the resultant wear properties of the materials should be considered potential artifacts of the hip simulator tests. Increasing the volume of the lubricant bath reduced the maximum temperatures for tests running at 1.5 Nz but had little affect at 2 Hz. Reducing the cycling rate is an effective way to avoid overheating of the specimens, but this necessarily extends the time required to complete a test. (C) 1999 John Wiley & Sons, Inc.
引用
收藏
页码:458 / 464
页数:7
相关论文
共 50 条
  • [21] Computational hip joint simulator for wear and heat generation
    Fialho, Jorge C.
    Fernandes, Paulo R.
    Eca, Luis
    Folgado, Joao
    JOURNAL OF BIOMECHANICS, 2007, 40 (11) : 2358 - 2366
  • [22] Experimental Studies of the Unevenness of Wear of Hip Joint Endoprostheses
    Kolesnik A.I.
    Lipatov V.A.
    Frolov E.B.
    Biomedical Engineering, 2017, 50 (6) : 376 - 379
  • [23] IRREGULARITIES IN WEAR OF PLASTICS MATERIALS IN HIP-JOINT
    TRAVENIUO, SZM
    PLASTICS AND RUBBER, 1976, 1 (05): : 194 - 196
  • [24] Simultaneous measurement of friction and wear in hip simulators (vol 230, pg 373, 2016)
    Haider, Hani
    Weisenburger, Joel N.
    Garvin, Kevin L.
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2018, 232 (04) : 434 - 434
  • [25] CHOICE OF A RATIONAL HIP JOINT WEAR SIMULATOR DESIGN
    Pakhaliuk, Volodymyr
    Pashkov, Yevgen
    Desyatov, Ivan
    Manchuk, Vitaly
    ANNALS OF DAAAM FOR 2009 & PROCEEDINGS OF THE 20TH INTERNATIONAL DAAAM SYMPOSIUM, 2009, 20 : 1559 - 1560
  • [26] Forecasting Wear of Head and Acetabulum in Hip Joint Implant
    Szarek, Arkadiusz
    Korytkowski, Marcin
    Rutkowski, Leszek
    Scherer, Rafal
    Szyprowski, Janusz
    ARTIFICIAL INTELLIGENCE AND SOFT COMPUTING, PT II, 2012, 7268 : 341 - 346
  • [27] A triaxial physiological hip joint simulator for the wear testing of hip joint prostheses - Drive and control principles
    Springstein, D
    Wendt, Z
    Boenick, U
    BIOMEDIZINISCHE TECHNIK, 2003, 48 (04): : 106 - 109
  • [28] Wear of the artificial hip joint material under lubrication
    Waters, E. P. J.
    Spedding, P. L.
    Dohert, A. P.
    Spedding, R. L.
    ASIA-PACIFIC JOURNAL OF CHEMICAL ENGINEERING, 2009, 4 (01) : 80 - 89
  • [29] Quantitative evaluation of parameters used in wear testing simulators of total hip arthroplasty components
    Oliveira, A. L. L.
    Trigo, F. C.
    Martins, F. P. R.
    WEAR, 2014, 313 (1-2) : 1 - 10
  • [30] CALCULATION OF WEAR (FI WEAR MODULUS) IN PLASTIC CUP OF A HIP-JOINT PROSTHESIS
    LIGTERINK, DJ
    WEAR, 1975, 35 (01) : 113 - 121