Collimator Design for a Brain SPECT/MRI Insert

被引:14
|
作者
Salvado, Debora [1 ]
Erlandsson, Kjell [1 ]
Bousse, Alexandre [1 ]
Occhipinti, Michele [2 ]
Busca, Paolo [2 ]
Fiorini, Carlo [2 ]
Hutton, Brian F. [1 ,3 ]
机构
[1] UCL, Inst Nucl Med, London W1E 6BT, England
[2] Politecn Milan, Dipartimento Elettron Informaz & Bioingn, I-20133 Milan, Italy
[3] Univ Wollongong, Ctr Med Radiat Phys, Sydney, NSW, Australia
关键词
Collimator design; multi-modality; pinhole; slit-slat; SPECT insert; stationary system; SLIT-SLAT; EFFECTIVE DIAMETER; SMALL ANIMALS; RESOLUTION; SYSTEM; SIMULATION; PERFORMANCE; IMAGER;
D O I
10.1109/TNS.2015.2450017
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This project's goal is to design a SPECT insert for a clinical MRI system for simultaneous brain SPECT/MR imaging, with a high-sensitivity collimator and high-resolution detectors. We have compared eight collimator designs, four multi-pinhole and four multi-slit slit-slat configurations. The collimation was designed for a system with 2 rings of 25 5 x 5 cm detectors. We introduce the concept of 1/2-pinhole and 1/2-slit, which are transaxially shared between two adjacent detectors. Analytical geometric efficiency was calculated for an activity distribution corresponding to a human brain and a range of intrinsic detector resolutions R-i and target resolutions R-t at the centre of the FOV. Noise-free data were simulated with and without depth-of-interaction (DOI) information, 0.8 mm R-i and 10 mm R-t FWHM, and reconstructed for uniform, Defrise, Derenzo, and Zubal brain phantoms. Comparing the multi-pinhole and multi-slit slit-slat collimators, the former gives better reconstructed uniformity and transaxial resolution, while the latter gives better axial resolution. Although the 2 x 2-pin-hole and 2-slit designs give the highest sensitivities, they result in a sub-optimal utilisation of the detector FOV. The best options are therefore the 5+2 1/2-pinhole and the 1+2 1/2-slit systems, with sensitivities of 1.8 x 10(-4) and 3.2 x 10(-4), respectively. Noiseless brain phantom reconstructions with the multi-pinhole collimator are slightly superior as compared to slit-slat, in terms of symmetry and accuracy of the activity distribution, but the same is not true when noise is included. DOI information reduces artefacts and improves uniformity in geometric phantoms. Further evaluation is needed with prototype collimators.
引用
收藏
页码:1716 / 1724
页数:9
相关论文
共 50 条
  • [41] DESIGN AND CLINICAL UTILITY OF A FAN BEAM COLLIMATOR FOR SPECT IMAGING OF THE HEAD
    TSUI, BMW
    GULLBERG, GT
    EDGERTON, ER
    GILLAND, DR
    PERRY, JR
    MCCARTNEY, WH
    JOURNAL OF NUCLEAR MEDICINE, 1986, 27 (06) : 810 - 819
  • [42] Single and multipinhole collimator design evaluation method for small animal SPECT
    Vunckx, Kathleen
    Beque, Dirk
    Defrise, Michel
    Nuyts, Johan
    2005 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2005, : 2223 - 2227
  • [43] DESIGN CONSIDERATIONS AND SIMULATIONS FOR RING SPECT SYSTEMS WITH A MODULAR FOCUSED COLLIMATOR
    CHANG, W
    TIAN, C
    LI, S
    TSUI, BMW
    REZAI, K
    WILLIAMS, JJ
    EHRHARDT, JE
    KIRCHNER, PT
    JOURNAL OF NUCLEAR MEDICINE, 1986, 27 (06) : 1017 - 1017
  • [44] Single and multipinhole collimator design evaluation method for small animal SPECT
    Vunckx, Kathleen
    Beque, Dirk
    Defrise, Michel
    Nuyts, Johan
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2008, 27 (01) : 36 - 46
  • [45] A new collimator insert system for stereotactic radiotherapy
    Paschalis, T
    Sandilos, P
    Karaiskos, P
    Chatzigiannis, H
    Koutsouveli, E
    Dardoufas, K
    Vlachos, L
    RADIOTHERAPY AND ONCOLOGY, 2005, 76 : S212 - S212
  • [46] Design and initial results of an RF-penetrable TOF PET insert dedicated for brain PET/MRI
    Chang, Chen-Ming
    Lee, Brian J.
    Sacco, Ilaria
    Watkins, Ronald D.
    Dong, Qian
    Levin, Craig S.
    2017 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2017,
  • [47] Fabrication and Calibration of a Novel High-Sensitivity Collimator for Brain SPECT Imaging
    Park, M.
    Kijewski, M.
    Horky, L.
    Keijzers, M.
    Keijzers, R.
    Kalfin, L.
    Crough, J.
    Goswami, M.
    Moore, S.
    MEDICAL PHYSICS, 2014, 41 (06) : 106 - +
  • [48] Influence of collimator hole dimensions on parallel and cone-beam brain SPECT
    Kamphuis, C
    Beekman, FJ
    Hutton, BF
    1999 IEEE NUCLEAR SCIENCE SYMPOSIUM - CONFERENCE RECORD, VOLS 1-3, 1999, : 1047 - 1051
  • [49] Half-millimeter animal SPECT imaging on a clinical SPECT scanner with highly flexible collimator design
    Ma, Tianyu
    Dai, Tiantian
    Liu, Hui
    Cui, Junjian
    Wang, Shi
    Liu, Yangqiang
    Wei, Qingyang
    Wu, Jing
    Jin, Yongjie
    2010 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD (NSS/MIC), 2010,
  • [50] Clinical, MRI and brain SPECT correlations in dementias
    Muthukrishnan Rethinam
    Sivakumar Rajappa
    A Anirudhan
    S Shelley
    M Indirani
    Sanjeev Sivakumar
    Annals of General Psychiatry, 5 (Suppl 1)