Collimator Design for a Brain SPECT/MRI Insert

被引:14
|
作者
Salvado, Debora [1 ]
Erlandsson, Kjell [1 ]
Bousse, Alexandre [1 ]
Occhipinti, Michele [2 ]
Busca, Paolo [2 ]
Fiorini, Carlo [2 ]
Hutton, Brian F. [1 ,3 ]
机构
[1] UCL, Inst Nucl Med, London W1E 6BT, England
[2] Politecn Milan, Dipartimento Elettron Informaz & Bioingn, I-20133 Milan, Italy
[3] Univ Wollongong, Ctr Med Radiat Phys, Sydney, NSW, Australia
关键词
Collimator design; multi-modality; pinhole; slit-slat; SPECT insert; stationary system; SLIT-SLAT; EFFECTIVE DIAMETER; SMALL ANIMALS; RESOLUTION; SYSTEM; SIMULATION; PERFORMANCE; IMAGER;
D O I
10.1109/TNS.2015.2450017
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This project's goal is to design a SPECT insert for a clinical MRI system for simultaneous brain SPECT/MR imaging, with a high-sensitivity collimator and high-resolution detectors. We have compared eight collimator designs, four multi-pinhole and four multi-slit slit-slat configurations. The collimation was designed for a system with 2 rings of 25 5 x 5 cm detectors. We introduce the concept of 1/2-pinhole and 1/2-slit, which are transaxially shared between two adjacent detectors. Analytical geometric efficiency was calculated for an activity distribution corresponding to a human brain and a range of intrinsic detector resolutions R-i and target resolutions R-t at the centre of the FOV. Noise-free data were simulated with and without depth-of-interaction (DOI) information, 0.8 mm R-i and 10 mm R-t FWHM, and reconstructed for uniform, Defrise, Derenzo, and Zubal brain phantoms. Comparing the multi-pinhole and multi-slit slit-slat collimators, the former gives better reconstructed uniformity and transaxial resolution, while the latter gives better axial resolution. Although the 2 x 2-pin-hole and 2-slit designs give the highest sensitivities, they result in a sub-optimal utilisation of the detector FOV. The best options are therefore the 5+2 1/2-pinhole and the 1+2 1/2-slit systems, with sensitivities of 1.8 x 10(-4) and 3.2 x 10(-4), respectively. Noiseless brain phantom reconstructions with the multi-pinhole collimator are slightly superior as compared to slit-slat, in terms of symmetry and accuracy of the activity distribution, but the same is not true when noise is included. DOI information reduces artefacts and improves uniformity in geometric phantoms. Further evaluation is needed with prototype collimators.
引用
收藏
页码:1716 / 1724
页数:9
相关论文
共 50 条
  • [31] Image characteristics of brain perfusion SPECT/CT using a new multi-focal collimator: Comparison with conventional SPECT with LEHR collimator
    Shibutani, Takayuki
    Onoguchi, Masahisa
    Nishiyama, Tomoki
    Vija, Alexander Hans
    Massanes, Francesc
    Shimizu, Takeshi
    Yoneyama, Hiroto
    Konishi, Takahiro
    Saito, Shintaro
    Nakajima, Kenichi
    JOURNAL OF NUCLEAR MEDICINE, 2022, 63
  • [32] Collimator design for improved spatial resolution in SPECT and planar scintigraphy
    Kimiaei, S
    Larsson, SA
    Jacobsson, H
    JOURNAL OF NUCLEAR MEDICINE, 1996, 37 (08) : 1417 - 1421
  • [33] THE DESIGN AND CLINICAL UTILITIES OF A FAN BEAM COLLIMATOR FOR A SPECT SYSTEM
    TSUI, BMW
    GULLBERG, GT
    EDGERTON, ER
    GILLAND, DR
    RHO, T
    JOHNSTON, RE
    PERRY, JR
    MCCARTNEY, WH
    JOURNAL OF NUCLEAR MEDICINE, 1984, 25 (05) : P5 - P6
  • [34] Erratum to: Compatibility of metal additive manufactured tungsten collimator for SPECT/MRI integration
    Amine M. Samoudi
    Karen Van Audenhaege
    Gunter Vermeeren
    Luc Martens
    Roel Van Holen
    Wout Joseph
    EJNMMI Physics, 2 (1)
  • [35] COLLIMATOR SELECTION FOR SPECT BRAIN IMAGING - THE ADVANTAGE OF HIGH-RESOLUTION
    MUELLER, SP
    POLAK, JF
    KIJEWSKI, MF
    HOLMAN, BL
    JOURNAL OF NUCLEAR MEDICINE, 1986, 27 (11) : 1729 - 1738
  • [36] Standardization of brain dopamine transporter SPECT with a fan-beam collimator
    Morita, Keishin
    Maebatake, Akira
    Iwasaki, Rina
    Shiotsuki, Yuki
    Miwa, Kenta
    Himuro, Kazuhiko
    Baba, Shingo
    Sasaki, Masayuki
    JOURNAL OF NUCLEAR MEDICINE, 2016, 57
  • [37] A NEW COLLIMATOR FOR I-123-IMP SPECT IMAGING OF THE BRAIN
    OYAMADA, H
    FUKUKITA, H
    TANAKA, E
    SHIBAHARA, N
    KAWAI, H
    TERUI, S
    JOURNAL OF NUCLEAR MEDICINE, 1985, 26 (05) : P12 - P12
  • [38] Pinhole PET (pPET): A multi-pinhole collimator insert for small animal SPECT imaging on PET cameras
    Cardi, Christopher A.
    Cao, Zixiong
    Thakur, Mathew L.
    Karp, Joel S.
    Acton, Paul D.
    2005 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2005, : 1973 - 1976
  • [39] Development of a high-resolution detection module for the INSERT SPECT/MRI system
    Paolo Busca
    Carlo Fiorini
    Arslan D Butt
    Michele Occhipinti
    Riccardo Quaglia
    Paolo Trigilio
    Gabor Nemeth
    Peter Major
    Tamas Bukki
    Kalman Nagy
    Claudio Piemonte
    Alessandro Ferri
    Alberto Gola
    Jan Rieger
    Thoralf Niendorf
    EJNMMI Physics, 1 (Suppl 1)
  • [40] Validation and Performance Assessment of a Preclinical SiPM-Based SPECT/MRI Insert
    Carminati, M.
    Baratelli, F. M.
    Occhipinti, M.
    Erlandsson, K.
    Nagy, K.
    Nyitrai, Z.
    Czeller, M.
    Kuehne, A.
    Niendorf, T.
    Valtorta, S.
    Belloli, S.
    Moresco, R. M.
    Savi, A.
    Iadanza, A.
    Falini, A.
    Politi, L. S.
    Cadioli, M.
    Hutton, B. E.
    Fiorini, C.
    IEEE TRANSACTIONS ON RADIATION AND PLASMA MEDICAL SCIENCES, 2019, 3 (04) : 483 - 490