Collimator Design for a Brain SPECT/MRI Insert

被引:14
|
作者
Salvado, Debora [1 ]
Erlandsson, Kjell [1 ]
Bousse, Alexandre [1 ]
Occhipinti, Michele [2 ]
Busca, Paolo [2 ]
Fiorini, Carlo [2 ]
Hutton, Brian F. [1 ,3 ]
机构
[1] UCL, Inst Nucl Med, London W1E 6BT, England
[2] Politecn Milan, Dipartimento Elettron Informaz & Bioingn, I-20133 Milan, Italy
[3] Univ Wollongong, Ctr Med Radiat Phys, Sydney, NSW, Australia
关键词
Collimator design; multi-modality; pinhole; slit-slat; SPECT insert; stationary system; SLIT-SLAT; EFFECTIVE DIAMETER; SMALL ANIMALS; RESOLUTION; SYSTEM; SIMULATION; PERFORMANCE; IMAGER;
D O I
10.1109/TNS.2015.2450017
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This project's goal is to design a SPECT insert for a clinical MRI system for simultaneous brain SPECT/MR imaging, with a high-sensitivity collimator and high-resolution detectors. We have compared eight collimator designs, four multi-pinhole and four multi-slit slit-slat configurations. The collimation was designed for a system with 2 rings of 25 5 x 5 cm detectors. We introduce the concept of 1/2-pinhole and 1/2-slit, which are transaxially shared between two adjacent detectors. Analytical geometric efficiency was calculated for an activity distribution corresponding to a human brain and a range of intrinsic detector resolutions R-i and target resolutions R-t at the centre of the FOV. Noise-free data were simulated with and without depth-of-interaction (DOI) information, 0.8 mm R-i and 10 mm R-t FWHM, and reconstructed for uniform, Defrise, Derenzo, and Zubal brain phantoms. Comparing the multi-pinhole and multi-slit slit-slat collimators, the former gives better reconstructed uniformity and transaxial resolution, while the latter gives better axial resolution. Although the 2 x 2-pin-hole and 2-slit designs give the highest sensitivities, they result in a sub-optimal utilisation of the detector FOV. The best options are therefore the 5+2 1/2-pinhole and the 1+2 1/2-slit systems, with sensitivities of 1.8 x 10(-4) and 3.2 x 10(-4), respectively. Noiseless brain phantom reconstructions with the multi-pinhole collimator are slightly superior as compared to slit-slat, in terms of symmetry and accuracy of the activity distribution, but the same is not true when noise is included. DOI information reduces artefacts and improves uniformity in geometric phantoms. Further evaluation is needed with prototype collimators.
引用
收藏
页码:1716 / 1724
页数:9
相关论文
共 50 条
  • [21] Characterization of the Detection Module of the INSERT SPECT/MRI Clinical System
    Occhipinti, M.
    Carminati, M.
    Busca, P.
    Butt, A. D.
    Montagnani, G. L.
    Trigilio, P.
    Piemonte, C.
    Ferri, A.
    Gola, A.
    Bukki, T.
    Czeller, M.
    Nyitrai, Z.
    Papp, Z.
    Nagy, K.
    Fiorini, C.
    IEEE TRANSACTIONS ON RADIATION AND PLASMA MEDICAL SCIENCES, 2018, 2 (06) : 554 - 563
  • [22] Design optimization and evaluation of a human brain SPECT-MRI insert based on high-resolution detectors and slit-slat collimators
    Erlandsson, Kjell
    Salvado, Debora
    Bousse, Alexandre
    Hutton, Brian F.
    2013 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2013,
  • [23] Design considerations for INSERT: A new multi-modality SPECT/MRI system for preclinical and clinical imaging
    Erlandsson, Kjell
    Busca, Paolo
    Bukki, Tamas
    Gola, Alberto
    Salvado, Debora
    Butt, Arslan
    Nemeth, Gabor
    Piemonte, Claudio
    Hutton, Brian
    Fiorini, Carlo
    JOURNAL OF NUCLEAR MEDICINE, 2014, 55
  • [24] Design considerations for INSERT: A new multi-modality SPECT/MRI system for preclinical and clinical imaging
    Erlandsson, Kjell
    Busca, Paolo
    Bukki, Tamas
    Gola, Alberto
    Salvado, Debora
    Butt, Arslan
    Nemeth, Gabor
    Piemonte, Claudio
    Hutton, Brian
    Fiorini, Carlo
    JOURNAL OF NUCLEAR MEDICINE, 2014, 55
  • [25] Compatibility of metal additive manufactured tungsten collimator for SPECT/MRI integration
    Amine M Samudi
    Karen Van Audenhaege
    Gunter Vermeeren
    Luc Martens
    Roel Van Holen
    Wout Joseph
    EJNMMI Physics, 2 (Suppl 1)
  • [26] Evaluation of collimator blurring, attenuation, and scatter corrections for brain SPECT
    Zeniya, T.
    Sakimoto, T.
    Ishida, K.
    Watabe, H.
    Hirano, Y.
    Sohlberg, A.
    Minato, K.
    Iida, H.
    JOURNAL OF NUCLEAR MEDICINE, 2009, 50
  • [27] Performance of a novel collimator for high-sensitivity brain SPECT
    El Fakhri, G
    Ouyang, J
    Zimmerman, RE
    Fischman, AJ
    Kijewski, MF
    MEDICAL PHYSICS, 2006, 33 (01) : 209 - 215
  • [28] Introduction of a novel ultrahigh sensitivity collimator for brain SPECT imaging
    Park, Mi-Ae
    Kijewski, Marie Foley
    Keijzers, Ronnie
    Keijzers, Mark
    Lyon, Morgan C.
    Horky, Laura
    Moore, Stephen C.
    MEDICAL PHYSICS, 2016, 43 (08) : 4734 - 4741
  • [29] Design of a dual-head multi-pinhole collimator for brain SPECT with improved axial sampling
    Konik, Arda
    Kalluri, Kesava
    Chen, Ling
    Mok, Greta
    Fromme, Timothy
    Zubal, George
    Furenlid, Lars
    King, Michael
    JOURNAL OF NUCLEAR MEDICINE, 2018, 59
  • [30] A SiPM-Based Clinical MRI-Compatible SPECT Insert
    Carminati, M.
    D'Adda, I.
    Morahan, A. J.
    Erlandsson, K.
    Nagy, K.
    Nyitrai, Z.
    Czeller, M.
    Moresco, R.
    Savi, A.
    Van Mullekom, P.
    Hutton, B. F.
    Fiorini, C.
    2019 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2019,