Collimator Design for a Brain SPECT/MRI Insert

被引:14
|
作者
Salvado, Debora [1 ]
Erlandsson, Kjell [1 ]
Bousse, Alexandre [1 ]
Occhipinti, Michele [2 ]
Busca, Paolo [2 ]
Fiorini, Carlo [2 ]
Hutton, Brian F. [1 ,3 ]
机构
[1] UCL, Inst Nucl Med, London W1E 6BT, England
[2] Politecn Milan, Dipartimento Elettron Informaz & Bioingn, I-20133 Milan, Italy
[3] Univ Wollongong, Ctr Med Radiat Phys, Sydney, NSW, Australia
关键词
Collimator design; multi-modality; pinhole; slit-slat; SPECT insert; stationary system; SLIT-SLAT; EFFECTIVE DIAMETER; SMALL ANIMALS; RESOLUTION; SYSTEM; SIMULATION; PERFORMANCE; IMAGER;
D O I
10.1109/TNS.2015.2450017
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This project's goal is to design a SPECT insert for a clinical MRI system for simultaneous brain SPECT/MR imaging, with a high-sensitivity collimator and high-resolution detectors. We have compared eight collimator designs, four multi-pinhole and four multi-slit slit-slat configurations. The collimation was designed for a system with 2 rings of 25 5 x 5 cm detectors. We introduce the concept of 1/2-pinhole and 1/2-slit, which are transaxially shared between two adjacent detectors. Analytical geometric efficiency was calculated for an activity distribution corresponding to a human brain and a range of intrinsic detector resolutions R-i and target resolutions R-t at the centre of the FOV. Noise-free data were simulated with and without depth-of-interaction (DOI) information, 0.8 mm R-i and 10 mm R-t FWHM, and reconstructed for uniform, Defrise, Derenzo, and Zubal brain phantoms. Comparing the multi-pinhole and multi-slit slit-slat collimators, the former gives better reconstructed uniformity and transaxial resolution, while the latter gives better axial resolution. Although the 2 x 2-pin-hole and 2-slit designs give the highest sensitivities, they result in a sub-optimal utilisation of the detector FOV. The best options are therefore the 5+2 1/2-pinhole and the 1+2 1/2-slit systems, with sensitivities of 1.8 x 10(-4) and 3.2 x 10(-4), respectively. Noiseless brain phantom reconstructions with the multi-pinhole collimator are slightly superior as compared to slit-slat, in terms of symmetry and accuracy of the activity distribution, but the same is not true when noise is included. DOI information reduces artefacts and improves uniformity in geometric phantoms. Further evaluation is needed with prototype collimators.
引用
收藏
页码:1716 / 1724
页数:9
相关论文
共 50 条
  • [1] Collimator Design for a Brain SPECT/MRI Insert
    Institute of Nuclear Medicine, University College London, London
    W1E 6BT, United Kingdom
    不详
    20133, Italy
    不详
    20133, Italy
    不详
    IEEE Trans Nucl Sci, 4 (1716-1724):
  • [2] Collimator design for a multipinhole brain SPECT insert for MRI
    Van Audenhaege, Karen
    Van Holen, Roel
    Vanhove, Christian
    Vandenberghe, Stefaan
    MEDICAL PHYSICS, 2015, 42 (11) : 6679 - 6689
  • [3] Collimator design for a clinical brain SPECT/MRI insert
    Debora Salvado
    Kjell Erlandsson
    Alexandre Bousse
    Michele Occhipinti
    Paolo Busca
    Carlo Fiorini
    Brian Hutton
    EJNMMI Physics, 1 (Suppl 1)
  • [4] SPECT/MRI INSERT Compatibility: Assessment, Solutions, and Design Guidelines
    Carminati, Marco
    Montagnani, Giovanni L.
    Occhipinti, Michele
    Kuehne, Andre
    Niendorf, Thoralf
    Nagy, Kalman
    Nagy, Andras
    Czeller, Miklos
    Fiorini, Carlo
    IEEE TRANSACTIONS ON RADIATION AND PLASMA MEDICAL SCIENCES, 2018, 2 (04) : 369 - 379
  • [5] Evaluation of a partial ring design for the INSERT SPECT/MRI system
    Kell Erlandsson
    Debora Slavado
    Alexandre Bousse
    Brian Hutton
    EJNMMI Physics, 2 (Suppl 1)
  • [6] INSERT: A Novel Clinical Scanner for Simultaneous SPECT/MRI Brain Studies
    Hutton, B. F.
    Erlandsson, K.
    Salvado, D.
    Occhipinti, M.
    Papp, Z.
    Tolgyesi, B.
    Willems, J.
    Piemonte, C.
    Carminati, M.
    Bukki, T.
    Legrady, G.
    Kuehne, A.
    Nyitrai, Z.
    Niendorf, T.
    Van Mullekom, P.
    Waiczies, H.
    Nagy, K.
    De Francesco, I.
    Mathe, D.
    Ottobrini, L.
    Short, S. C.
    Fiorini, C.
    2017 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2017,
  • [7] Design of an Optimized Multi-pinhole Collimator for Brain SPECT
    Chen, L.
    Mok, G. S.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2015, 42 : S212 - S212
  • [8] Collimator Design in SPECT, an Optimisation Tool
    Fuin, Niccolo
    Bousse, Alexandre
    Pedemonte, Stefano
    Arridge, Simon
    Ourselin, Sebastien
    Hutton, Brian
    2010 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD (NSS/MIC), 2010, : 3149 - 3154
  • [9] The design and manufacture of an annular variable-focusing collimator for highsensitivity brain SPECT
    Ouyang, Jinsong
    El Fakhri, Georges
    Xia, Weishi
    Kijewski, Marie Foley
    Genna, Sebastian
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2006, 53 (05) : 2613 - 2618
  • [10] Multi-pinhole collimator design in different numbers of projections for brain SPECT
    Huang, Wenbo
    Mok, Greta S. P.
    FRONTIERS IN MEDICINE, 2023, 10