Values for Macular Perimetry Using the MP-1 Microperimeter in Normal Subjects

被引:26
|
作者
Shah, Vinay A. [1 ]
Chalam, K. V. [2 ]
机构
[1] Univ Missouri, Kansas City, MO 64110 USA
[2] Univ Florida, Coll Med, Jacksonville, FL USA
关键词
Microperimeter; MP-1; Normative data; Retinal sensitivity; age; SCANNING LASER OPHTHALMOSCOPE; FUNDUS PERIMETRY; RETINAL SENSITIVITY; VISUAL-FIELD; THRESHOLD; EDEMA;
D O I
10.1159/000162111
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To establish values for the fixation pattern and macular sensitivity to light in healthy subjects using the MP-1 microperimeter (Nideck Technologies, Vigonza, Italy). Method: A prospective study of 37 healthy subjects (age: 19-71 years), approved by the institutional review board. Full-threshold microperimetry of the central 12 degrees of the retina (the macula) was performed prospectively, utilizing 16 points with the MP-1. The fixation stability was simultaneously recorded during microperimetry testing. The relationship between the fixation pattern and light sensitivity with age was calculated using linear regression analysis. Results: Microperimetry was performed on 66 eyes. Total mean sensitivity was 18.33 dB (range: 13.0-19 dB). Mean macular sensitivities at 2 degrees, 4 degrees and 6 degrees as measured by the MP-1 were 18.4 +/- 2.0, 18.2 +/- 2.3 and 18.2 +/- 2.1 dB, respectively. The linear regression analysis revealed a 0.019240, 0.0248 and 0.0285 dB per year age-related decline in mean macular sensitivity at 2 degrees, 4 degrees and 6 degrees from the center of the macula, respectively. The total mean fixation stability values within 2 degrees and 4 degrees were 86 and 96%, respectively. The linear regression analysis also revealed a 0.344% and 0.122% per year decline in age-related fixation stability within 2 degrees and 4 degrees, respectively. There were no statistically significant differences between the median retinal sensitivity or fixation stability values for sex or race in the study. Conclusions: The MP-1 may be a reliable tool to measure macular sensitivity. Among normal healthy subjects, there was a linear decline in light sensitivity with increasing age. Copyright (C) 2008 S. Karger AG, Basel
引用
收藏
页码:9 / 13
页数:5
相关论文
共 50 条
  • [21] Fixation Behavior in Primary Open Angle Glaucoma at Early and Moderate Stage Assessed by the MicroPerimeter MP-1
    Shi, Yan
    Liu, Mugen
    Wang, Xin
    Zhang, Chun
    Huang, Ping
    JOURNAL OF GLAUCOMA, 2013, 22 (02) : 169 - 173
  • [22] Assessment of Macular Sensitivity and Fixation Stability by MP-1 Microperimetry in Diabetic Macular Edema
    Kusbeci, Tuncay
    Yavas, Guliz
    Inan, Umit Ubeyt
    TURK OFTALMOLOJI DERGISI-TURKISH JOURNAL OF OPHTHALMOLOGY, 2012, 42 (05): : 310 - 315
  • [23] FIXATION STABILITY MEASUREMENT USING THE MP1 MICROPERIMETER
    Crossland, Michael D.
    Dunbar, Hannah M. P.
    Rubin, Gary S.
    RETINA-THE JOURNAL OF RETINAL AND VITREOUS DISEASES, 2009, 29 (05): : 651 - 656
  • [24] Reduced macular sensitivity in patients with clinically significant diabetic macular edema evaluated with Microperimetry MP-1
    Weingessel, B.
    Richter-Mueksch, S.
    Sacu, S.
    Schmidt-Erfurth, U.
    Vecsei-Marlovits, P. V.
    SPEKTRUM DER AUGENHEILKUNDE, 2008, 22 (03) : 158 - 162
  • [25] Test Retest Reproducibility of the Microperimeter MP3 With Fundus Image Tracking in Healthy Subjects and Patients With Macular Disease
    Palkovits, Stefan
    Hirnschall, Nino
    Georgiev, Stefan
    Leisser, Christoph
    Findl, Oliver
    TRANSLATIONAL VISION SCIENCE & TECHNOLOGY, 2018, 7 (01):
  • [26] Evaluation of fixation stability using different targets with the MP1 microperimeter
    Cesareo, M.
    Manca, D.
    Ciuffoletti, E.
    De Giovanni, V.
    Ricci, F.
    Nucci, C.
    Cerulli, L.
    INTERNATIONAL OPHTHALMOLOGY, 2015, 35 (01) : 11 - 17
  • [27] Evaluation of fixation stability using different targets with the MP1 microperimeter
    M. Cesareo
    D. Manca
    E. Ciuffoletti
    V. De Giovanni
    F. Ricci
    C. Nucci
    L. Cerulli
    International Ophthalmology, 2015, 35 : 11 - 17
  • [28] Introducing parallel processing concepts using the Maspar MP-1 computer
    Duckworth, R.James
    SIGCSE Bulletin (Association for Computing Machinery, Special Interest Group on Computer Science Education), 1994, 26 (01): : 353 - 356
  • [29] MP-1 Biofeedback: Luminous Pattern Stimulus Versus Acoustic Biofeedback in Age Related Macular degeneration (AMD)
    Enzo M. Vingolo
    Serena Salvatore
    Paolo G. Limoli
    Applied Psychophysiology and Biofeedback, 2013, 38 : 11 - 16
  • [30] MP-1 Biofeedback: Luminous Pattern Stimulus Versus Acoustic Biofeedback in Age Related Macular degeneration (AMD)
    Vingolo, Enzo M.
    Salvatore, Serena
    Limoli, Paolo G.
    APPLIED PSYCHOPHYSIOLOGY AND BIOFEEDBACK, 2013, 38 (01) : 11 - 16